WAVELIGHT EX500 EXCIMER LASER 8065990794Recalls & reported problems

Alcon Laboratories, Inc. · WAVELIGHT EX500 EX LASER US · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Alcon WaveLight EX500 excimer laser, used in eye surgery, has 109 adverse event reports, with no deaths and 92 injuries. Many reports describe blurred vision, visual disturbances, or corneal inflammation after laser eye surgery, and a few describe the treatment not working as expected. Reports fell to 5 in the last 12 months from 30 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
109
Deaths reported
0
Injuries reported
92
Recalls naming it
0

Summary

The FDA's MAUDE database contains 109 adverse event reports that list this exact device (DI 00380659907944), received between Mar 4, 2021 and Dec 26, 2025 — classified as 92 injuries, 17 malfunctions.

The most frequently reported specific device problem is “Failure to Capture” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 28 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202628
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem76
  • Failure to Capture9
  • Insufficient Device Problem Information8
  • Energy Output Problem7
  • Failure to Align3
  • Contamination2
  • Product Quality Problem2
  • Device Displays Incorrect Message1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available34
  • Blurred Vision34
  • Visual Disturbances34
  • Keratitis30
  • Visual Impairment19
  • Dry Eye(s)10
  • Insufficient Information10
  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 25 most recent of 109 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT VISUAL ACUITY WAS STABLE AND CLEAR BUT THE PATIENT HAS EXPERIENCED LONGSTANDING FLOATERS AND FLASHES IN PERIPHERY IN THE RIGHT EYE AFTER REFRACTIVE SURGERY ALONG WITH APPLICATION OF MEDICATION. PATIENT REPORTS HAS NOT BEEN USING LID SCRUBS. THERE ARE TWO RELATED REPORTS FOR THIS PATIENT. THIS REPORT ADDRESSES THE PATIENT'S RIGHT EYE AND ANOTHER MANUFACTURER REPORT WILL BE FILED FOR THE FELLOW EYE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23896152 ↗
  • InjuryDec 26, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT VISUAL ACUITY WAS STABLE AND CLEAR BUT THE PATIENT HAS EXPERIENCED LONGSTANDING FLOATERS AND FLASHES IN PERIPHERY IN THE LEFT EYE AFTER REFRACTIVE SURGERY ALONG WITH APPLICATION OF MEDICATION. PATIENT REPORTS HAS NOT BEEN USING LID SCRUBS. THERE ARE TWO RELATED REPORTS FOR THIS PATIENT. THIS REPORT ADDRESSES THE PATIENT'S LEFT EYE AND ANOTHER MANUFACTURER REPORT WILL BE FILED FOR THE FELLOW EYE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 23896151 ↗
  • MalfunctionDec 16, 2025Contamination; Product Quality Problem; Contamination; Product Quality Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DIAGNOSTIC DEVICE IMAGE USED LOOKED DIFFERENT AND MEIBOMIAN DEBRIS STILL PRESENT IN LEFT EYE DURING REFRACTIVE SURGERY. THERE ARE TWO RELATED REPORTS FOR THIS PATIENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), LEFT EYE AND ANOTHER MANUFACTURER REPORT WILL BE FILED FOR THE FELLOW EYE.

    Patient problems: Insufficient Information

    MAUDE report 23813801 ↗
  • MalfunctionDec 12, 2025Contamination; Product Quality Problem; Contamination; Product Quality Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE SYSTEM WAS READY TO CAPTURE THE INFRARED IMAGE THE DIAGNOSTIC DEVICE IMAGE APPEARED DIFFERENT THAN EXPECTED. ADDITIONALLY, IT WAS REPORTED THAT MEIBOMIAN DEBRIS WAS PRESENT IN THE RIGHT EYE POSTOPERATIVELY REFRACTIVE SURGERY. THERE ARE TWO RELATED REPORTS FOR THIS PATIENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), RIGHT EYE AND ANOTHER MANUFACTURER REPORT WILL BE FILED FOR THE FELLOW EYE

    Patient problems: Insufficient Information

    MAUDE report 23789077 ↗
  • InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED OVER CORRECTION IN THE LEFT EYE WITH THE REFRACTIVE OUTCOME WAS GREATER THAN ONE DIOPTER IN MANIFEST REFRACTION SPHERICAL EQUIVALENT TREATMENT VALUES AFTER REFRACTIVE SURGERY.

    Patient problems: Visual Impairment

    MAUDE report 23641402 ↗
  • MalfunctionSep 19, 2025Unexpected Therapeutic Results; Unexpected Therapeutic Results

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE TREATMENT IN THE LEFT EYE OF A PATIENT, THE SYSTEM FLAP HAS STARTED TO COME INTO RANGE WHILE THE TREATMENT WAS FIRING. THE LIGHTS NEEDED TO BE TURNED DOWN TO EXPAND A SMALL PUPIL, THE DOCTOR COULD NOT SEE THE FLAP GOING ONTO THE ABLATION AREA ALERTED AND PULLED THE FLAP BACK ACCORDINGLY BUT SOME OF THE TREATMENT APPEARED TO BE BLOCKED BY THE FLAP PRIOR RESULTING IN AN INCOMPLETE CORNEAL ABLATION DURING LASIK SURGERY. THE PATIENT IS EXPERIENCING GOOD VISUAL OUTCOMES IN THAT LEFT EYE. RIGHT WAS UNAFFECTED AND TREATED WITH NO ISSUES.

    Patient problems: Insufficient Information

    MAUDE report 23095165 ↗
  • InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21650616 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED PATIENT DEVELOPED DIFFUSE LAMELLAR KERATITIS OR CENTRAL TOXIC KERATOPATHY WITH POST OPERATIVE SYMPTOMS OF BLURRED VISION, LIGHT SENSITIVITY AND DISCOMFORT IN THE IN THE RIGHT EYE OF THE PATIENT AFTER LASIK SURGERY. PATIENT SYMPTOMS HAVE NOT BEEN RESOLVED AND CURRENTLY UNDERGOING TREATMENT WITH ORALS AND DUREZOL. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT AP, RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21645580 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21645566 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT SR, LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21645330 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT MB, LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21645328 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT JH, LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21645195 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT BL, LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21645194 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT MR, LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21645042 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (UNK), RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21644904 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21644102 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21643922 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT SR, RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21643791 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT BL, RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21643659 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21643522 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT MR, RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21643511 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT BK, RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21643298 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT SL, RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21643296 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE RIGHT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (B)(6), RIGHT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21642857 ↗
  • InjuryMar 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT HAD EXPERIENCED DIFFUSE LAMELLAR KERATITIS AND CENTRAL TOXIC KERATOPATHY, WITH SPECIFIC SYMPTOMS INCLUDING BLURRY VISION AND LIGHT SENSITIVITY IN THE LEFT EYE AFTER LASIK SURGERY. THE PATIENT IS CURRENTLY UNDERGOING TREATMENT WITH ORAL MEDICATIONS AND DUREZOL, BUT THE SYMPTOMS HAVE NOT RESOLVED. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE PATIENT (B)(6)'S, IN THE LEFT EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Blurred Vision; Keratitis; Visual Disturbances

    MAUDE report 21642641 ↗

Other Excimer Laser System devices

All Excimer Laser System devices →

Sources and updates