Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Recalls & reported problems
Alcon Laboratories, Inc. · Constellation Vision System 7500 CPM UltraVit Probe Straight Endoilluminator 25+ TotalPlus Vitrectomy Pak EdgePlus Blade Valved Entry System · Class 2
All models of Constellation, Ultravit, 25+, TotalPlus, EdgePlus (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Alcon Constellation Vision System is a machine used in cataract and eye surgery. Many of its 400 adverse event reports describe the cutter failing to cut. The reports list 6 injuries and no deaths. There are 2 open recalls, for possibly damaged lids on some units that could compromise sterility and for sharpness complaints about certain knives. Reports fell from 89 in the prior 12 months to 28 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 399
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 399 adverse event reports that list this exact device (DI 00380657516179), received between Mar 24, 2021 and Sep 30, 2026 — classified as 6 injuries, 393 malfunctions.
The most frequently reported specific device problem is “Failure to Cut” (162 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 28, 2025Class IIOpen, Classified
Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being compromised.
Z-2109-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Jul 15, 2024Class IIOpen, Classified
Alcon has detected an increase in complaint reports related to sharpness for specific ClearCut Dual Bevel Sideport and A-OK Comeal/Scleral V-Lance Ophthalmic knives.
Z-2673-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Cut162
- Fluid/Blood Leak39
- Break37
- Leak/Splash34
- Suction Failure28
- Suction Problem22
- No Flow17
- Complete Blockage16
Patient problems
- No Clinical Signs, Symptoms or Conditions332
- Insufficient Information75
- Intraocular Pressure Decreased3
- Retinal Tear2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Capsular Bag Tear1
- Choroidal Hemorrhage1
- Corneal Abrasion1
Recent adverse event reports
The 25 most recent of 399 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Suction Problem; Failure to Cut
A nurse reported that cutter failed to cut and aspirate during the vitrectomy surgery. The surgery was completed by replacing the cutter. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943551 ↗MalfunctionSep 30, 2026Break
A healthcare professional reported that tip of the vitrectomy probe broke off inside the eye during surgery. The surgery type was unknown. The broken tip was successfully removed from the eye without further complications. The surgery completions details were unknown. There was no information about patient impact.Additional information was requested and non-received till date.
Patient problems: Insufficient Information
MAUDE report 50943104 ↗MalfunctionSep 28, 2026Failure to Cut
A nurse reported that the cutter was not cut during vitrectomy surgery. The surgery was completed on the same day by replacing the product with new one. There was no patient impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915635 ↗MalfunctionAug 17, 2026Use of Device Problem
A health care professional reported that the cutter (vitrectomy probe) was not performing during vitrectomy surgery. The surgery was completed on same day. There was no information about any patient impact.Additional information has been requested but none received till date.
Patient problems: Insufficient Information
MAUDE report 50271405 ↗MalfunctionAug 12, 2026Failure to Cut
Patient problems: Insufficient Information
MAUDE report 50224294 ↗MalfunctionAug 12, 2026Failure to Cut
Patient problems: Insufficient Information
MAUDE report 50224293 ↗MalfunctionAug 12, 2026Failure to Cut
Patient problems: Insufficient Information
MAUDE report 50224287 ↗MalfunctionAug 11, 2026Failure to Cut
Patient problems: Insufficient Information
MAUDE report 50206363 ↗MalfunctionJul 27, 2026Failure to Cut
A non-health care professional reported that cutter did not cut when the doctor pressed the foot pedal at an unknown timing. The procedure details were unknown. There was no information about patient impact.
Patient problems: Insufficient Information
MAUDE report 50005037 ↗MalfunctionJan 23, 2026Material Twisted/Bent
A HEALTH CARE PROFESSIONAL REPORTED THAT PROBE GOT BEND AND THEN DETACHED DURING SURGERY. PROCEDURE TYPE WAS UNKNOWN. THE SURGERY WAS COMPLETED ON SAME DAY. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 24159071 ↗MalfunctionJan 12, 2026Fluid/Blood Leak
A PHARMACIST REPORTED THAT THE VALVED TROCARS BEGAN TO LEAK DURING CALIBRATION OF VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY AFTER REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 24033764 ↗MalfunctionJan 8, 2026Noise, Audible; Air/Gas in Device
A PHYSICIAN REPORTED THAT A STRANGE SOUND WAS HEARD FROM THE CASSETTE AND ALSO, AIR GOT INTO THE CASSETTE A LOT DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24000517 ↗MalfunctionJan 7, 2026Suction Problem; Failure to Cut
A NURSE REPORTED THAT THE CUTTER HAD POOR ACTUATION AND ASPIRATION OF THE CUTTER COULD NOT BE PERFORMED DURING VITRECTOMY SURGERY. THE PROCEDURE WAS COMPLETED BY REPLACING THE PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23989890 ↗MalfunctionJan 2, 2026Disconnection
A HEALTH CARE PROFESSIONAL REPORTED THAT DURING THE VITRECTOMY SURGERY THE TIP OF THE VITRECTOR WAS DETACHED FROM THE VITRECTOR. THE SURGERY WAS COMPLETED ON THE SAME DAY BUT IT WAS UNKNOWN HOW IT WAS COMPLETED. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23942551 ↗MalfunctionDec 29, 2025Break; Failure to Cut; Break; Failure to Cut
AN OPHTHALMIC SURGEON REPORTED THAT CUTTER WAS NOT WORKING WELL (CUTTER WITH 7500 CUTS PER MINUTE WAS NOT WORKING WELL AND HE REDUCED IT TO 5000 CUTS PER MINUTE) AND CUTTER BROKE INSIDE THE TROCAR DURING SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED ON SAME DAY. PROCEDURE TYPE WAS UNKNOWN AND THERE WAS NO INFORMATION ABOUT THE PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23904928 ↗MalfunctionDec 26, 2025Failure to Cut; Ambient Noise Problem; Air/Gas in Device; Failure to Cut; Ambient Noise Problem; Air/Gas in Device
A HEALTH CARE PROFESSIONAL REPORTED THAT CUTTING HAD SOUND AND CREATED A LOT OF BUBBLE AND EVENTUALLY GOT STUCK AT AN UNKNOWN TIMING OF SURGERY. THE PROCEDURAL TYPE WAS UNKNOWN. THE SURGERY COMPLETION AND REPLACEMENT DETAILS WERE UNKNOWN. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23894785 ↗MalfunctionDec 24, 2025Suction Problem; Suction Problem
A HEALTH CARE PROFESSIONAL REPORTED THAT PROBE HAD INSUFFICIENT ASPIRATION EFFICIENCY BEFORE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING WITH NEW PRODUCT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23887519 ↗MalfunctionDec 15, 2025Failure to Cut; Failure to Cut
A NURSE REPORTED THAT PROBE HAD NO CUTTING DURING THE VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY; HOWEVER, THE METHOD OF COMPLETION WAS UNKNOWN. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23803802 ↗MalfunctionDec 15, 2025Suction Problem; Failure to Cut; Suction Problem; Failure to Cut
A NURSE REPORTED THAT THE VITRECTOME STOPPED CUTTING, THE VITRECTOMY PORT WAS BLOCKED BY A JAMMED BLADE DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY AFTER REPLACING THE PRODUCT WITH ANOTHER ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23801888 ↗MalfunctionDec 11, 2025Disconnection; Ambient Noise Problem; Disconnection; Ambient Noise Problem
A NURSE REPORTED THAT EXCESSIVE VIBRATIONS OF THE VITRECTOMY AND GRADUAL LOOSENING OF THE VITRECTOMY TIP DURING VITRECTOMY SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED ON SAME DAY. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23777717 ↗MalfunctionDec 4, 2025Failure to Cut; Failure to Cut
A PHARMACIST REPORTED THAT PROBE DID NOT PERFORM CUTTING AND ASPIRATION WAS NORMAL DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACING WITH NEW PROBE. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23715143 ↗MalfunctionDec 4, 2025Failure to Cut; Failure to Cut
A HEALTHCARE PROFESSIONAL REPORTED THAT CUTTER HEAD STOPPED WORKING DURING VITRECTOMY SURGERY. THE NEW PRODUCT WAS REPLACED TO COMPLETE THE SURGERY. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED THAT THE CUTTING WAS NOT WORKING. THE ASPIRATION AND ACTUATION CONDITION WAS UNKNOWN. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23714821 ↗MalfunctionNov 25, 2025Failure to Cut; Failure to Cut
A PHYSICIAN REPORTED THAT THE VITRECTOMY PROBE HAD POOR CUTTING DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED AFTER REPLACEMENT OF PRODUCT WITH NEW ONE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645072 ↗MalfunctionNov 24, 2025Failure to Cut; Failure to Cut
A CUSTOMER REPORTED VIA HEALTH AUTHORITY THAT THE VITRECTOMY PROBE COULD NOT CUT THE VITREOUS DURING VITRECTOMY SURGERY. THE SURGERY WAS COMPLETED ON THE SAME DAY AFTER REPLACING THE PRODUCT WITH A NEW VITRECTOMY PROBE. THERE WAS NO INFORMATION ABOUT PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 23628528 ↗MalfunctionNov 11, 2025Failure to Cut; Failure to Cut
A HEALTH CARE PROFESSIONAL REPORTED THAT IN A PARS PLANA VITRECTOMY (PPV) SURGERY, THE CUTTER PROBES WERE NOT CUTTING. NURSE CHANGED CUTTER PROBE TWICE UNTIL THE PROBE COULD CUT PROPERLY AND FINISHED CASE IN THAT DAY. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23520633 ↗
FDA clearance and classification
The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617
Has the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617 been recalled?
Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-2109-2025 (Class II), began on Apr 28, 2025: Due to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617?
The FDA's MAUDE database lists 399 adverse event reports for this device, received between Mar 24, 2021 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Cut” (162), “Fluid/Blood Leak” (39) and “Break” (37); the most common patient problems are “Intraocular Pressure Decreased” (3), “Retinal Tear” (2) and “Capsular Bag Tear” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617?
None of the reports are classified as deaths, 6 as injuries and 393 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617, and is it still sold?
The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand Constellation, Ultravit, 25+, TotalPlus, EdgePlus, model 8065751617). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Unit, phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation723 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation631 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation550 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls303 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation367 reports
- Constellation, Ultravit, TotalPlus, EdgePlus 8065751613Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page