ORA System® VerifEye™ 8065998328Recalls & reported problems
Alcon Laboratories, Inc. · ORA System with VerifEyeTM+, Aberometer 190.5 · Class 1
All models of ORA System VerifEye (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00380659983283), received between Aug 1, 2022 and May 23, 2025 — classified as 5 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement8
- Incorrect, Inadequate or Imprecise Result or Readings3
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- Insufficient Information5
- Blurred Vision1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 23, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT SYSTEM HAD DIFFICULTY IN MEASURING THE EYES DURING CATARACT SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22080645 ↗InjuryMar 27, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A NON-HEALTHCARE PROFESSIONAL REPORTED SYSTEM PROVIDED INCONSISTENT MEASUREMENTS RESULTED IN AN INTRAOCULAR LENS EXCHANGE IN THE UNKNOWN EYE OF THE PATIENT DURING CATARACT SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 21710560 ↗InjuryMar 27, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A NURSE THAT REPORTED A SYSTEM PROVIDED INCONSISTENT MEASUREMENTS RESULTED IN AN INTRAOCULAR LENS EXCHANGE IN THE UNKNOWN EYE OF THE PATIENT DURING CATARACT SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 21709177 ↗InjuryJan 14, 2025Incorrect Measurement; Incorrect Measurement
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE SYSTEM WITH MEASUREMENTS WAS NOT CORRECT WHICH CAUSED BLURRED VISION AND RESULTED IN EXPLANT IN THE LEFT EYE OF A PATIENT, AFTER CATARACT SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blurred Vision
MAUDE report 21150221 ↗MalfunctionSep 18, 2024Incorrect Measurement; Incorrect Measurement
THE NON HEALTHCARE PROFESSIONAL REPORTED THAT THE SYSTEM WITH MEASUREMENTS WERE OFF DURING SURGERY IN THE UNKNOWN EYE OF A PATIENT.
Patient problems: Insufficient Information
MAUDE report 20244642 ↗InjuryJun 4, 2024Incorrect Measurement; Incorrect Measurement
A HEALTHCARE PROFESSIONAL REPORTED THAT THE INTRAOCULAR LENS WAS EXPLANTED BECAUSE OF THE WRONG LENS IMPLANTED AS PER THE SYSTEM RECOMMENDATION IN THE UNKNOWN EYE OF A PATIENT. THE ADVANCED TECHNOLOGY INTRAOCULAR LENS WAS IMPLANTED. THE SURGEON TOOK THE READING THREE TIMES DURING THE PROCEDURE AND THE SYSTEM SUGGESTED THE WRONG DIOPTER LENS TO IMPLANT. THE SURGEON IMPLANTED THE LENS WHICH WAS RECOMMENDED BY THE SYSTEM. DURING THE POST OPERATIVE VISIT, FOUND THAT THE VISION CORRECTION WAS NOT ACCURATE WITH THE LENS SELECTED. THE SURGEON EXCHANGED THE LENS OF THE PATIENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19458022 ↗InjuryDec 8, 2023Incorrect Measurement; Incorrect Measurement
A NON-HEALTHCARE PROFESSIONAL REPORTED THAT AN EXPLANTED INTRAOCULAR LENS WITH REASON INACCURATE POWER, UNEXPECTED OUTCOME WITH POST-OPERATIVE TREATMENT RESULTS WAS LESS THAN TWO DIOPTERS IN THE RIGHT EYE OF A PATIENT, AFTER SURGERY.
Patient problems: Insufficient Information
MAUDE report 18290825 ↗MalfunctionMay 8, 2023Incorrect Measurement; Incorrect Measurement
A CUSTOMER REPORTED A SYSTEM PROVIDED INCORRECT VALUES AND READINGS. ADDITIONAL INFORMATION HAS BEEN RECEIVED REPORTED SURGEON ADJUSTED ACCORDING TO THEIR PLAN, NO UNEXPECTED OUTCOMES WERE REPORTED. NO ALIGNMENT ISSUE IN ABERROMETER TO CAUSE INCORRECT LENS SUGGESTIONS AND DATABASE RESET WAS PERFORMED.
Patient problems: Insufficient Information
MAUDE report 16886622 ↗MalfunctionMay 4, 2023Incorrect Measurement; Incorrect Measurement
A HEALTH CARE PROFESSIONAL REPORTED THAT INACCURATE INTRAOCULAR LENS CALCULATION (MEASUREMENT CONCERNS ¿ PREOPERATIVE PLAN LENS, CALCULATED AND IMPLANTED LENS WAS DIFFERENT), THE LENS OVER-CORRECTED AND WILL LIKELY LEAD TO AN INTRAOCULAR LENS EXCHANGE, INTERMITTENT CART RESOLUTION ISSUES, FREEZING, TOUCH SCREEN DIFFICULTIES AND SHOWED UPSIDE DOWN ERROR MESSAGE (PHOTONI KEYNOTE), IN AN UNKNOWN EYE OF A PATIENT, DURING CATARACT SURGERY
Patient problems: Insufficient Information
MAUDE report 16869757 ↗MalfunctionDec 26, 2022Incorrect Measurement; Incorrect Measurement
A CUSTOMER REPORTED AN EXPLANTED INTRAOCULAR LENS DUE TO BLURRY VISION, INCORRECT MEASUREMENT (MISCALCULATION) AND SPHERICAL CORRECTIONS IN THE UNKNOWN EYE OF A PATIENT, AFTER SURGERY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 16047748 ↗MalfunctionAug 1, 2022Incorrect Measurement; Incorrect Measurement
A NURSE REPORTED THAT FOLLOWING AN INTRAOCULAR LENS (IOL) IMPLANT PROCEDURE, THE PATIENT EXPERIENCED BLURRED VISION. ADDITIONAL INFORMATION RECEIVED STATED RECOMMENDED A T7 AT THE INITIAL SURGERY DUE TO INCORRECT MEASUREMENTS.
Patient problems: Insufficient Information
MAUDE report 15139505 ↗
FDA clearance and classification
The FDA regulates product code NCF (aberrometer, ophthalmic) under 21 CFR 886.1760 as a Class I device. 21 CFR 886.1760 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ORA System® VerifEye™ 8065998328
Has the ORA System® VerifEye™ 8065998328 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00380659983283), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ORA System® VerifEye™ 8065998328?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Aug 1, 2022 and May 23, 2025. In those reports, the most common specific device problems are “Incorrect Measurement” (8) and “Incorrect, Inadequate or Imprecise Result or Readings” (3); the most common patient problem is “Blurred Vision” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ORA System® VerifEye™ 8065998328?
None of the reports are classified as deaths, 5 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ORA System® VerifEye™ 8065998328 FDA approved?
Devices of this type (product code NCF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Alcon Laboratories, Inc.
Who makes the ORA System® VerifEye™ 8065998328, and is it still sold?
The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand ORA System® VerifEye™, model 8065998328). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Aberrometer, ophthalmic devices
- ORA System® VerifEye™ 8065753093Alcon Laboratories, Inc. · Aberrometer, ophthalmic17 reports
- ORA System® VerifEye™ 8065998329Alcon Laboratories, Inc. · Aberrometer, ophthalmic16 reports
- ORA System® VerifEye™ 8065998330Alcon Laboratories, Inc. · Aberrometer, ophthalmic7 reports
- ORA System® VerifEye™ 8065753001Alcon Laboratories, Inc. · Aberrometer, ophthalmic1 report
- ORA System® VerifEye™ 8065753003Alcon Laboratories, Inc. · Aberrometer, ophthalmic1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page