CENTURION® 8065752914Recalls & reported problems

Alcon Laboratories, Inc. · Centurion Vision System Active Sentry® Handpiece · Class 2

All models of Centurion (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Alcon Centurion Active Sentry handpiece is used in cataract surgery. Its 560 reports mostly describe energy that fails to deliver, overheating and suction problems. The reports include no deaths and 194 injuries, with some describing eye burns, corneal swelling and toxic inflammation inside the eye. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
550
Deaths reported
0
Injuries reported
184
Recalls naming it
0

Summary

The FDA's MAUDE database contains 550 adverse event reports that list this exact device (DI 00380657529148), received between Mar 23, 2021 and Sep 28, 2026 — classified as 184 injuries, 366 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver Energy” (118 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 11 reportsFeb 2024: 7 reportsMar 2024: 3 reportsApr 2024: 12 reportsMay 2024: 8 reportsJun 2024: 4 reportsJul 2024: 14 reportsAug 2024: 3 reportsSep 2024: 10 reportsOct 2024: 11 reportsNov 2024: 10 reportsDec 2024: 2 reportsJan 2025: 8 reportsFeb 2025: 4 reportsMar 2025: 5 reportsApr 2025: 9 reportsMay 2025: 8 reportsJun 2025: 4 reportsJul 2025: 0 reportsAug 2025: 10 reportsSep 2025: 4 reportsOct 2025: 3 reportsNov 2025: 11 reportsDec 2025: 12 reportsJan 2026: 7 reportsFeb 2026: 2 reportsMar 2026: 4 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 16 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem119
  • Failure to Deliver Energy118
  • Energy Output Problem95
  • Overheating of Device89
  • Appropriate Device Problem Term/Code Not Available73
  • Suction Problem37
  • No Device Output26
  • Complete Blockage23

Patient problems

  • No Clinical Signs, Symptoms or Conditions289
  • Toxic Anterior Segment Syndrome (TASS)85
  • Insufficient Information61
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available55
  • Eye Burn52
  • Fibrosis41
  • Corneal Edema38
  • Blurred Vision17

Recent adverse event reports

The 25 most recent of 550 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026No Device Output

    Patient problems: Insufficient Information

    MAUDE report 50924040 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that patients developed Toxic Anterior Segment Syndrome (TASS) following a procedure involving an ophthalmic handpiece. Procedure details and current condition of patients have not been provided. Additional information has been requested but is not available at the time of this report.Additional information received indicated that a cataract procedure on the right eye had been scheduled.

    Patient problems: Toxic Anterior Segment Syndrome (TASS)

    MAUDE report 50815342 ↗
  • MalfunctionSep 15, 2026No Device Output

    A customer reported that during cataract surgery handpiece stopped working during phacoemulfication. Procedure was completed by using new handpiece and there was no patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50772309 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the tenth of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756252 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the nineth of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756251 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the fourth of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756250 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the eighth of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756249 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the sixth of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756247 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the twelfth of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756245 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the third of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756244 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the fifth of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756243 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the second of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756242 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the eleventh of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50756241 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    A physician reported that, following cataract surgeries performed using an ophthalmic system and ophthalmic handpieces, multiple patients experienced significant corneal edema and swelling on postoperative day one. The physician stated that all 16 patients evaluated on the first postoperative day presented with significant corneal edema, reduced visual acuity of 20/100 or worse, at least 1+ Descemet¿s folds, and 1+ microcystic edema. The patients' current condition is unknown. The physician further reported that the ophthalmic system and handpieces successfully passed Prime and Tune testing, and no Advisory Codes were generated.It was reported that there was no indication that these handpieces were used in the procedures.This report pertains to the first of twelve different handpieces reported from the same facility.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Corneal Edema; Swelling/ Edema

    MAUDE report 50755946 ↗
  • MalfunctionSep 9, 2026Particulates; Obstruction of Flow; Excessive Heating

    A customer reported that milky-white smoke, as described by the surgeon, came out of the tip of an ophthalmic handpiece during a cataract removal procedure with intraocular lens placement. The event occurred during the procedure while the system was in quadrant removal mode, and an occlusion message was displayed. The handpiece was removed and replaced with a different handpiece, and the procedure was completed on the same day. There was patient contact however, no visible patient impact was observed, and no additional medical intervention was required.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50689251 ↗
  • MalfunctionSep 4, 2026Excessive Heating

    FYI: A customer reported that ophthalmic system messages were displayed while using an ophthalmic handpiece and handpiece became hot. Procedure details and timing of the event are unknown. It is also unclear whether the surgery was completed. There was no patient impact. No further information was expected, as the customer was unwilling to provide additional details.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50608393 ↗
  • MalfunctionJul 31, 2026Suction Problem

    Patient problems: Eye Burn

    MAUDE report 50100266 ↗
  • InjuryJul 20, 2026Adverse Event Without Identified Device or Use Problem

    A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY WITH INTRAOCULAR LENS IMPLANTATION, PERFORMED USING TWO DIFFERENT OPHTHALMIC HANDPIECES, THE PATIENT EXPERIENCED AN INCISIONAL BURN AT THE MAIN INCISION SITE. A SUTURE WAS PLACED TO CLOSE THE WOUND, AND THE PROCEDURE WAS COMPLETED. THE EVENT DID NOT APPEAR TO HAVE RESULTED IN SIGNIFICANT ASTIGMATISM. THE PATIENT'S CURRENT CONDITION WAS REPORTED AS RECOVERING WELL. THIS REPORT PERTAINS TO TWO OF TWO DIFFERENT OPHTHALMIC HANDPIECES REPORTED FROM THE SAME FACILITY.

    Patient problems: Eye Burn; Visual Impairment

    MAUDE report 25875365 ↗
  • InjuryJul 20, 2026Adverse Event Without Identified Device or Use Problem

    A PHYSICIAN REPORTED THAT DURING CATARACT SURGERY WITH INTRAOCULAR LENS IMPLANTATION, PERFORMED USING TWO DIFFERENT OPHTHALMIC HANDPIECES, THE PATIENT EXPERIENCED AN INCISIONAL BURN AT THE MAIN INCISION SITE. A SUTURE WAS PLACED TO CLOSE THE WOUND, AND THE PROCEDURE WAS COMPLETED. THE EVENT DID NOT APPEAR TO HAVE RESULTED IN SIGNIFICANT ASTIGMATISM. THE PATIENT'S CURRENT CONDITION WAS REPORTED AS RECOVERING WELL. THIS REPORT PERTAINS TO ONE OF TWO DIFFERENT OPHTHALMIC HANDPIECES REPORTED FROM THE SAME FACILITY.

    Patient problems: Eye Burn; Visual Impairment

    MAUDE report 25875363 ↗
  • MalfunctionApr 20, 2026Failure to Calibrate; Improper Flow or Infusion; Output below Specifications

    A CUSTOMER REPORTED THAT AN OPHTHALMIC HANDPIECE EXHIBITED FLOW FAILURE AND PHACOEMULSIFICATION ISSUES. PROCEDURE DETAILS AND PATIENT IMPACT HAVE NOT BEEN PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Insufficient Information

    MAUDE report 24933833 ↗
  • InjuryMar 26, 2026Particulates

    A HEALTHCARE PROFESSIONAL REPORTED THAT FOREIGN PARTICLES WERE FOUND IN THE ANTERIOR CHAMBER OF A PATIENT¿S EYE AFTER CATARACT SURGERY, WHICH WERE RELEASED FROM THE OPHTHALMIC HANDPIECE. THE PATIENT REQUIRED AN ADDITIONAL SURGICAL PROCEDURE TO REMOVE THE PARTICLES. THE CURRENT CONDITION OF THE PATIENT IS UNKNOWN. THIS COMPLAINT IS PERTAINING THE FIRST OF THREE PATIENTS RECEIVED FROM THE INITIAL REPORTER.

    Patient problems: Insufficient Information

    MAUDE report 24696867 ↗
  • InjuryMar 26, 2026Particulates

    ADDITIONAL INFORMATION RECEIVED REPORTING THAT THE CURRENT PATIENT CONDITION WAS RESOLVED.

    Patient problems: Insufficient Information

    MAUDE report 24696865 ↗
  • MalfunctionMar 18, 2026Improper Flow or Infusion

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING SURGERY, NO WATER CAME OUT OF THE OPHTHALMIC HANDPIECE. PROCEDURE DETAILS AND PATIENT IMPACT HAVE NOT BEEN PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Insufficient Information

    MAUDE report 24629692 ↗
  • InjuryMar 5, 2026Suction Problem

    A NURSE REPORTED THAT PATIENT EXPERIENCED WOUND BURN, CORNEAL INJURY AND CORNEAL OPACITY AND STRICHES REQUIRED. PATIENT WAS PRESCRIBED EYE DROPS POST SURGERY AND ALSO OPHTHALMIC HANDPIECE EXHIBITED NO ASPIRATION DURING THE SCULP MODE OPHTHALMIC SYSTEM AND ALSO DISPLAYED SYSTEM MESSAGE.

    Patient problems: Anterior Capsule Opacification; Corneal Scar; Eye Burn

    MAUDE report 24518722 ↗
  • MalfunctionFeb 26, 2026No Device Output; Failure to Cut

    A CUSTOMER REPORTED THAT OPHTHALMIC HANDPIECE WAS NOT OSCILLATING OR CUTTING DURING SETUP. DETAILS RELATED TO PATIENT HARM HAVE NOT BEEN REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT IS NOT AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24452610 ↗

FDA clearance and classification

The FDA regulates product code HQC (unit, phacofragmentation) under 21 CFR 886.4670 as a Class II device. 21 CFR 886.4670 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CENTURION® 8065752914

Has the CENTURION® 8065752914 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380657529148), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CENTURION® 8065752914?

The FDA's MAUDE database lists 550 adverse event reports for this device, received between Mar 23, 2021 and Sep 28, 2026. In those reports, the most common specific device problems are “Failure to Deliver Energy” (118), “Energy Output Problem” (95) and “Overheating of Device” (89); the most common patient problems are “Toxic Anterior Segment Syndrome (TASS)” (85), “Eye Burn” (52) and “Fibrosis” (41). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CENTURION® 8065752914?

None of the reports are classified as deaths, 184 as injuries and 366 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CENTURION® 8065752914 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the CENTURION® 8065752914, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand CENTURION®, model 8065752914). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Unit, phacofragmentation devices

All Unit, phacofragmentation devices →