Skip to content

Alinity 08P0521Recalls & reported problems

Abbott GmbH · Alinity i Anti-HCV Reagent Kit 200 Tests · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740130657), received between May 14, 2024 and Aug 10, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (P050042); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 4, 2024Open, Classified

    Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).

    Z-1480-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • False Negative Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 10, 2026False Negative Result

    The customer observed false nonreactive Alinity i Anti-HCV results on a patient. The results provided were:Initial=0.33 S/CO (< 1.00 S/CO=nonreactive) /repeated =0.33 and 0.34 S/COPrevious history in APRIL on Cobas Platform=6.88.There was no reported impact to patient management.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50195581 ↗
  • MalfunctionMay 14, 2024False Negative Result; False Negative Result

    THE CUSTOMER REPORTED A FALSE NONREACTIVE ALINITY I ANTI-HCV RESULT ON A PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: (B)(6) 2024 SID (B)(6) = 0.40 S/CO, REPEATED (B)(6) 2024 ON ARCHITECT = 5.59 / 5.46 S/CO. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19311120 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P050042. One submission often covers many models, so check the product and lot details in each record.

  • Z-1479-2024 · Mar 4, 2024 · Open, Classified
    ARCHITECT Anti-HCV Reagent Kit, List Numbers: a) 1L79-25, b) 1L79-35
  • Z-0413-2007 · Jan 2, 2007 · Terminated
    ARCHITECT i System Assay CD-ROM, in vitro diagnostic analyzer, U.S. Version 21.0; for use on the ARCHITECT i System; Abbott Laboratories, Abbott Park, IL 60064; list number 06E58-2

Other Assay, enzyme linked immunosorbent, hepatitis c virus devices

All Assay, enzyme linked immunosorbent, hepatitis c virus devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.