ARCHITECT 2K43-20Recalls & reported problems

Abbott Ireland Diagnostics · ARCHITECT STAT Myoglobin Reagent Kit · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380740003296), received between May 10, 2021 and Sep 17, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Product Quality Problem” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 21, 2024Open, Classified

    The reason for the recall is the failure of calibration and quality controls due to a manufacturing issue of microparticles not meeting labeled claim. This may lead to incorrect myoglobin results and delayed diagnosis of myocardial infarction.

    Z-1476-2024 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Product Quality Problem6
  • Failure to Calibrate3
  • Incorrect, Inadequate or Imprecise Result or Readings3
  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2026Low Test Results

    The customer observed falsely depressed ARCHITECT Myoglobin results for four samples. The following data was provided:Sample 1 initial ARCHITECT Myoglobin result 12 ng/mL, retest 40 ng/mL,Sample 2 initial ARCHITECT Myoglobin result 9.2 ng/mL, retest 31.4 ng/mL,Sample 3 initial ARCHITECT Myoglobin result 91 ng/mL, retest 412.1 ng/mL,Sample 4 initial ARCHITECT Myoglobin result 70 ng/mL, retest 336.6 ng/mL. No impact to patient management was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50803312 ↗
  • MalfunctionJun 21, 2024Product Quality Problem; Product Quality Problem

    THE CUSTOMER OBSERVED THAT QUALITY CONTROL WERE OUT OF RANGE HIGH FOR ALL THREE LEVELS WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23 ON THE ARCHITECT I1000 PROCESSING MODULE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19579989 ↗
  • MalfunctionJun 20, 2024Product Quality Problem; Product Quality Problem

    THE CUSTOMER OBSERVED THAT QUALITY CONTROL WERE OUT OF RANGE HIGH FOR ALL THREE LEVELS WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23 ON THE ARCHITECT I1000 PROCESSING MODULE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19579218 ↗
  • MalfunctionJun 4, 2024Product Quality Problem; Failure to Calibrate; Product Quality Problem; Failure to Calibrate

    THE CUSTOMER OBSERVED THAT MICROPARTICLES IN THE MICROPARTICLES BOTTLE APPEARED LIGHTER THAN USUAL AND QUALITY CONTROL RESULTS ISSUES WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23 ON THE ARCHITECT I2000 PROCESSING MODULE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19461962 ↗
  • MalfunctionMar 18, 2024Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER OBSERVED THAT BIO-RAD QUALITY CONTROLS WAS OUT OF RANGE HIGH AND MICROPARTICLES IN THE MICROPARTICLES BOTTLE APPEARED DIFFERENT THAN USUAL WHILE USING ARCHITECT STAT MYOGLOBIN REAGENT LIST NUMBER 02K43-20, LOT NUMBER 60104UN23. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18925017 ↗
  • MalfunctionMar 14, 2024Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate; Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate

    THE CUSTOMER OBSERVED THAT QUALITY CONTROLS (QC) WERE OUT OF RANGE LOW AFTER USE OF ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23. THE CUSTOMER PROVIDED FOLLOWING DATA FOR THE QC. QC1 55.9 79.95 WITH THE RANGE 63.96-95.94 NG/ML QC2 298.3 449.89 WITH THE RANGE 359.93-539.85 NG/ML SAMPLE COMPARISON RECOVERED WITHIN RANGE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18905298 ↗
  • MalfunctionMar 14, 2024Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate; Product Quality Problem; Incorrect, Inadequate or Imprecise Result or Readings; Failure to Calibrate

    THE CUSTOMER OBSERVED THAT QUALITY CONTROLS (QC) WERE OUT OF RANGE LOW AFTER USE OF ARCHITECT STAT MYOGLOBIN REAGENT LIST 02K43-20, LOT 60104UN23. THE CUSTOMER PROVIDED FOLLOWING DATA FOR THE QC. LOW LEVEL: 48.9 NG/ML, 47.8 NG/ML WITH THE RANGE 60.2-83.0 NG/ML AND HIGH LEVEL WAS 257 NG/ML WITH THE RANGE 339-459 NG/ML. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18904133 ↗
  • MalfunctionMay 10, 2021High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT STAT MYOGLOBIN RESULTS ON ONE PATIENT. THE RESULTS PROVIDED WERE: SID (B)(6) INITIAL=1003.7 NG/ML / REPEATED DUE TO QUESTIONED BY PHYSICIAN =51 NG/ML THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11801795 ↗

Other MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL devices

All MYOGLOBIN, ANTIGEN, ANTISERUM, CONTROL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.