ARCHITECT 1P35-25Recalls & reported problems

Abbott Laboratories · ARCHITECT iValproic Acid Reagent Kit · Class 2

All models of ARCHITECT (51)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740001827), received between Oct 3, 2019 and Oct 3, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Low Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Low Test Results1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 3, 2019Low Test Results; Low Test Results

    THE CUSTOMER REPORTED FALSE DECREASED IVALPROIC ACID RESULTS WHEN PROCESSING ON THE ARCHITECT I2000SR. THE INITIAL RESULT FOR (B)(6) WAS GREATER THAN 150 (209.639 UG/ML). RETEST AFTER DILUTION WAS 77.40 UG/ML. RETEST RESULT OF THE DILUTED SAMPLE WAS 78.29 UG/ML. THE SAMPLE WAS TESTED UNDILUTED AND THE RESULT WAS GREATER THAN 150 AND THIS SAMPLE WAS DILUTED AND GENERATED A RESULT OF 40.98 UG/ML. ADDITIONAL TESTING USING GASCHROMATOGRAPHY GENERATED A RESULT OF 75. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9150292 ↗