Skip to content

ARCHITECT 1L79-35Recalls & reported problems

Abbott GmbH · ARCHITECT Anti-HCV Reagent Kit · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380740001421), received between May 26, 2017 and May 14, 2021 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (P050042); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 4, 2024Open, Classified

    Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).

    Z-1479-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • False Negative Result2
  • Incorrect Or Inadequate Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2021False Negative Result; False Negative Result

    THE CUSTOMER GENERATED FALSELY (B)(6) ARCHITECT (B)(6) RESULTS FOR 2 PATIENTS. THE FOLLOWING INFORMATION WAS PROVIDED: SID (B)(6) INITIAL RESULT (B)(6). THE CUSTOMER REPORTED THIS PATIENT AS (B)(6). SID (B)(6) INITIAL RESULT (B)(6) (GRAYZONE), REPEAT (B)(6) THE CUSTOMER REPORTED THIS PATIENT AS (B)(6). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11824991 ↗
  • MalfunctionApr 6, 2021False Negative Result; False Negative Result

    THE CUSTOMER GENERATED (B)(6) ARCHITECT ANTI-HCV RESULTS FOR 2 PATIENTS. THE FOLLOWING INFORMATION WAS PROVIDED: SID (B)(6), INITIAL RESULT (B)(6). THE CUSTOMER REPORTED THIS PATIENT AS (B)(6). SID (B)(6), INITIAL RESULT (B)(6). THE CUSTOMER REPORTED THIS PATIENT AS (B)(6). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11624047 ↗
  • MalfunctionMay 26, 2017Incorrect Or Inadequate Test Results; Incorrect Or Inadequate Test Results

    THE CUSTOMER OBSERVED (B)(6) RESULTS WHILE USING THE ARCHITECT ANTI-HCV REAGENTS. THE FOLLOWING INFORMATION WAS PROVIDED. (B)(6). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6595746 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P050042. One submission often covers many models, so check the product and lot details in each record.

  • Z-1480-2024 · Mar 4, 2024 · Open, Classified
    Alinity i Anti-HCV Reagent Kit, List Number 08P0521
  • Z-0413-2007 · Jan 2, 2007 · Terminated
    ARCHITECT i System Assay CD-ROM, in vitro diagnostic analyzer, U.S. Version 21.0; for use on the ARCHITECT i System; Abbott Laboratories, Abbott Park, IL 60064; list number 06E58-2

Other ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS devices

All ASSAY,ENZYME LINKED IMMUNOSORBENT,HEPATITIS C VIRUS devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.