ARCHITECT 2K43-20Recalls & reported problems
Abbott Ireland Diagnostics · ARCHITECT STAT Myoglobin Reagent Kit · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380740003296), received between May 10, 2021 and Sep 17, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Product Quality Problem” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Feb 21, 2024Open, Classified
The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.
Z-1476-2024 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Product Quality Problem6
- Failure to Calibrate3
- Incorrect, Inadequate or Imprecise Result or Readings3
- High Test Results1
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2026Low Test Results
The customer observed falsely depressed ARCHITECT Myoglobin results for four samples. The following data was provided:Sample 1 initial ARCHITECT Myoglobin result 12 ng/mL, retest 40 ng/mL,Sample 2 initial ARCHITECT Myoglobin result 9.2 ng/mL, retest 31.4 ng/mL,Sample 3 initial ARCHITECT Myoglobin result 91 ng/mL, retest 412.1 ng/mL,Sample 4 initial ARCHITECT Myoglobin result 70 ng/mL, retest 336.6 ng/mL. No impact to patient management was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50803312 ↗MalfunctionJun 21, 2024Product Quality Problem; Product Quality Problem
THE CUSTOMER OBSERVED THAT QUALITY CONTROL WERE OUT OF RANGE HIGH FOR ALL THREE LEVELS WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23 ON THE ARCHITECT I1000 PROCESSING MODULE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19579989 ↗MalfunctionJun 20, 2024Product Quality Problem; Product Quality Problem
THE CUSTOMER OBSERVED THAT QUALITY CONTROL WERE OUT OF RANGE HIGH FOR ALL THREE LEVELS WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23 ON THE ARCHITECT I1000 PROCESSING MODULE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19579218 ↗MalfunctionJun 4, 2024Product Quality Problem; Failure to Calibrate; Product Quality Problem; Failure to Calibrate
THE CUSTOMER OBSERVED THAT MICROPARTICLES IN THE MICROPARTICLES BOTTLE APPEARED LIGHTER THAN USUAL AND QUALITY CONTROL RESULTS ISSUES WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23 ON THE ARCHITECT I2000 PROCESSING MODULE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19461962 ↗MalfunctionMar 18, 2024Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER OBSERVED THAT BIO-RAD QUALITY CONTROLS WAS OUT OF RANGE HIGH AND MICROPARTICLES IN THE MICROPARTICLES BOTTLE APPEARED DIFFERENT THAN USUAL WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST NUMBER 02K43-20, LOT NUMBER 60104UN23. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18925017 ↗MalfunctionMar 14, 2024Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate; Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate
THE CUSTOMER OBSERVED THAT QUALITY CONTROLS (QC) WERE OUT OF RANGE LOW AFTER USE OF ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23. THE CUSTOMER PROVIDED FOLLOWING DATA FOR THE QC. QC1 55.9 79.95 WITH THE RANGE 63.96-95.94 NG/ML QC2 298.3 449.89 WITH THE RANGE 359.93-539.85 NG/ML SAMPLE COMPARISON RECOVERED WITHIN RANGE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18905298 ↗MalfunctionMar 14, 2024Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate; Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate
THE CUSTOMER OBSERVED THAT QUALITY CONTROLS (QC) WERE OUT OF RANGE LOW AFTER USE OF ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23. THE CUSTOMER PROVIDED FOLLOWING DATA FOR THE QC. LOW LEVEL: 48.9 NG/ML, 47.8 NG/ML WITH THE RANGE 60.2-83.0 NG/ML AND HIGH LEVEL WAS 257 NG/ML WITH THE RANGE 339-459 NG/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18904133 ↗MalfunctionMay 10, 2021High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT STAT MYOGLOBIN RESULTS ON ONE PATIENT. THE RESULTS PROVIDED WERE: SID (B)(6) INITIAL=1003.7 NG/ML / REPEATED DUE TO QUESTIONED BY PHYSICIAN =51 NG/ML THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11801795 ↗
Other MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL devices
- PATHFAST Myo-II N/APhc Corporation · Myoglobin, Antigen, Antiserum, Control1 recall0 reports
- Triage 97300EUQuidel Cardiovascular Inc. · MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL8 reports
- VITROS 6801042Ortho Clinical Diagnostics · MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL7 reports
- Alere Triage® 97000HSAlere San Diego, Inc. · Myoglobin, Antigen, Antiserum, Control6 reports
- Elecsys Myoglobin 07027583190Roche Diagnostics GmbH · MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL6 reports
- Tinaquant Myoglobin Gen2 4580010190Roche Diagnostics GmbH · MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL1 report
- Alinity 04V3720Abbott Ireland Diagnostics · MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL1 report
- Tina-quant Myoglobin Gen.2 08252629190Roche Diagnostics GmbH · MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.