MEL 90 1900-109Recalls & reported problems
Carl Zeiss Meditec AG · The Carl Zeiss Meditec MEL 90 Excimer Laser System is designed for refractive surgery based on the ablation of corneal tissue achieved with a short pulse excimer laser having a wavelength of 193 nanometers. The laser head emits 4 to 7 nanosecond pulses (FWHM nominal pulse duration) with a repetition rate of 500 Hz. The MEL 90 Excimer Laser is a spot-scanning laser that utilizes a Gaussian beam with a 0.7 mm spot diameter. Indications for Use: The MEL 90 is indicated for use in primary Laser-Assisted in situ Keratomileusis (LASIK) treatments for the reduction or elimination of: • Myopia less in magnitude than -10.00 D sphere (in minus-cylinder notation), with and without cylinder up to -4.00 D, when MRSE is less in magnitude than -10.00 D; • Hyperopia up to +4.00 D sphere (in plus-cylinder notation), with and without cylinder up to +3.00 D, when MRSE is up to +5.00 D; and • Mixed astigmatism with cylinder from >1.00 D up to 4.00 D; in patients who are 18 years of age or older with documentation of stable manifest refraction over the past year as demonstrated by a change in MRSE within +/- 0.50 D. · Class 3
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04049471095074), received between Jan 23, 2026 and Sep 23, 2026 — classified as 10 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak13
Patient problems
- Unspecified Respiratory Problem10
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Gas/Air Leak
During a routine maintenance inspection, a ZEISS engineer detected an irritating odor and confirmed a leakage of the mixed fluorine argon gas (ArF gas) which is necessary to use the excimer laser of MEL 90 device. The ZEISS engineer detected abnormal gas consumption and identified the main valve Premix Valve MV-ArF to be defective and causing a gas leakage. The ZEISS engineer corrected the issue during the maintenance of the device. After replacing the Premix ArF valve, the equipment returned to normal.Based on the available information, no negative impact, e.g. injury, nausea, feeling unwell or nose bleeding, was reported for any patient, user, third party, or the ZEISS engineer involved in the service activity.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50870477 ↗MalfunctionFeb 2, 2026Gas/Air Leak
AT A CUSTOMER SITE IN INDIA, IT WAS REPORTED THAT THE MAIN VALVE (MAGNETIC VALVE CLOSE TO PREMIX GAS CYLINDER) HAD A GAS LEAKAGE AND IT WAS OBSERVED THAT THE ELECTRO MAGNETIC VALVE HAD CRACKS. THE O-RING (SEALING RING) WAS REPLACED, THE GAS LEAK ARRESTED. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24239763 ↗MalfunctionFeb 2, 2026Gas/Air Leak
CUSTOMER IN GERMANY REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE MAGNETIC VALVE. THE CUSTOMER WAS INSTRUCTED BY PHONE HOW TO CLOSE THE GAS VALVE ON THE GAS BOTTLE. THE CUSTOMER WAS ALSO INSTRUCTED NOT TO USE THE DEVICE. THERE WAS NO ADVERSE EVENT AND NO SERIOUS INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24235382 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172925 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172874 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172773 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172747 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172499 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172367 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172270 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24172043 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24171757 ↗InjuryJan 23, 2026Gas/Air Leak
CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.
Patient problems: Unspecified Respiratory Problem
MAUDE report 24171436 ↗
Other Excimer Laser System devices
- WAVELIGHT EX500 EXCIMER LASER 8065990794Alcon Laboratories, Inc. · Excimer Laser System109 reports
- STAR S4 IR 0030-4864Johnson & Johnson Surgical Vision, Inc. · Excimer laser system98 reports
- IDESIGN ADVANCED WAVESCAN STUDIO G300Johnson & Johnson Surgical Vision, Inc. · Excimer laser system20 reports
- IDESIGN RS G301Johnson & Johnson Surgical Vision, Inc. · Excimer laser system10 reports
- Quest™ EC-5000Nidek CO.,Ltd. · Excimer Laser System3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.