Skip to content

ORA System® VerifEye™ 8065998330Recalls & reported problems

Alcon Laboratories, Inc. · ORA System with VerifEye + Technology , Abberometer 195.5mm Transient, non-invasive, active therapeutic device, intended for diagnosis and monitoring in ophthalmic surgical procedure. · Class 1

All models of ORA System VerifEye (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00380659983306), received between Nov 12, 2024 and Feb 13, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings5
  • Defective Component1
  • Incorrect Measurement1

Patient problems

  • Insufficient Information5
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Visual Impairment1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2026Defective Component

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT RESIGHT DOVETAIL ON ABERROMETER WAS BROKEN IN THE UNKNOWN EYE IN THE UNKNOWN TIMING.

    Patient problems: Insufficient Information

    MAUDE report 24332588 ↗
  • MalfunctionFeb 6, 2026Incorrect, Inadequate or Imprecise Result or Readings

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE SYSTEM READINGS WERE OFF IN THE UNKNOWN EYE DURING UNKNOWN TIMING OF THE REFRACTIVE SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 24275490 ↗
  • MalfunctionJun 15, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A NON HEALTHCARE PROFESSIONAL REPORTED THAT SYSTEM PROVIDED INCORRECT READINGS AND VALUES IN THE UNKNOWN EYE OF A PATIENT DURING CATARACT SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 22208563 ↗
  • InjuryApr 1, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A HEALTHCARE PROFESSIONAL REPORTED THAT THE PATIENT EXPERIENCED UNEXPECTED OUTCOME DUE TO SYSTEM MEASUREMENTS IN THE LEFT EYE OF A PATIENT, AFTER LASIK SURGERY. THE MEASUREMENTS WAS DONE IN PHAKIC MODE INSTEAD OF NOT APHAKIC MEASUREMENT MODE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21739731 ↗
  • MalfunctionMar 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A HEALTHCARE PROFESSIONAL REPORTED THAT THE SYSTEM READING WAS DIFFERENT THAN PRE OPERATIVE MEASUREMENT VALUES IN THE RIGHT EYE OF A PATIENT, DURING CATARACT SURGERY. THERE WAS NO REPORTED UNEXPECTED OUTCOME AT THIS TIME.

    Patient problems: Insufficient Information

    MAUDE report 21662528 ↗
  • InjuryFeb 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A NON-HEALTHCARE PROFESSIONAL REPORTED THAT THE SURGEON QUESTIONED ABOUT THE SYSTEM READINGS AND THE INTRAOCULAR LENS WAS ROTATED BY ONE HUNDRED TWENTY DEGREES, AS THE RESULT THE PATIENT HAD INDUCED ASTIGMATISM OF THREE DIOPTERS IN THE RIGHT EYE OF A PATIENT, DURING CATARACT SURGERY. THERE ARE TWO RELATED REPORTS FOR THIS PATIENT. THIS REPORT ADDRESSES THE PATIENT'S RIGHT EYE AND ANOTHER MANUFACTURER REPORT WILL BE FILED FOR THE FELLOW EYE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Visual Impairment

    MAUDE report 21432692 ↗
  • MalfunctionNov 12, 2024Incorrect Measurement; Incorrect Measurement

    THE HEALTHCARE PROFESSIONAL REPORTED THAT THE SYSTEM WITH WRONG READING DURING POST TORIC MEASUREMENTS IN THE UNKNOWN EYE OF A PATIENT, DURING SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 20661110 ↗

FDA clearance and classification

The FDA regulates product code NCF (aberrometer, ophthalmic) under 21 CFR 886.1760 as a Class I device. 21 CFR 886.1760 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the ORA System® VerifEye™ 8065998330

Has the ORA System® VerifEye™ 8065998330 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380659983306), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ORA System® VerifEye™ 8065998330?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Nov 12, 2024 and Feb 13, 2026. In those reports, the most common specific device problems are “Incorrect, Inadequate or Imprecise Result or Readings” (5), “Defective Component” (1) and “Incorrect Measurement” (1); the most common patient problem is “Visual Impairment” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ORA System® VerifEye™ 8065998330?

None of the reports are classified as deaths, 2 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ORA System® VerifEye™ 8065998330 FDA approved?

Devices of this type (product code NCF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Alcon Laboratories, Inc.

Who makes the ORA System® VerifEye™ 8065998330, and is it still sold?

The FDA's device database lists Alcon Laboratories, Inc. as the company responsible for it (brand ORA System® VerifEye™, model 8065998330). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Aberrometer, ophthalmic devices

All Aberrometer, ophthalmic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page