ARCHITECT 3L80-21Recalls & reported problems
Abbott Laboratories · Carbon Dioxide · Class 2
All models of ARCHITECT (51)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00380740005955), received between Jun 24, 2018 and Feb 10, 2023 — classified as 14 malfunctions.
The most frequently reported specific device problem is “High Test Results” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- High Test Results11
- Low Test Results3
Patient problems
- No Known Impact Or Consequence To Patient9
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 10, 2023High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CO2 RESULTS GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR MULTIPLE SAMPLES. (NORMAL RANGE: 22-29 MEQ/L). THE CUSTOMER REPORTS THE RESULTS OF PATIENTS BEGIN TO INCREASE IN VALUE DURING THE DAY. RESULTS NEAR 35 OR 36 MEQ/L, AFTER CALIBRATING GENERATE NORMAL RESULTS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16349807 ↗MalfunctionAug 17, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED BICARBONATE (CO2) RESULTS FOR THE PAST MONTH ON PATIENT SAMPLES ON THE ARCHITECT ANALYZER. THE CUSTOMER PROVIDED TWO EXAMPLES OF RESULTS: ON (B)(6) 2020 (B)(6) INITIAL = 40MMOL/L / RETEST TWICE =20MMOL/L (NORMAL RANGE 22-29MMOL/L). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10417521 ↗MalfunctionAug 4, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED FALSE ELEVATED ARCHITECT CO2 ASSAY RESULTS FOR THREE PATIENTS. THE CUSTOMER PROVIDED: (B)(6) INITIAL = >50 MMOL/L, REPEAT = 27 MMOL/L; (B)(6) INITIAL = >50 MMOL/L, REPEAT = 24 MMOL/L; (B)(6) INITIAL = >50 MMOL/L, REPEAT 20.3, 20.4 MMOL/L (RANGE: 23.0 TO 31 MMOL/L). THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10365161 ↗MalfunctionJul 20, 2020Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DECREASED CO2 RESULTS ON THE ARCHITECT C4000 ANALYZER FOR TWO PREGNANT PATIENTS. ON (B)(6) 2020 SID (B)(6) RESULT 14 MEQ/L. ON (B)(6) 2020 SID (B)(6) RESULT 15 MEQ/L. THE CUSTOMER'S NORMAL RANGE IS 22 - 32 MEQ/L. VALUES LESS THAN 16 MEQ/L ARE CONSIDERED CRITICAL. BOTH PATIENTS HAD NORMAL ARTERIAL BLOOD GAS CO2 VALUES (21 - 23 MEQ/L). NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10296192 ↗MalfunctionJul 15, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED BICARBONATE (CO2) RESULTS FOR THE PAST MONTH ON PATIENT SAMPLES ON THE ARCHITECT ANALYZER. THE CUSTOMER PROVIDED TWO EXAMPLES OF RESULTS: ON (B)(6) 2020 (B)(6) INITIAL = 40MMOL/L / RETEST TWICE = 20MMOL/L (NORMAL RANGE 22-29MMOL/L). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10276595 ↗MalfunctionJan 28, 2020High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED CO2 RESULTS FOR 7 PATIENTS GENERATED ON THE ARCHITECT ANALYZER. THE RESULTS PROVIDED WERE: ON (B)(6) 2020 SID (B)(6) = 29 / 32 / 33; SID (B)(6) = 23 / 36MEQ/L (NORMAL RANGE 22-29MEQ/L). ON (B)(6) 2020 ADDITIONAL RESULTS PROVIDED: SID (B)(6) = 33 /22; SID (B)(6) = 30 /25; SID (B)(6) = 33 / 22; SID (B)(6) = 33 /25; AND SID (B)(6) = 33 / 22. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9638791 ↗MalfunctionFeb 5, 2019Low Test Results; Low Test Results
THE CUSTOMER INQUIRED AS TO WHETHER AN ELEVATED TRIGLYCERIDE RESULT (3642 MG/DL) COULD CAUSE A FALSELY DEPRESSED CO2 RESULT OF 11 MMOL/L FOR A PATIENT SAMPLE TESTED ON THE ARCHITECT C16000 SYSTEM. THE SAMPLE TESTED AT 26 MMOL/L USING AN ALTERNATE CO2 METHOD. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8307829 ↗MalfunctionDec 21, 2018High Test Results; High Test Results
THE CUSTOMER REPORTED MULTIPLE ARCHITECT CARBON DIOXIDE RESULTS THAT WERE EITHER FALSELY ELEVATED OR REPEATED LOWER WHEN RERUN. THE CUSTOMER PROVIDED THE FOLLOWING DISCREPANT RESULTS: INITIAL 36, REPEAT 20 DISCREPANT; INITIAL 25, REPEAT 18 DISCREPANT; INITIAL 27, REPEAT 18 DISCREPANT; INITIAL 36, REPEAT 23 DISCREPANT; INITIAL 41, REPEAT 20 DISCREPANT; INITIAL 32, REPEAT 27 DISCREPANT; INITIAL 25, REPEAT 21 DISCREPANT; INITIAL 31, REPEAT 27 DISCREPANT; INITIAL 27, REPEAT 21 DISCREPANT; INITIAL 24, REPEAT 19 DISCREPANT. THE CUSTOMER USES A NORMAL RANGE OF 22-29 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8191268 ↗MalfunctionNov 28, 2018Low Test Results; Low Test Results
THE CUSTOMER REPORTED OF FALSELY DEPRESSED ARCHITECT CARBON DIOXIDE RESULTS FOR MULTIPLE SAMPLES. THE CUSTOMER INDICATED THE INITIAL RESULTS WERE APPROXIMATELY 18 MMOL/L AND RETESTED ON A NEW LOT WITH A RESULTS OF APPROXIMATELY 22 MMOL/L. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8112107 ↗MalfunctionSep 27, 2018High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CO2 RESULTS WHILE USING THE CLINICAL CHEMISTRY CO2 ASSAY. THE CUSTOMER PROVIDED THE FOLLOWING DATA FOR 4 PATIENTS (NORMAL RANGE IS 23-31 MEQ/L): SID (B)(6): INITIAL 32 MEQ/L, RETESTED USING NEW REAGENT 27 MEQ/L. SID (B)(6): INITIAL 35 MEQ/L, RETESTED USING NEW REAGENT 29 MEQ/L. SID (B)(6): INITIAL 32 MEQ/L, RETESTED USING NEW REAGENT 27 MEQ/L. SID (B)(6): INITIAL 31 MEQ/L, RETESTED USING NEW REAGENT 27 MEQ/L . NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7914125 ↗MalfunctionJul 9, 2018High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED CO2 RESULTS ON MULTIPLE PATIENTS. THE RESULTS PROVIDED WERE FOR 5 SAMPLES: SID (B)(6) = 33.27 / 24.67; SID (B)(6) = 31.89 / 24.26; SID (B)(6) = 31.75 / 23.68; SID (B)(6) = 32.47 / 24.67; SID (B)(6) = 32.87 / 24.52. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7671710 ↗MalfunctionJun 28, 2018High Test Results; High Test Results
THE ACCOUNT OBSERVED ELEVATED ARCHITECT CARBON DIOXIDE RESULTS RANGING 33 TO 35 MMOL/L ON 5 PATIENT SAMPLES. THE SAMPLES WERE REPEATED AFTER RECALIBRATION WITH PATIENT RESULTS IN THE ACCOUNT NORMAL RANGE OF 20 TO 31 MMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED. NO SPECIFIC PATIENT INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7649241 ↗MalfunctionJun 25, 2018High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED CO2 RESULTS ON MULTIPLE PATIENTS. THE RESULTS PROVIDED WERE FOR 5 SAMPLES: SID (B)(6) = 33.27 / 24.67; (B)(6) = 31.89 / 24.26; SID (B)(6) = 31.75 / 23.68; SID (B)(6) = 32.47 / 24.67; SID (B)(6) = 32.87 / 24.52. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7634405 ↗MalfunctionJun 24, 2018High Test Results; High Test Results
THE ACCOUNT GENERATED ELEVATED ARCHITECT CARBON DIOXIDE OF 30 MMOL/L ON SID (B)(6) THAT REPEATED 27 MMOL/L AT A REFERENCE LAB. HISTORICAL CARBON DIOXIDE PATIENT RESULTS WERE 28 MMOL/L ((B)(6) 2017), 24 MMOL/L ((B)(6) 2017), 25 MMOL/L ((B)(6) 2017), 28 MMOL/L ((B)(6) 2016). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7632770 ↗
FDA clearance and classification
The FDA regulates product code KHS (enzymatic, carbon-dioxide) under 21 CFR 862.1160 as a Class II device. 21 CFR 862.1160 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ARCHITECT 3L80-21
Has the ARCHITECT 3L80-21 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00380740005955), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ARCHITECT 3L80-21?
The FDA's MAUDE database lists 14 adverse event reports for this device, received between Jun 24, 2018 and Feb 10, 2023. In those reports, the most common specific device problems are “High Test Results” (11) and “Low Test Results” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARCHITECT 3L80-21?
None of the reports are classified as deaths, 0 as injuries and 14 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ARCHITECT 3L80-21, and is it still sold?
The FDA's device database lists Abbott Laboratories as the company responsible for it (brand ARCHITECT, model 3L80-21). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ENZYMATIC, CARBON-DIOXIDE devices
- AU OSR6237Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls1 report
- AU OSR6637Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls0 reports
- AU OSR6137Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls0 reports
- Alinity 07P7220Abbott GmbH · ENZYMATIC, CARBON-DIOXIDE89 reports
- VITROS 8262396Ortho-Clinical Diagnostics, Inc. · ENZYMATIC, CARBON-DIOXIDE1 recall0 reports
- ARCHITECT 3L80-22Abbott GmbH · ENZYMATIC, CARBON-DIOXIDE31 reports
- Atellica CH CO2_c 11097521Siemens Healthcare Diagnostics Inc. · ENZYMATIC, CARBON-DIOXIDE21 reports
- Dimension Vista® Flex® reagent cartridge CO2 10445163Siemens Healthcare Diagnostics Inc. · ENZYMATIC, CARBON-DIOXIDE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page