Skip to content

AU OSR6237Recalls & reported problems

Beckman Coulter, Inc. · Bicarbonate · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15099590020187), received between Mar 18, 2026 and Mar 18, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “High Test Results” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

2 other recalls cover devices cleared under the same FDA submission (K131546); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 14, 2025Open, Classified

    Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

    Z-0997-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

  2. Apr 5, 2023Open, Classified

    There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.

    Z-1606-2023 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • High Test Results1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 18, 2026High Test Results

    THIS EVENT WAS INITIALLY REPORTED UNDER MDR 9612296-2025-00950. FOLLOWING A REVIEW COMPLETED ON FEBRUARY 20, 2026, IT WAS DETERMINED THAT THIS EVENT SHOULD BE REPORTED UNDER THE REAGENT, RATHER THAN THE ANALYZER. ON (B)(6) 2025, IT WAS REPORTED BY THE CUSTOMER THAT THERE WERE ERRONEOUSLY HIGH CARBON DIOXIDE (CO2) PATIENT RESULTS WHILE USING THE DXC 700 AU CHEMISTRY ANALYZER. THE CUSTOMER INDICATED THAT THE CO2 RESULTS FROM THE ARTERIAL BLOOD GAS WERE "SIGNIFICANTLY DIFFERENT" THAN THE TOTAL CO2 MEASURED ON THE BECKMAN COULTER DXC 700 AU ANALYZER. THE CUSTOMER ALSO INDICATED THAT THE PATIENT WAS "VERY SICK" WITH MULTIPLE MORBIDITIES. PER THE CUSTOMER, THE PATIENT WAS SUSPECTED TO HAVE HIGH LACTATE DEHYDROGENASE (LDH) AND LACTATE (LAC) RESULTS. THE CUSTOMER REPORTED A POSSIBLE DELAY IN CHANGING THE DOSAGE OF CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT) DUE TO THE ERRONEOUSLY HIGH CO2 RESULTS. THE CUSTOMER REPORTED THE PATIENT PASSED AWAY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24635522 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K131546. One submission often covers many models, so check the product and lot details in each record.

  • Z-1607-2023 · Apr 5, 2023 · Open, Classified
    Beckman Coulter Bicarbonate, REF: OSR6237, 4x50 mL
  • Z-1608-2023 · Apr 5, 2023 · Open, Classified
    Beckman Coulter Bicarbonate, REF: OS6637, 4x173 mL

Other Enzymatic, Carbon-Dioxide devices

All Enzymatic, Carbon-Dioxide devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.