MEL 90 1900-109Recalls & reported problems

Carl Zeiss Meditec AG · The Carl Zeiss Meditec MEL 90 Excimer Laser System is designed for refractive surgery based on the ablation of corneal tissue achieved with a short pulse excimer laser having a wavelength of 193 nanometers. The laser head emits 4 to 7 nanosecond pulses (FWHM nominal pulse duration) with a repetition rate of 500 Hz. The MEL 90 Excimer Laser is a spot-scanning laser that utilizes a Gaussian beam with a 0.7 mm spot diameter. Indications for Use: The MEL 90 is indicated for use in primary Laser-Assisted in situ Keratomileusis (LASIK) treatments for the reduction or elimination of: • Myopia less in magnitude than -10.00 D sphere (in minus-cylinder notation), with and without cylinder up to -4.00 D, when MRSE is less in magnitude than -10.00 D; • Hyperopia up to +4.00 D sphere (in plus-cylinder notation), with and without cylinder up to +3.00 D, when MRSE is up to +5.00 D; and • Mixed astigmatism with cylinder from >1.00 D up to 4.00 D; in patients who are 18 years of age or older with documentation of stable manifest refraction over the past year as demonstrated by a change in MRSE within +/- 0.50 D. · Class 3

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04049471095074), received between Jan 23, 2026 and Sep 23, 2026 — classified as 10 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 10 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak13

Patient problems

  • Unspecified Respiratory Problem10
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Gas/Air Leak

    During a routine maintenance inspection, a ZEISS engineer detected an irritating odor and confirmed a leakage of the mixed fluorine argon gas (ArF gas) which is necessary to use the excimer laser of MEL 90 device. The ZEISS engineer detected abnormal gas consumption and identified the main valve Premix Valve MV-ArF to be defective and causing a gas leakage. The ZEISS engineer corrected the issue during the maintenance of the device. After replacing the Premix ArF valve, the equipment returned to normal.Based on the available information, no negative impact, e.g. injury, nausea, feeling unwell or nose bleeding, was reported for any patient, user, third party, or the ZEISS engineer involved in the service activity.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870477 ↗
  • MalfunctionFeb 2, 2026Gas/Air Leak

    AT A CUSTOMER SITE IN INDIA, IT WAS REPORTED THAT THE MAIN VALVE (MAGNETIC VALVE CLOSE TO PREMIX GAS CYLINDER) HAD A GAS LEAKAGE AND IT WAS OBSERVED THAT THE ELECTRO MAGNETIC VALVE HAD CRACKS. THE O-RING (SEALING RING) WAS REPLACED, THE GAS LEAK ARRESTED. NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24239763 ↗
  • MalfunctionFeb 2, 2026Gas/Air Leak

    CUSTOMER IN GERMANY REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE MAGNETIC VALVE. THE CUSTOMER WAS INSTRUCTED BY PHONE HOW TO CLOSE THE GAS VALVE ON THE GAS BOTTLE. THE CUSTOMER WAS ALSO INSTRUCTED NOT TO USE THE DEVICE. THERE WAS NO ADVERSE EVENT AND NO SERIOUS INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24235382 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172925 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172874 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172773 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172747 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172499 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172367 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172270 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24172043 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem

    MAUDE report 24171757 ↗
  • InjuryJan 23, 2026Gas/Air Leak

    CUSTOMER IN CHINA REPORTED AN ABNORMAL SMELL FROM THE MEL 90 DEVICE. THE MEL 90 EQUIPMENT AT THE CUSTOMER'S SITE EXPERIENCED A LEAKAGE OF EXCIMER GAS DUE TO A MALFUNCTION OF THE SOLENOID VALVE. TEN PEOPLE WERE IN THE ROOM AT THE TIME OF THE INCIDENT AND THE AFFECTED INDIVIDUALS EXPERIENCED THROAT AND RESPIRATORY DISCOMFORT, FURTHER MEDICAL EXAMINATIONS WERE CONDUCTED. THE EVENT MAY HAVE JEOPARDIZED THE INVOLVED PEOPLE AND MAY HAVE REQUIRED MEDICAL OR SURGICAL INTERVENTION (TREATMENT) TO ADDRESS THESE SYMPTOMS. A SERIOUS MEDICALLY IMPORTANT EVENT OCCURRED BECAUSE OF EXPOSURE TO THE GAS.

    Patient problems: Unspecified Respiratory Problem