Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380740011123), received between Dec 13, 2017 and Oct 4, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result2
  • Low Test Results1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 4, 2023False Negative Result; False Negative Result

    THE CUSTOMER OBSERVED A FALSE NONREACTIVE ARCHITECT CORE RESULT FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: 24AUG2023: ANTI-HBCORE TOTAL RESULT = 0.08 (NONREACTIVE). ANTI-HBCORE IGM RESULT = 11, REPEAT = 12 (REACTIVE). THE PHYSICIAN IS QUESTIONING THE RESULT SINCE THE ANTI-HBCORE TOTAL RESULT IS LESS THAN THE ANTI-HBCORE M RESULT. THE SAMPLE WAS REPEATED ON 13SEP2023 AND GENERATED THE SAME RESULTS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17870991 ↗
  • MalfunctionMay 14, 2018False Negative Result; False Negative Result

    THE CUSTOMER OBSERVED A FALSE NEGATIVE HBCORE RESULT ON THE ARCHITECT I2000SR ANALYZER. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL 0.931, REPEATS 4.713, 4.927 S/CO. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7512743 ↗
  • MalfunctionDec 13, 2017Low Test Results; Low Test Results

    THE CUSTOMER IS REQUESTING AN EXPLANATION FOR WHY THE ARCHITECT TOTAL ANTI-HBCORE IS NONREACTIVE ON A PATIENT, BUT THE ARCHITECT ANTI-HBCORE IGM IS REACTIVE. THE RESULTS PROVIDED WERE: ANTI-HBCORE-M =1.62S/CO (>/=1.21S/CO = REACTIVE); AUSAB = >1000. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED.

    Patient problems: No Known Impact Or Consequence To Patient