VITROS 6801697Recalls & reported problems

Ortho-Clinical Diagnostics, Inc. · VITROS Chemistry Products Calibrator Kit 12 · Class 2

All models of VITROS (71)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10758750006618), received between Aug 23, 2024 and Jun 4, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “High Test Results” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results3
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2025High Test Results; Low Test Results; High Test Results; Low Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS VALPROIC ACID (VALP) REAGENT LOT 2511-34-2390 RESULTS WERE OBTAINED FROM A SINGLE LEVEL OF NON-VITROS BIORAD MULTIQUAL LOT 56750 CONTROL AND A SINGLE LEVEL OF VITROS THERAPEUTIC DRUG MONITORING PERFORMANCE VERIFIER (TDM PV) FLUID ON TWO DIFFERENT VITROS XT7600 SYSTEMS. J76001978 BIORAD L2 VITROS VALP RESULTS OF 538.1 AND 538.1 UMOL/L VS. THE VITROS PEER MEAN OF 448.1 UMOL/L. J76001977 BIORAD L2 VITROS VALP RESULT OF 357.8 UMOL/L VS. THE VITROS PEER MEAN OF 448.1 UMOL/L. VITROS TDM D2151 RESULT OF 398.28 UMOL/L VS. THE EXPECTED RESULT OF 517.0 UMOL/L. THE HIGHER THAN EXPECTED VITROS VALP RESULTS WERE OBTAINED FROM NON-PATIENT QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THIS REPORT IS NUMBER ONE OF TWO 3500A FORMS BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22143497 ↗
  • MalfunctionJan 10, 2025High Test Results; High Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VALPROIC ACID (VALP) RESULTS WERE OBTAINED FROM NON-VITROS BIORAD QUALITY CONTROL (QC) FLUID AS WELL AS FROM AN ADMIXTURE OF PATIENT SAMPLES AND INSTITUTE FOR QUALITY MANAGEMENT IN HEALTHCARE (IQMH) PROFICIENCY SAMPLES USING VITROS CHEMISTRY PRODUCTS VALP REAGENT LOT 2511-33-2028 ON A VITROS XT 7600 INTEGRATED SYSTEM. IQMH/PATIENT SAMPLE ADMIXTURE 1 RESULT OF 404? MOL/L VERSUS THE EXPECTED REPEAT RESULT OF 331? MOL/L IQMH/PATIENT SAMPLE ADMIXTURE 6 RESULT OF 471? MOL/L VERSUS THE EXPECTED REPEAT RESULT OF 383? MOL/L BIORAD QC LEVEL 2 RESULTS 649, 620, 623, 631, 638, AND 639 UMOL/L WHEN COMPARED TO THE BR LEVEL 2 PEER MEAN OF 515.7 UMOL/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE HIGHER-THAN-EXPECTED RESULTS WERE FROM AN ADMIXTURE OF IQMH PROFICIENCY SAMPLES AND PATIENT SAMPLES AND THESE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY. THERE WAS NO REPORTED ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS IN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21124559 ↗
  • MalfunctionAug 23, 2024High Test Results; High Test Results

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VALPROIC ACID (VALP) RESULTS WERE OBTAINED FROM MULTIPLE PATIENT SAMPLES PROCESSED DURING REAGENT LOT TO LOT CORRELATION TESTING USING VITROS CHEMISTRY PRODUCTS VALP REAGENT LOT 2511-34-2390 ON TWO DIFFERENT VITROS XT 7600 INTEGRATED SYSTEMS. VITROS XT 7600 INTEGRATED SYSTEM: S/N (B)(6) (J1) PATIENT 2 SAMPLE RESULT OF 545.10 UMOL/L VS AN EXPECTED RESULT OF 437.40 UMOL/L VITROS XT 7600 INTEGRATED SYSTEM: S/N (B)(6) (J2) PATIENT 4 SAMPLE RESULT OF 312.10 UMOL/L VS AN EXPECTED RESULT OF 237.10 UMOL/L PATIENT 8 SAMPLE RESULT OF 452.90 UMOL/L VS AN EXPECTED RESULT OF 365.80 UMOL/L PATIENT 9 SAMPLE RESULT OF 607.90 UMOL/L VS AN EXPECTED RESULT OF 501.80 UMOL/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE HIGHER-THAN-EXPECTED RESULTS WERE FROM PATIENT SAMPLES THAT WERE PROCESSED DURING REAGENT LOT TO LOT CORRELATION TESTING AND WERE NOT REPORTED OUTSIDE OF THE LABORATORY. NO ERRONEOUS PATIENT SAMPLE RESULTS WERE REPORTED FROM THE LABORATORY. THERE WAS NO REPORTED ALLEGATION OF P

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20053870 ↗

FDA clearance and classification

The FDA regulates product code DLJ (calibrators, drug specific) under 21 CFR 862.3200 as a Class II device. 21 CFR 862.3200 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VITROS 6801697

Has the VITROS 6801697 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10758750006618), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VITROS 6801697?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Aug 23, 2024 and Jun 4, 2025. In those reports, the most common specific device problems are “High Test Results” (3) and “Low Test Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VITROS 6801697?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VITROS 6801697 FDA approved?

Devices of this type (product code DLJ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Ortho-Clinical Diagnostics, Inc.

Who makes the VITROS 6801697, and is it still sold?

The FDA's device database lists Ortho-Clinical Diagnostics, Inc. as the company responsible for it (brand VITROS, model 6801697). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.