LenSx 8065998162Recalls & reported problems

Alcon Laboratories, Inc. · LenSx Laser System · Class 2

All models of LenSx (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Alcon LenSx is a femtosecond laser system used in eye surgery. Of the 603 reports naming it, many describe the laser failing to align or to finish a cut during refractive surgery, with some describing capsule tears. The reports include 222 injuries and no deaths, and reports fell from 143 in the previous 12 months to 57 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
591
Deaths reported
0
Injuries reported
215
Recalls naming it
0

Summary

The FDA's MAUDE database contains 591 adverse event reports that list this exact device (DI 00380659981623), received between Mar 1, 2021 and Sep 21, 2026 — classified as 215 injuries, 376 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (264 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 23 reportsFeb 2024: 39 reportsMar 2024: 17 reportsApr 2024: 2 reportsMay 2024: 4 reportsJun 2024: 6 reportsJul 2024: 13 reportsAug 2024: 11 reportsSep 2024: 7 reportsOct 2024: 12 reportsNov 2024: 6 reportsDec 2024: 12 reportsJan 2025: 11 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 45 reportsMay 2025: 7 reportsJun 2025: 30 reportsJul 2025: 13 reportsAug 2025: 10 reportsSep 2025: 1 reportsOct 2025: 4 reportsNov 2025: 5 reportsDec 2025: 13 reportsJan 2026: 6 reportsFeb 2026: 8 reportsMar 2026: 4 reportsApr 2026: 5 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202645
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align264
  • Adverse Event Without Identified Device or Use Problem121
  • Appropriate Device Problem Term/Code Not Available90
  • Failure to Cut65
  • Unintended Movement17
  • Application Program Freezes or Fails to Launch13
  • Insufficient Device Problem Information8
  • Decrease in Suction7

Patient problems

  • Insufficient Information235
  • No Clinical Signs, Symptoms or Conditions145
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available119
  • Capsular Bag Tear92
  • Corneal Edema7
  • Corneal Perforation7
  • Blurred Vision3
  • Corneal Abrasion2

Recent adverse event reports

The 25 most recent of 591 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Failure to Cut

    A non healthcare professional reported that flap creation was poor during refractive surgery in the unknown eye of the patient.

    Patient problems: Insufficient Information

    MAUDE report 50841117 ↗
  • MalfunctionSep 17, 2026Failure to Cut

    A non healthcare professional reported that the flap resistance in the unknown eye of a patient, during refractive surgery.

    Patient problems: Insufficient Information

    MAUDE report 50798805 ↗
  • MalfunctionAug 24, 2026Failure to Cut

    A non-healthcare professional reported that the in this report of the during the capsulotomy procedure, at approximately the twelve-o clock position, the laser failed to complete the cut in unknown eye during refractive surgery.

    Patient problems: Insufficient Information

    MAUDE report 50402378 ↗
  • MalfunctionAug 17, 2026Failure to Cut

    A non-healthcare professional reported uncut capsulotomy about three to four clock hours inferior in the left eye during refractive surgery.There was no error message displayed, suction was maintained for the entire procedure. The Capsulorhexis was completed manually.There are multiple related reports for this facility. This report addresses the patient JJ, in the left eye and other manufacturer reports will be filed.

    Patient problems: Insufficient Information

    MAUDE report 50279604 ↗
  • MalfunctionApr 21, 2026Failure to Cut

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING CREATION OF FLAP A SURGEON OBSERVED STICKY FLAP IN THE RIGHT EYE OF A PATIENT DURING SURGERY ISSUE DURING REFRACTIVE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 24941549 ↗
  • MalfunctionApr 16, 2026Failure to Cut; Appropriate Device Problem Term/Code Not Available

    A HEALTHCARE PROFESSIONAL REPORTED THAT SYSTEM DISPLAYED OPTICAL COHERENCE TOMOGRAPHY ERROR DURING DOCKING AND ADDITIONALLY FLAP WAS INCOMPLETE IN THE UNKNOWN EYE OF THE PATIENT DURING REFRACTIVE SURGERY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 24895354 ↗
  • MalfunctionApr 16, 2026Unintended Movement

    A NON HEALTHCARE PROFESSIONAL REPORTED GANTRY WAS MOVING DOWN ON ITS OWN INTERMITTENTLY IN THE UNKNOWN EYE OF THE PATIENT DURING UNKNOWN TIMING.

    Patient problems: Insufficient Information

    MAUDE report 24893121 ↗
  • InjuryApr 8, 2026Adverse Event Without Identified Device or Use Problem

    A PHYSICIAN REPORTED THAT A PATIENT'S ANTERIOR CAPSULE MEMBRANE WAS TORN IN A NINETY-HUNDRED-DEGREE DIRECTION IN LEFT EYE OF THE PATIENT DURING REFRACTIVE SURGERY. THE INTRAOCULAR LENS WAS IMPLANTED, AND THE SURGERY WAS COMPLETED. THE PATIENT'S POSTOPERATIVE VISUAL EFFECTS HAD NOT BEEN AFFECTED

    Patient problems: Capsular Bag Tear

    MAUDE report 24813803 ↗
  • MalfunctionApr 3, 2026Display or Visual Feedback Problem; Failure to Align

    A PHYSICIAN REPORTED THAT CAPSULOTOMY WAS DECENTERED AND AS A RESULT SYSTEM GENERATED AN UNEXPECTED IMAGE IN THE RIGHT EYE OF THE PATIENT DURING REFRACTIVE SURGERY. THE TREATMENT OUTCOME WAS UNFORTUNATE BUT THE PATIENT IS EXPECTED TO DO WELL POSTOPERATIVELY. THERE WAS NO CAPSULAR TEAR AND THE SURGEON DID NOT RE DOCK AS WELL. HE DOCKED ON THE EYE ONCE AND MAINTAINED SUCTION IN THE ENTIRE PROCEDURE. FURTHER THE SURGEON WHEN PUSHED THE ADVANCED PREPOSITIONING BUTTON, THE SYSTEM SHOULD HAVE PROMPTED THE DEVICE TO CAPTURE AN UPDATED DOCKED IMAGE BUT IT DID NOT.

    Patient problems: Insufficient Information

    MAUDE report 24783857 ↗
  • MalfunctionMar 13, 2026Failure to Cut

    A PHYSICIAN REPORTED THAT THE INADEQUATE INCISION PART WAS FOUND ON THE FLAP CREATION, WHEN THEY CHECKED A LINE-SHAPED SCRATCH WAS FOUND ON THE PATIENT INTERFACE SURFACE, FLAP INCISION WAS PERFORMED MANUALLY IN UNKNOWN EYE OF THE PATIENT DURING REFRACTIVE SURGERY. THE CORNEA WAS UNEVEN, SURGEON WAS NOTICED SIMILAR DEFECTS SINCE LAST MONTH AND TOO MANY PROBLEMS.

    Patient problems: Insufficient Information

    MAUDE report 24598128 ↗
  • MalfunctionMar 11, 2026Appropriate Device Problem Term/Code Not Available

    A NURSE REPORTED THAT DURING THE CREATION OF FLAP, THE FLAP BECOMES SUPERFICIAL IN THE TEMPORAL REGION IN UNKNOWN EYE OF THE PATIENT DURING REFRACTIVE SURGERY. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 24566739 ↗
  • MalfunctionMar 9, 2026Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE SYSTEM NOT CUTTING THE CAPSULOTOMY IN PROPER POSITION IN THE UNKNOWN EYE OF A PATIENT, DURING REFRACTIVE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 24542584 ↗
  • MalfunctionMar 6, 2026Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE ENERGY CALIBRATION WAS NOT REMAINING STABLE AND UNABLE TO PERFORM THE FLAP RESULTED THE CUT WAS INCOMPLETE IN THE UNKNOWN EYE OF A PATIENT, DURING REFRACTIVE SURGERY. THE PROCEDURE WAS NOT COMPLETED. THE SYSTEM DID NOT COMPLETE THE RHEXIS. THE SURGEON COMPLETED THE RHEXIS MANUALLY IN THE OPERATING ROOM.

    Patient problems: Insufficient Information

    MAUDE report 24530582 ↗
  • MalfunctionFeb 27, 2026Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT FLAP RESISTANCE IN THE UNKNOWN EYE DURING REFRACTIVE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 24469757 ↗
  • InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT PATIENT HAD EXPERIENCED ANTERIOR CAPSULAR TEAR THAT TORE AROUND TO THE POSTERIOR CAPSULE IN RIGHT EYE OF THE PATIENT DURING REFRACTIVE SURGERY.

    Patient problems: Capsular Bag Tear

    MAUDE report 24455399 ↗
  • MalfunctionFeb 26, 2026Failure to Cut; Firing Problem

    A PHYSICIAN REPORTED THAT SYSTEM NOT EMITTING LASER AND FLAP BECAME FIRM AS A RESULT FLAP COULD NOT BE CUT PROPERLY IN THE LEFT EYE OF THE PATIENT DURING REFRACTIVE SURGERY, ADDITIONALLY POSTOPERATIVE VISUAL OUTCOME NOT REACHING THE EXPECTED LEVEL. PATIENT REPORTED VISUAL DISTURBANCE WHICH PERSISTED. THE SYMPTOMS ARE CONTINUING.

    Patient problems: Visual Impairment

    MAUDE report 24452406 ↗
  • InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT THE ANTERIOR TEAR IN THE UNKNOWN EYE OF A PATIENT, DURING REFRACTIVE SURGERY.

    Patient problems: Capsular Bag Tear

    MAUDE report 24402760 ↗
  • InjuryFeb 20, 2026Failure to Cut

    A PHYSICIAN REPORTED THAT DURING THE CREATION OF THE FLAP, THE CONTINUOUS CURVILINEAR CAPSULORHEXIS WAS PARTIALLY INCOMPLETE IN THE UNKNOWN EYE DURING REFRACTIVE SURGERY. THE SURGEON SUBSEQUENTLY COMPLETED THE CAPSULORHEXIS USING FORCEPS, AND THE SURGERY PROCEEDED WITHOUT FURTHER COMPLICATION. THE PROCEDURE COMPLETED IN SAME DAY, NO POSTOPERATIVE HEALTH ISSUES WERE REPORTED FOR THE PATIENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24398379 ↗
  • MalfunctionFeb 17, 2026Failure to Align

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT SYSTEM CUTS ARE NOT CENTERED IN UNKNOWN EYE OF THE PATIENT DURING REFRACTIVE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 24361822 ↗
  • InjuryFeb 14, 2026Failure to Cut

    A PHYSICIAN REPORTED THAT DURING THE CREATION OF A FLAP A CONTINUOUS CURVILINEAR CAPSULORHEXIS, OCCURRED AT THE FOUR TO SIX O¿CLOCK POSITION. THE SURGEON PROCEEDED TO COMPLETE THE SURGERY USING FORCEPS IN AN UNKNOWN EYE DURING A REFRACTIVE PROCEDURE. NO PATIENT HARM WAS REPORTED. THE SURGEON MANUALLY RECOVERED THE CONTINUOUS CURVILINEAR CAPSULORHEXIS WITHOUT ANY ISSUES. THE INVOLVED EYE AND THE PATIENT IDENTIFIER ARE NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24340869 ↗
  • MalfunctionFeb 2, 2026Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE TREATMENTS WAS OFF BY TWO MILLIMETERS IN CUT PLACEMENT FROM THE ACTUAL CUT, STATING THAT THE CUTS WERE MORE SUPERIOR THAN THE SURGEON PLACED THEM WITH THE OVERLAYS IN THE UNKNOWN EYE OF A PATIENT, DURING REFRACTIVE SURGERY. THERE ARE MULTIPLE RELATED REPORTS FOR THIS FACILITY. THIS REPORT ADDRESSES THE PATIENT'S TWO UNKNOWN EYE AND ADDITIONAL MANUFACTURER REPORTS WILL BE FILED FOR THE OTHER PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24232012 ↗
  • MalfunctionJan 29, 2026Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE TREATMENTS WAS OFF BY TWO MILLIMETERS IN CUT PLACEMENT FROM THE ACTUAL CUT, STATING THAT THE CUTS WERE MORE SUPERIOR THAN THE SURGEON PLACED THEM WITH THE OVERLAYS IN THE UNKNOWN EYE OF A PATIENT, DURING REFRACTIVE SURGERY. THERE ARE TWO RELATED REPORTS FOR THIS FACILITY. THIS REPORT ADDRESSES THE PATIENT'S ONE UNKNOWN EYE AND ADDITIONAL MANUFACTURER REPORTS WILL BE FILED FOR THE OTHER PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 24207032 ↗
  • InjuryJan 16, 2026Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PROCEDURE, THERE IS NO INCOMPLETE CONTINUOUS CURVILINEAR CAPSULORHEXIS AND HAD LINEAR TEARS APPEARED ALONG THE CAPSULE OCCURRED IN THE UNKNOWN EYE DURING REFRACTIVE SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE UNKNOWN PATIENT INITIAL'S, IN THE UNKNOWN EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Capsular Bag Tear

    MAUDE report 24095574 ↗
  • InjuryJan 16, 2026Adverse Event Without Identified Device or Use Problem

    A HEALTHCARE PROFESSIONAL REPORTED THAT DURING THE PROCEDURE, AN INCOMPLETE CONTINUOUS CURVILINEAR CAPSULORHEXIS AND CAPSULAR WEAKENING AND POSTERIOR CAPSULE TEAR OCCURRED IN THE UNKNOWN EYE DURING REFRACTIVE SURGERY. THE SURGERY WAS COMPLETED ON SAME DAY. THERE ARE MULTIPLE RELATED REPORTS FOR THIS EVENT. THIS REPORT ADDRESSES THE UNKNOWN PATIENT INITIAL'S, IN THE UNKNOWN EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Capsular Bag Tear

    MAUDE report 24095322 ↗
  • MalfunctionJan 13, 2026Failure to Cut

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE TAG OCCURRED AND DID NOT PERFORM THE MAIN INCISION AND NOT ABLE TO COMPLETE CAPSULORHEXIS IN THE UNKNOWN EYE OF A PATIENT, DURING REFRACTIVE SURGERY. THERE ARE MULTIPLE RELATED REPORTS FOR THIS FACILITY. THIS REPORT ADDRESSES THE PATIENT 1, THE UNKNOWN EYE AND OTHER MANUFACTURER REPORTS WILL BE FILED.

    Patient problems: Insufficient Information

    MAUDE report 24050984 ↗

FDA clearance and classification

Cleared through 510(k) K123120 on Dec 12, 2012 (decision: Substantially Equivalent). Submitted by Alcon Lensx, Inc. as “Lensx Laser System”. The FDA review took 70 days. FDA record ↗

The FDA regulates product code OOE (ophthalmic femtosecond laser) under 21 CFR 886.4390 as a Class II device. 21 CFR 886.4390 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the LenSx 8065998162

Has the LenSx 8065998162 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380659981623), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LenSx 8065998162?

The FDA's MAUDE database lists 591 adverse event reports for this device, received between Mar 1, 2021 and Sep 21, 2026. In those reports, the most common specific device problems are “Failure to Align” (264), “Failure to Cut” (65) and “Unintended Movement” (17); the most common patient problems are “Capsular Bag Tear” (92), “Corneal Edema” (7) and “Corneal Perforation” (7). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LenSx 8065998162?

None of the reports are classified as deaths, 215 as injuries and 376 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the LenSx 8065998162 FDA approved?

It was cleared through 510(k) K123120 on Dec 12, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the LenSx 8065998162, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand LenSx, model 8065998162). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Ophthalmic femtosecond laser devices

All Ophthalmic femtosecond laser devices →