Skip to content

LIQUID OPTICS INTERFACE 0180-1201Recalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · LIQUID OPTICS INTERFACE, SIZE 12, GEN2 · Class 2

All models of LIQUID OPTICS INTERFACE (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Johnson & Johnson Surgical Vision Liquid Optics Interface is a suction piece used with an ophthalmic femtosecond laser during eye surgery. Most of the 51 reports describe the suction dropping out or breaking down, sometimes during laser firing, which interrupted procedures; a few mention contamination or packaging that was not sealed or not sterile. Reports describe 2 injuries and 49 malfunctions, with no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 05050474609778), received between Aug 15, 2019 and Aug 7, 2026 — classified as 2 injuries, 48 malfunctions.

The most frequently reported specific device problem is “Decrease in Suction” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Decrease in Suction38
  • Contamination4
  • Delivered as Unsterile Product4
  • Manufacturing, Packaging or Shipping Problem3
  • Unsealed Device Packaging2
  • Break1
  • Device Difficult to Setup or Prepare1
  • Failure to Sense1

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • Chemosis1
  • Corneal Perforation1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2026Decrease in Suction

    Customer reported that their suction broke during scans and laser firing. Procedure was not completed successfully. No patient injury reported. \\Note: This is report 2 of 4.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50184047 ↗
  • MalfunctionAug 7, 2026Decrease in Suction

    Customer reported that their suction broke during scans and laser firing. Procedure was not completed successfully. No patient injury reported. \\Note: This is report 1 of 4.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50184006 ↗
  • MalfunctionAug 7, 2026Manufacturing, Packaging or Shipping Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50177710 ↗
  • MalfunctionJul 30, 2026Decrease in Suction

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50093969 ↗
  • MalfunctionJan 9, 2026Break; Manufacturing, Packaging or Shipping Problem

    THE CUSTOMER REPORTED THAT THEIR LIQUID OPTICS INTERFACE HAD AN EYEPIECE THAT WAS FOUND TO BE LOOSE OUT OF THE BOX. THE PROCEDURE WAS COMPLETED WITH ANOTHER LIQUID OPTICS INTERFACE. NO FURTHER INFORMATION PROVIDED. FOLLOW-UP INFORMATION RECEIVED ON DECEMBER 15, 2025, INDICATED THAT THE STERILE BARRIER WAS OPEN AND THE PRODUCT CAME OUT OF THE PACKAGE, BUT IT WAS NOT LOOSE WITHIN THE PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24030675 ↗
  • InjuryOct 27, 2025Decrease in Suction; Failure to Sense; Decrease in Suction; Failure to Sense

    IT WAS REPORTED THAT A LIQUID OPTICS INTERFACE (LOI) DIRECTION ERROR WAS DISPLAYED AFTER NEGATIVE PRESSURE CONNECTION WHICH OCCURRED BEFORE LASER FIRED AND WAS UNABLE TO CONTINUE THE NEXT STEP. THEN, THERE WAS A LARGER WHITE BUBBLE ABOVE THE MARK POINT VISIBLE ON THE SCREEN, WHICH COULD BE SUCCESSFULLY IDENTIFIED AFTER REPLACING THE LOI. HOWEVER, THE PATIENT'S CONJUNCTIVA WAS RELAXED, THE NEGATIVE PRESSURE WAS REMOVED DURING THE CUTTING PROCESS, THE CUTTING FAILED TO COMPLETE, AND THE ASTIGMATISM RELEASE INCISION FAILED TO BE COMPLETED. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Chemosis; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23392567 ↗
  • MalfunctionOct 20, 2025Decrease in Suction; Decrease in Suction

    THE CUSTOMER REPORTED THAT THERE WAS NO SUCTION ON THE LIQUID OPTICS INTERFACE (LOI) WHICH OCCURRED DURING DOCKING/PROCEDURE ACTIVATION AND DURING LASER TREATMENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23336988 ↗
  • MalfunctionSep 10, 2025Decrease in Suction; Decrease in Suction

    THE CUSTOMER REPORTED RECURRING LOSS OF SUCTION DESPITE MULTIPLE ATTEMPTS AT REDOCKING, CAUSED BY A SMALL EYE FISSURE AND LOW PATIENT COOPERATION. THEY WERE ABLE TO REACH THE TREATMENT STAGE BUT LOST SUCTION DURING CAPSULOTOMY. THE PROCEDURE WAS STOPPED BECAUSE SUCTION COULD NOT BE MAINTAINED. NO PATIENT INJURIES OR ADDITIONAL SURGICAL INTERVENTIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017862 ↗
  • MalfunctionAug 26, 2025Contamination; Contamination

    CUSTOMER REPORTED DEBRIS WAS NOTED ON THE EYEPIECE DURING THE DOCKING/PROCEDURE ACTIVATION, AND SUCTION TO THE EYE WAS ACHIEVED. HOWEVER, THE LASER PROCEDURE WAS COMPLETED SUCCESSFULLY, AS A DIFFERENT KIT WAS USED. THERE WAS NO PATIENT INJURY OR SURGICAL INTERVENTION REQUIRED. NO FURTHER INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22900680 ↗
  • InjuryJun 4, 2025Decrease in Suction; Decrease in Suction

    IT WAS REPORTED THAT THERE WAS A LOSS OF SUCTION DURING A PROCEDURE USING THE CATALYSE-I LASER. THE SUCTION WAS REESTABLISHED AND AT THE END IT WAS FOUND THAT THE ARCUATE INCISION HAD PERFORATED THE CORNEA, MAKING IT NECESSARY TO SUTURE THE INCISION. FOLLOW-UP INFORMATION REVEALED THAT THE PROCEDURE WAS COMPLETED USING A SINGLE LIQUID OPTICAL INTERFACE (LOI) CONE. THE PATIENT IS CURRENTLY WAITING FOR THE REMOVAL OF THE CORNEAL SUTURE. NO ADDITIONAL INFORMATION HAVE BEEN PROVIDED. THIS REPORT PERTAINS THE LIQUID OPTICS INTERFACE(LOI). A SEPARATE REPORT IS BEING SUBMITTED FOR THE CATALYSE SYSTEM .

    Patient problems: Corneal Perforation; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22146110 ↗
  • MalfunctionApr 28, 2025Decrease in Suction; Decrease in Suction

    THIS REPORT SUMMARIZES 28 MALFUNCTION EVENTS. THE EVENT WAS RELATED TO SUCTION LOSS WHILE LASER WAS FIRING. THERE WERE NO PATIENT INJURIES REPORTED ASSOCIATED TO THE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21923372 ↗
  • MalfunctionJan 16, 2025Decrease in Suction; Decrease in Suction

    THIS REPORT SUMMARIZES 25 MALFUNCTION EVENTS. THE EVENTS WERE RELATED TO SUCTION LOSS DURING LASER FIRE. THERE WERE NO PATIENT INJURIES REPORTED ASSOCIATED TO THE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21172942 ↗
  • MalfunctionNov 27, 2024Contamination; Contamination

    CUSTOMER REPORTED DEBRIS ON THEIR LIQUID OPTICS INTERFACE (LOI). THE ISSUE WAS OBSERVED DURING DOCKING/PROCEDURE ACTIVATION. A NEW LOI WAS REQUIRED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20791978 ↗
  • MalfunctionOct 28, 2024Decrease in Suction; Decrease in Suction

    THIS REPORT SUMMARIZES 21 MALFUNCTION EVENTS. THE EVENT WAS RELATED TO SUCTION LOSS WHILE LASER FIRING. THERE WERE NO PATIENT INJURIES REPORTED ASSOCIATED TO THE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20553437 ↗
  • MalfunctionOct 17, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT THE OUTER PACKAGING WAS INTACT, HOWEVER THE INNER PACKAGING OF THE PATIENT INTERFACE WAS NOT AND THE KIT WAS COMPROMISED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20476937 ↗
  • MalfunctionOct 1, 2024Delivered as Unsterile Product; Manufacturing, Packaging or Shipping Problem; Delivered as Unsterile Product; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT THE PACKAGE LIQUID OPTICS INTERFACE (LOI) WAS OPENED AND THERE WAS A BREACH IN STERILITY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20348126 ↗
  • MalfunctionJul 25, 2024Decrease in Suction; Decrease in Suction

    THIS REPORT SUMMARIZES 45 MALFUNCTION EVENTS. THE EVENT WAS RELATED TO SUCTION LOSS WHILE LASER FIRING. THERE WERE NO PATIENT INJURIES REPORTED ASSOCIATED TO THE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19833951 ↗
  • MalfunctionJul 23, 2024Contamination; Contamination

    IT WAS REPORTED THAT THE SURGEON FOUND FOREIGN PARTICLE WHEN DOCKING THE PATIENT EYE. THERE WAS NO DELAY IN TREATMENT OR OTHER INTERVENTIONS PERFORMED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19812394 ↗
  • MalfunctionMay 13, 2024Contamination; Contamination

    IT WAS REPORTED THAT THERE WAS HAZE ON THE CONE OF THE LIQUID OPTICS INTERFACE (LOI). ATTEMPTS WERE MADE TO SEE IF THERE WAS ANY PATIENT CONTACT HOWEVER, NO RESPONSE WAS RECEIVED. THE PROCEDURE WAS COMPLETED BY SWITCHING TO A NEW LOI.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19305937 ↗
  • MalfunctionApr 24, 2024Decrease in Suction; Decrease in Suction

    B5 - THIS REPORT SUMMARIZES 24 MALFUNCTION EVENTS. THE EVENTS WERE RELATED TO SUCTION LOSS DURING LASER FIRE. THERE WERE NO PATIENT INJURIES REPORTED ASSOCIATED TO THE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19177728 ↗
  • MalfunctionNov 20, 2023Decrease in Suction; Mechanical Problem; Decrease in Suction; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS A LOSS OF SUCTION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER INTERFACE. ACCORDING TO THE DOCTOR, THIS UNIT HAD A STEEPER TUBING INCLINATION THAN THE OTHERS. THROUGH FOLLOW-UP, IT WAS CONFIRMED THAT THE SUCTION LOSS OCCURRED DURING LASER FIRING. NO ADDITIONAL INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18177342 ↗
  • MalfunctionNov 6, 2023Decrease in Suction; Device Difficult to Setup or Prepare; Decrease in Suction; Device Difficult to Setup or Prepare

    IT WAS REPORTED THAT THERE WERE DIFFICULTIES IN INSERTING THE LIQUID OPTICAL INTERFACE (LOI) AND THERE WAS SUCTION LOSS. THROUGH FOLLOW-UP, WE LEARNED THAT THE SUCTION LOSS OCCURRED DURING THE LASER FIRING. THE PROCEDURE WAS TEMPORARILY ABORTED AND THEN RESUMED AS A STANDARD CATARACT SURGERY. NO ADDITIONAL INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18076423 ↗
  • MalfunctionJul 28, 2023Decrease in Suction; Decrease in Suction

    THIS REPORT SUMMARIZES 27 MALFUNCTION EVENTS. THE EVENT WAS RELATED TO SUCTION LOSS DURING LASER FIRING. THERE WERE NO PATIENT INJURIES REPORTED ASSOCIATED TO THE EVENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17420806 ↗
  • MalfunctionMay 26, 2023Decrease in Suction; Decrease in Suction

    IT WAS REPORTED THAT A SUCTION LOSS DURING LASER TREATMENT OCCURRED USING THE LIQUID OPTICS INTERFACE (LOI). NO PATIENT INJURY OR SURGICAL INTERVENTION WAS NEEDED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17011742 ↗
  • MalfunctionMay 19, 2023Decrease in Suction; Decrease in Suction

    IT WAS REPORTED THAT THEY WERE UNABLE TO FIXATE AND LOST SUCTION HOWEVER, UNSURE IF THIS OCCURRED PRIOR OR AFTER LASER FIRING. THE LASER PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT INJURY OR SURGICAL INTERVENTION NEEDED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16969761 ↗

FDA clearance and classification

The FDA regulates product code OOE (ophthalmic femtosecond laser) under 21 CFR 886.4390 as a Class II device. 21 CFR 886.4390 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the LIQUID OPTICS INTERFACE 0180-1201

Has the LIQUID OPTICS INTERFACE 0180-1201 been recalled?

No FDA recall record in our data names this exact device identifier (DI 05050474609778), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the LIQUID OPTICS INTERFACE 0180-1201?

The FDA's MAUDE database lists 50 adverse event reports for this device, received between Aug 15, 2019 and Aug 7, 2026. In those reports, the most common specific device problems are “Decrease in Suction” (38), “Contamination” (4) and “Delivered as Unsterile Product” (4); the most common patient problems are “Chemosis” (1) and “Corneal Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the LIQUID OPTICS INTERFACE 0180-1201?

None of the reports are classified as deaths, 2 as injuries and 48 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the LIQUID OPTICS INTERFACE 0180-1201, and is it still sold?

The FDA's device database lists Johnson & Johnson Surgical Vision, Inc. as the company responsible for it (brand LIQUID OPTICS INTERFACE, model 0180-1201). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Ophthalmic femtosecond laser devices

All Ophthalmic femtosecond laser devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page