Skip to content

Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT 8065752438Recalls & reported problems

Alcon Laboratories, Inc. · 25+ TOTAL PLUS Vitrectory Pak, Beveled 10,000 CPM ULTRAVIT Probe Wide Angle Endoilluminator · Class 2

All models of Constellation, EDGE PLUS, TOTAL PLUS, ULTRAVIT (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Alcon Constellation Total Plus vitrectomy cutter has 207 adverse event reports: no deaths and 5 injuries. Most reports describe the cutter failing to cut during eye surgery, and some describe suction problems. Reports fell to 47 in the last 12 months from 78 in the previous 12. Two recalls are open: one over lids damaged during manufacturing that could affect sterility, and one over sharpness complaints in certain knives.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
203
Deaths reported
0
Injuries reported
5
Recalls naming it
2

Summary

The FDA's MAUDE database contains 203 adverse event reports that list this exact device (DI 00380657524389), received between Jun 9, 2021 and Sep 24, 2026 — classified as 5 injuries, 198 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (115 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 6 reportsApr 2024: 2 reportsMay 2024: 5 reportsJun 2024: 4 reportsJul 2024: 8 reportsAug 2024: 8 reportsSep 2024: 3 reportsOct 2024: 4 reportsNov 2024: 5 reportsDec 2024: 9 reportsJan 2025: 8 reportsFeb 2025: 3 reportsMar 2025: 10 reportsApr 2025: 0 reportsMay 2025: 7 reportsJun 2025: 10 reportsJul 2025: 6 reportsAug 2025: 9 reportsSep 2025: 6 reportsOct 2025: 5 reportsNov 2025: 14 reportsDec 2025: 11 reportsJan 2026: 8 reportsFeb 2026: 2 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 28, 2025Class IIOpen, Classified

    Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being compromised.

    Z-2108-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jul 15, 2024Class IIOpen, Classified

    Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.

    Z-2677-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Cut115
  • Suction Failure30
  • Suction Problem25
  • Air/Gas in Device12
  • Ambient Noise Problem10
  • Leak/Splash7
  • Break6
  • Disconnection5

Patient problems

  • No Clinical Signs, Symptoms or Conditions174
  • Insufficient Information25
  • Retinal Detachment2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Unspecified Eye / Vision Problem1

Recent adverse event reports

The 25 most recent of 203 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Failure to Cut

    A pharmacist reported that the cutter speed remained at two thousand five hundred cuts per min instead of the indicated ten thousand cut per minute. Restarting the cassette did not increase the cutting speed during retinal detachment procedure. The surgery was completed by replacing the cassette. There was no information about patient impact.

    Patient problems: Insufficient Information

    MAUDE report 50877493 ↗
  • MalfunctionSep 24, 2026Failure to Cut

    A pharmacist reported that the cutter speed was limited to 2,500 cuts per minute, whereas the packaging indicated 10000 cuts per minute. It was impossible to increase the cutting rate during surgery. The procedure type was unknown. The surgery was completed on the same day after replacing the product with another one. There was no information about the patient impact. Additional information has been requested but is not available at the time of this report

    Patient problems: Insufficient Information

    MAUDE report 50877276 ↗
  • MalfunctionAug 12, 2026Failure to Cut

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50221638 ↗
  • MalfunctionAug 12, 2026Disconnection; Failure to Cut

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50218760 ↗
  • MalfunctionAug 11, 2026Disconnection; Ambient Noise Problem

    A physician reported that unusual movement and sound were felt in the handle and metal sleeve came loose and fell to the bottom during vitrectomy surgery. The surgery was completed on the same day, but it was unknown how it got completed. There was no patient impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50206354 ↗
  • MalfunctionMar 5, 2026Failure to Cut

    A NURSE REPORTED THAT CUTTER HEAD DID NOT CUT DURING VITREORETINAL SURGERY. THE CUTTER INITIALLY WORKED BUT GRADUALLY THE PERFORMANCE DECREASED DURING SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24516233 ↗
  • MalfunctionMar 5, 2026Suction Problem; Failure to Cut

    A NURSE REPORTED THAT CUTTER DID NOT ASPIRATE, DID NOT CUT AND RELEASED BUBBLES FROM THE ASPIRATION TUBE BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED BY REPLACING THE CUTTER HEAD WITH A NEW ONE WHICH FUNCTIONED NORMALLY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24516228 ↗
  • MalfunctionMar 4, 2026Suction Problem; Failure to Cut; Air/Gas in Device

    A NURSE REPORTED THAT THE CUTTERS DID NOT ASPIRATE, DID NOT CUT, AND RELEASED BUBBLES BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY AFTER REPLACING THE CUTTER HEAD. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24511827 ↗
  • MalfunctionFeb 18, 2026Break; Failure to Cut; Ambient Noise Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE CATHETER DID NOT CUT WHEN INSERTED, IT BEGAN TO MAKE STRANGE NOISE, AND METAL PART OF THE CATHETER DETACHED FROM THE VITRECTOR CONTROL DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED ON SAME DAY BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24376168 ↗
  • MalfunctionFeb 16, 2026Suction Problem; Defective Device

    A HEALTHCARE PROFESSIONAL REPORTED THAT VITRECTOME DID NOT ASPIRATE, IN THE OTHER THERE WAS A FACTORY DEFECT IN THE VITRECTOME TUBES AT AN UNKNOWN TIMING FOR VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED BY REPLACING WITH NEW PAK. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24343695 ↗
  • MalfunctionJan 28, 2026Infusion or Flow Problem

    A HEALTH CARE PROFESSIONAL REPORTED THAT NO INFUSION DURING CATARACT AND VITRECTOMY COMBINED PROCEDURE. THE SURGERY WAS COMPLETED ON THE SAME DAY, AND IT WAS UNKNOWN HOW IT GOT COMPLETED. THERE WAS NO PATIENT CONTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198245 ↗
  • MalfunctionJan 22, 2026Suction Problem; Failure to Cut

    A PHYSICIAN REPORTED THAT PROBE COULD NOT CUT AND ASPIRATE DURING VITRECTOMY SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24142730 ↗
  • MalfunctionJan 21, 2026Improper Flow or Infusion

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT AIR VALVE WOULD NOT OPEN AND WOULD NOT DRAIN DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED ON SAME DAY AFTER REPLACEMENT OF PRODUCT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24127869 ↗
  • MalfunctionJan 20, 2026Infusion or Flow Problem

    A HEALTH CARE PROFESSIONAL REPORTED THAT NO INFUSION DURING CATARACT/VITRECTOMY COMBINATION SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT CONTACT.

    Patient problems: Insufficient Information

    MAUDE report 24112407 ↗
  • MalfunctionJan 17, 2026Leak/Splash

    A PHYSICIAN REPORTED THAT THE LEAKAGE OCCURRED FROM THE BOTTOM OF CASSETTE BEFORE SURGERY. THE PROCEDURE TYPE AND OTHER SURGERY DETAILS WERE UNKNOWN. IT WAS UNKNOWN WHETHER THE SURGERY WAS COMPLETED OR NOT. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 24105304 ↗
  • MalfunctionJan 12, 2026Suction Problem; Failure to Cut

    A PHYSICIAN REPORTED THE ACTUATION OF THE CUTTER WAS STRANGE AND ASPIRATION OF CUTTER WAS ALSO POOR DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS PATIENT CONTACT WITH SUSPECT PRODUCT BUT NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24040160 ↗
  • MalfunctionJan 9, 2026Suction Problem

    A NURSE REPORTED THAT CUTTER WAS MALFUNCTIONED, SUCTION WAS NOT WORKING BEFORE INTRAOCULAR LENS IMPLANTATION SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY BY REPLACING THE PRODUCT. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 24019243 ↗
  • MalfunctionJan 5, 2026Suction Problem; Failure to Cut

    A NURSE REPORTED THAT THE CUTTING AND SUCTION ARE NOT POSSIBLE BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING WITH A NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23954668 ↗
  • MalfunctionDec 29, 2025Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT THE PROBE DID NOT CUT, DURING CATARACT AND VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED AFTER REPLACEMENT OF PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23904563 ↗
  • MalfunctionDec 28, 2025Fluid/Blood Leak; Fluid/Blood Leak

    A PHYSICIAN REPORTED THAT TWO CONSECUTIVE LEAKS FROM THE CLOSURE VALVE FROM THE TROCAR CLOSURE VALVE DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY AFTER REPLACING WITH A NEW PAK AND ONLY THE TROCAR WAS REPLACED. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.

    Patient problems: Insufficient Information

    MAUDE report 23903980 ↗
  • MalfunctionDec 15, 2025Failure to Cut; Failure to Cut

    A CUSTOMER REPORTED THAT CUTTER WAS NO CUTTING DURING SURGERY. THE PROCEDURE TYPE AND COMPLETION DETAILS WERE UNKNOWN. THE PROBE VACUUM WAS NORMAL. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED STATING THAT THE EVENT OCCURRED DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED ON SAME DAY AFTER REPLACEMENT OF PAK. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23801054 ↗
  • MalfunctionDec 14, 2025Failure to Cut; Failure to Cut

    A CUSTOMER REPORTED THAT THE PROBE COULD NOT CUT DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23799336 ↗
  • MalfunctionDec 9, 2025Failure to Cut; Failure to Cut

    A PHYSICIAN REPORTED THAT CUTTING FAILURE OF THE CUTTER OCCURRED DURING SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THE PROCEDURE TYPE WAS UNKNOWN. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23760062 ↗
  • MalfunctionDec 5, 2025Failure to Cut; Failure to Cut

    A NURSE REPORTED THAT BEFORE VITRECTOMY SURGERY, THE VITRECTOMY CUTTER (PROBE) DID NOT CUT. THE ASPIRATION FUNCTION WAS NORMAL. THE PROCEDURE WAS COMPLETED ON THE SAME DAY. THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23726213 ↗
  • MalfunctionDec 4, 2025Failure to Cut; Failure to Cut

    AN OPHTHALMOLOGIST REPORTED THAT PROBE STOPPED CUTTING AND ONLY ABLE TO ASPIRATE DURING THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED BY REPLACING THE CASSETTE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23712731 ↗

Other Vitrectomy, instrument cutter devices

All Vitrectomy, instrument cutter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page