Skip to content

PATHFAST Myo-II N/ARecalls & reported problems

Phc Corporation · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Apr 30, 2018Class IITerminated

    A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.

    Z-3185-2018 · Root cause: Process control · Recall details · FDA record ↗

Other Myoglobin, Antigen, Antiserum, Control devices

All Myoglobin, Antigen, Antiserum, Control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page