Alinity 07P6620Recalls & reported problems
Abbott Ireland · Alinity i Prolactin Reagent Kit 200 Tests · Class 1
All models of Alinity (22)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Abbott Alinity i reagent kit is used to measure prolactin levels in blood samples. The 26 reports all describe prolactin results reading far higher than the true value, and repeating the test or checking it on another lab's machine gave normal numbers. None of the reports mention deaths or injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 00380740121570), received between Mar 29, 2021 and Mar 9, 2026 — classified as 25 malfunctions.
The most frequently reported specific device problem is “High Test Results” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results25
Patient problems
- No Clinical Signs, Symptoms or Conditions25
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR A 62-YEAR-OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2025, SID: (B)(6): INITIAL RESULT (LOT: 72165UD00, (B)(6) = 1980.092 NG/ML, REPEAT WITH 1:10 DILUTION (B)(6) = 3083.830 NG/ML, REPEAT WITH NO DILUTION (B)(6) = 3230.905 NG/ML. ON (B)(6) 2025, SID: (B)(6): INITIAL RESULT (LOT: 75050UD00, (B)(6) = 2225.910 NG/ML, REPEAT (B)(6) = 3468.409 NG/ML, REPEAT WITH 1:10 DILUTION (B)(6) = 3229.661 NG/ML. ON (B)(6) 2026, SID: (B)(6): INITIAL RESULT = 3092.7 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24547904 ↗MalfunctionMar 9, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR A 62-YEAR-OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2025, SID (B)(6): INITIAL RESULT (LOT 72165UD00, (B)(6) = 1980.092 NG/ML, REPEAT WITH 1:10 DILUTION (B)(6) = 3083.830 NG/ML, REPEAT WITH NO DILUTION (B)(6) = 3230.905 NG/ML. (B)(6) 2025, SID (B)(6): INITIAL RESULT (LOT 75050UD00, (B)(6) = 2225.910 NG/ML, REPEAT (B)(6) = 3468.409 NG/ML, REPEAT WITH 1:10 DILUTION (B)(6) = 3229.661 NG/ML. (B)(6) 2026, SID (B)(6): INITIAL RESULT = 3092.7 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24547899 ↗MalfunctionMar 9, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR A 62-YEAR-OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2025, SID (B)(6): INITIAL RESULT (LOT 72165UD00, (B)(6) = 1980.092 NG/ML, REPEAT WITH 1:10 DILUTION (B)(6) = 3083.830 NG/ML, REPEAT WITH NO DILUTION (B)(6) = 3230.905 NG/ML. (B)(6) 2025, SID (B)(6): INITIAL RESULT (LOT 75050UD00, (B)(6) = 2225.910 NG/ML, REPEAT (B)(6) = 3468.409 NG/ML, REPEAT WITH 1:10 DILUTION (B)(6) = 3229.661 NG/ML. (B)(6) 2026, SID (B)(6): INITIAL RESULT = 3092.7 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24547898 ↗MalfunctionFeb 5, 2026High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULT A 53YRS OLD FEMALE PATIENT. THE RESULTS PROVIDED WERE: ON (B)(6) 2026, SID (B)(6), INITIAL= 3168.86 MIU/L (REFERENCE RANGE= 108.78 TO 557.13 MIU/L THE SPECIMEN IS SENT TO ANOTHER LABORATORY FOR POLYETHYLENE GLYCOL (PEG) PRECIPITATION. ON ROCHE ECLIA=536 UUI/ML (REFERENCE RANGE= 127 TO 637 UUI/ML), NO POLYETHYLENE GLYCOL (PEG) PRECIPITATION METHOD PERFORMED DUE TO NORMAL RESULTS ON ROCHE PLATFORM. SAME SPECIMEN REPEATED ON ALINITY ON (B)(6) 2026= 3232.61 MIU/L, 2968.92 MIU/L, 3299.16 MIU/L THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24269209 ↗MalfunctionNov 18, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I PROLACTIN RESULT ON A PATIENT. RESULTS PROVIDED: ON (B)(6) 2025, SID (B)(6) (62-YR OLD MALE) = >27.09, HISTORY OF PROLACTIN OF 5.80 NG/ML; ANOTHER LABORATORY CHEMILUMINESCENCE PLATFORM = 9.6 NG/ML, NORMAL RANGE MALE: 3.46-19.40 NG/ML, NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23579746 ↗MalfunctionNov 18, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I PROLACTIN RESULT ON A PATIENT. RESULTS PROVIDED: (B)(6) 2025: SID (B)(6) (41-YR OLD FEMALE) = 27.0 NG/ML, DXI RESULT = 20.89 NG/ML, NORMAL RANGE FEMALE: 5.18-26.53 NG/ML , NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23579709 ↗MalfunctionSep 25, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS GENERATED FOR A 9-YEAR-OLD FEMALE PATIENT WHO IS NOT PREGNANT. THE FOLLOWING DATA WAS PROVIDED: SAMPLE ID 10276218 RESULT = 110 NG/ML, RESULT ON SIEMENS PLATFORM = 24 NG/ML. SAMPLE WAS INITIALLY RUN AT ANOTHER LABORATORY ON ANOTHER ALINITY. SAMPLE ID (B)(6). RESULT = 110 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23144083 ↗MalfunctionSep 24, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR 9 YEAR OLD FEMALE PATIENT WHO IS NOT PREGNANT. THE SAMPLE WAS REPEATED USING THE SIEMENS METHOD WITH LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) ALINITY RESULT = 110 NG/ML. SENT TO ANOTHER LAB ALINITY RESULT FOR SID (B)(6) = 110 NG/ML. SIEMENS RESULT = 24 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23133535 ↗MalfunctionSep 4, 2025High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN. THE CUSTOMER INITIALLY REPORTED ELEVATED QC AND THEN REPORTED AN ELEVATED PATIENT RESULT. THE CUSTOMER PROVIDED THE FOLLOWING DATA: THE PROLACTIN RESULT WAS 68 NG/ML., AND WHEN TESTED AT ANOTHER FACILITY WITH THE ROCHE PLATFORM, THE RESULT WAS 27 NG/ML. THERE WAS NO REPORTED IMPACT ON PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22968371 ↗MalfunctionJun 2, 2025High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR ONE PATIENT. THE CUSTOMER OBSERVED A DIFFERENCE BETWEEN THE PLASMA AND SERUM SAMPLE RESULTS. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6) - HEPARINIZED PLASMA TUBE = 42.63 NG/ML SID (B)(6) ¿ SERUM TUBE = 26.19 NG/ML PATIENT WAS DRAWN AGAIN WITH NEW SAMPLE SID (B)(6) - SERUM = 23.75 NG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22124807 ↗MalfunctionJan 2, 2024High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR ONE 23-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2023: PROLACTIN TESTED WITH THE CUSTOMER USING AN ARCHITECT AT (B)(6) WITH A RESULT OF 18.25 NG/ML (NORMAL). ON (B)(6) 2023, SHE RETURNED TO DO ALL HER HORMONAL TESTS AGAIN AND THIS TIME THE PROLACTIN WAS PROCESSED FOR SID (B)(6) ON THE ALINITY AT (B)(6) WITH A RESULT OF 78.26 NG/ML (ELEVATED). THIS SAMPLE WAS REPEATED AND GENERATED A RESULT OF 73.79 NG/ML (ELEVATED). THE PATIENT FOLLOWS UP WITH HER DOCTOR AND HE RECOMMENDS GOING TO ANOTHER LABORATORY AND OBTAINED A NORMAL RESULT. THE PATIENT GOES 3 DAYS LATER TO YET ANOTHER LAB TO HAVE A THIRD SAMPLE TESTED AND THE RESULT OBTAINED IS NORMAL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18428444 ↗MalfunctionDec 4, 2023High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR MULTIPLE PATIENTS. THE SAMPLES WERE SENT TO ANOTHER LAB WITH CHEMILUMINESCENCE METHOD, AND THE RESULTS WERE LOWER. THE FOLLOWING DATA WAS PROVIDED: PATIENT 1: FEMALE ALINITY I PROLACTIN RESULT = 33.86 NG/ML; ANOTHER LAB¿S RESULT = 22.7 NG/ML. PATIENT 2: FEMALE ALINITY I PROLACTIN RESULT = 30.39 NG/ML; ANOTHER LAB¿S RESULT = 18.7 NG/ML. PATIENT 3: FEMALE ALINITY I PROLACTIN RESULT = 13.61 NG/ML; ANOTHER LAB¿S RESULT = 9.4 NG/ML. PATIENT 4: FEMALE ALINITY I PROLACTIN RESULT = 8.13 NG/ML; ANOTHER LAB¿S RESULT = 6.0 NG/ML. PATIENT 5: FEMALE ALINITY I PROLACTIN RESULT = 11.28 NG/ML; ANOTHER LAB¿S RESULT = 7.8 NG/ML. PATIENT 6: FEMALE ALINITY I PROLACTIN RESULT = 34.02 NG/ML; ANOTHER LAB¿S RESULT = 23.0 NG/ML. PATIENT 7: FEMALE ALINITY I PROLACTIN RESULT = 30.39 NG/ML; ANOTHER LAB¿S RESULT = 18.7 NG/ML. PATIENT 8: MALE ALINITY I PROLACTIN RESULT = 30.7 NG/ML; ANOTHER LAB¿S RESULT = 7.0 NG/ML NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18260263 ↗MalfunctionNov 13, 2023High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN AND PROVIDED THE FOLLOWING EXAMPLE DATA: FEMALE PATIENT DATA: CUSTOMER NORMAL VALUES ARE: 5.18 - 28 NG/ML. PATIENT 1 INITIAL RESULT WAS 28.95 NG/ML AND RETEST RESULT WAS 19.7 NG/ML. PATIENT 2 INITIAL RESULT WAS 65.15 NG/ML AND RETEST RESULT WAS 28.3 NG/ML. PATIENT 3 INITIAL RESULT WAS 26.72 NG/ML AND RETEST RESULT WAS 19.1 NG/ML. MALE DATA PROVIDED: CUSTOMER NORMAL VALUES ARE: 3.46 - 19.40 NG/ML. THE CUSTOMER TESTED 3 MALE SAMPLES THAT WERE FROM REPORTEDLY HEALTHY INDIVIDUALS AND OBTAINED THE FOLLOWING RESULTS: 23.9, 9.67, & 19.62. NO OTHER INFO AVAILABLE FOR THESE PATIENTS. THE CUSTOMER REPORTED LOT 52220UD00 AND LOT 53599UD00 WERE ON-BOARD THE ANALYZER WHEN THE DISCREPANT RESULTS WERE OBTAINED. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18120856 ↗MalfunctionNov 13, 2023High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN AND PROVIDED THE FOLLOWING EXAMPLE DATA: FEMALE PATIENT DATA: CUSTOMER NORMAL VALUES ARE: 5.18 - 28 NG/ML. PATIENT 1 INITIAL RESULT WAS 28.95 NG/ML AND RETEST RESULT WAS 19.7 NG/ML. PATIENT 2 INITIAL RESULT WAS 65.15 NG/ML AND RETEST RESULT WAS 28.3 NG/ML. PATIENT 3 INITIAL RESULT WAS 26.72 NG/ML AND RETEST RESULT WAS 19.1 NG/ML. MALE DATA PROVIDED: CUSTOMER NORMAL VALUES ARE: 3.46 - 19.40 NG/ML. THE CUSTOMER TESTED 3 MALE SAMPLES THAT WERE FROM REPORTEDLY HEALTHY INDIVIDUALS AND OBTAINED THE FOLLOWING RESULTS: 23.9, 9.67, & 19.62. NO OTHER INFO AVAILABLE FOR THESE PATIENTS. THE CUSTOMER REPORTED LOT 52220UD00 AND LOT 53599UD00 WERE ON-BOARD THE ANALYZER WHEN THE DISCREPANT RESULTS WERE OBTAINED. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18120832 ↗MalfunctionJul 20, 2023High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR A PATIENT. THE PATIENT PROVIDED ANOTHER SAMPLE THE NEXT DAY AND THE RESULT WAS NORMAL. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER¿S REFERENCE RANGE: 108.78 TO 557.13 MIU/L. (B)(6) 2023 SAMPLE 1 INITIAL RESULT = 2671.55 MIU/L. RESULT AFTER PEG TREATMENT = 2300 MIU/L. SIEMENS RESULT = 1947.01 MIU/L. (B)(6) 2023 SAMPLE 2 INITIAL RESULT = 239.79 MIU/L; REPEATED RESULT = CONSISTENT WITH INITIAL RESULT. SIEMENS RESULT = 171.08 MIU/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17360305 ↗MalfunctionApr 10, 2023High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR A 40-YEAR-OLD FEMALE PATIENT WHEN COMPARING THE PATIENT¿S RESULTS WHEN TESTED AT OTHER LABORATORIES. THE FOLLOWING RESULTS WERE PROVIDED: AT THE CUSTOMER¿S SITE ON (B)(6) 2023: PROLACTIN RESULT = 106.9 NG/ML; REPEAT RESULT = 106.9 NG/ML. ON (B)(6) 2023: PROLACTIN RESULT = 106.1 NG/ML. REFERENCE RANGE: FEMALE = 5.18-26.53 NG/ML. CVS LAB ON (B)(6) 2023: PROLACTIN RESULT = 28.96 NG/ML. REFERENCE RANGE: FEMALE = 2.80-29.20 NG/ML; POST MENOPAUSE = 1.80-20.30 NG/ML; PREGNANT FEMALES: 9.70-208.50 NG/ML. PARDINI LAB ON (B)(6) 2023: PROLACTIN RESULT = 11.6 NG/ML; REPEAT RESULT POST DILUTION = 16.5 NG/ML; REPEAT RESULT POST TREATMENT WITH PEG (PROLACTIN CONCENTRATION AFTER TREATMENT WITH POLYETHYLENE GLYCOL) = 10.88 NG/ML. REFERENCE RANGE: FEMALE = 2.80-29.20 NG/ML; POST MENOPAUSE = 1.80-20.30 NG/ML. A NEW SAMPLE WAS COLLECTED FROM THE PATIENT AND TESTED FOR PROLACTIN AT THE CUSTOMER¿S SITE. THE RESULT WAS 89.98 NG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16710661 ↗MalfunctionApr 10, 2023High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I PROLACTIN RESULT ON A 20-YR OLD MALE PATIENT. RESULTS PROVIDED: (B)(6) 2022 SID (B)(6) PROLACTIN = 520 MU/L, ANOTHER LABORATORY = 14.7 UG/L (REFERENCE RANGE: 4.6-21.4 UG/L). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16706957 ↗MalfunctionFeb 24, 2023High Test Results; High Test Results
THE CUSTOMER OBSERVED A FALSELY ELEVATED PROLACTIN RESULT FOR A 21-YEAR-OLD FEMALE WHEN COMPARED TO THE ROCHE METHOD. THE FOLLOWING RESULTS WERE PROVIDED: (B)(6) 2023. SID: (B)(6). INITIAL RESULT 41.10 NG /ML, SENT TO BIOMNIS REPEATED ON COBAS ROCHE RESULT = 22.2 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16434660 ↗MalfunctionDec 19, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR 6 PATIENTS. THE CUSTOMER ALSO NOTICED A SHIFT UP IN BOTH LEVELS OF CONTROLS. THE SAMPLES WERE TESTED ON A DIFFERENT METHOD WHICH GENERATED LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: PATIENT 1: SAMPLE COLLECTION DATE: (B)(6) 2022, ALINITY RESULT = 52.44 NG/ML, CHEMILUMINESCENCE METHOD RESULT = 36.2 NG/ML, CHEMILUMINESCENCE REFERENCE VALUES: WOMEN: PRE-MENOPAUSE = 2.8 TO 29.2 NG/ML, POST-MENOPAUSE = 1.8 TO 20.3 NG/ML; MEN = 2.1 TO 17.7 NG/ML. PATIENT 2: ALINITY RESULTS PERFORMED ON (B)(6) 2022 = 80.11 NG/ML, 81.50 NG/ML, RESULT FROM HERMES PARDINI (HP) = 21.0 NG/ML. PATIENT 3: ALINITY RESULTS PERFORMED ON (B)(6) 2022 = 31.27 NG/ML, 30.56 NG/ML, RESULT FROM HERMES PARDINI (HP) = 9.3 NG/ML. PATIENT 4: SAMPLE COLLECTION DATE: (B)(6) 2022, ALINITY RESULT = 15.41 NG/ML, CHEMILUMINESCENCE METHOD RESULT = 9.7 NG/ML, CHEMILUMINESCENCE REFERENCE VALUES: WOMEN: PRE-MENOPAUSE = 2.8 TO 29.2 NG/ML, POST-MENOPAUSE = 1.8 TO 20.3 NG/ML; MEN = 2.1 TO 17.7 NG/ML. PATIENT 5: SAMPLE COLLECTION DATE: (B)(6) 2022, ALINITY RESULT = 35.06 NG/ML, CHEMILUMINESCENCE METHOD RESULT = 25.7 NG/ML, CHEMILUMINESCENCE REFERENCE VALUES: WOMEN:
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16003811 ↗MalfunctionNov 7, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR MULTIPLE FEMALE PATIENTS. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE FOR ALINITY: FEMALE 108-557 MIU/L; CUSTOMER¿S REFERENCE RANGE FOR ADVIA: 59-619 MIU/L): (B)(6) 2022. PATIENT 1 SID (B)(6) RESULT ON ALINITY = 833.37 MIU/L, REPEAT ON ADVIA = 425 MIU/L. PATIENT 2 SID (B)(6) RESULT ON ALINITY = 729.17 MIU/L, REPEAT ON ADVIA = 281 MIU/L. PATIENT 3 SID (B)(6) RESULT ON ALINITY = 601.29 MIU/L, REPEAT ON ADVIA = 370 MIU/L. PATIENT 5 RESULT ON ALINITY = 736.61 MIU/L, REPEAT ON ADVIA = 320 MIU/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15743471 ↗MalfunctionOct 12, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR A 37 YEAR OLD FEMALE PATIENT. RESULTS PROVIDED: ON (B)(6) 2022, PROLACTIN: 35.4 NG/ML. PREVIOUS RESULTS: ON (B)(6) 2019, PROLACTIN: 575 NG/ML. TREATED WITH 1 TABLET CABASER 1.0 MG PARLODEL. THE PATIENT WAS ALREADY DIAGNOSED WITH PROLACTINOMA. ON (B)(6) 2019 PROLACTIN: AROUND 100 NG/ML, DIAGNOSED AS CABASER-RESISTANT PROLACTINOMA, 2¿3 TABLETS OF CABASER 1.0 MG. PROLACTIN REMAINED SLIGHTLY BELOW 100 NG/ML ON (B)(6) 2020, PROLACTIN: 3.6 NG/ML. ON (B)(6) 2020, PROLACTIN: 19.8 NG/ML. ON (B)(6) 2021, PROLACTIN: 62.3 NG/ML. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15587977 ↗MalfunctionSep 3, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS FOR ONE FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER PROVIDED REFERENCE RANGE FOR FEMALES IS 5.18 TO 26.53 NG/ML). PREVIOUS RESULT WAS >200 NG/ML. THE PATIENT RECEIVED TREATMENT AND HER PROLACTIN RESULTS GRADUALLY DECREASED. THE LOWEST RESULT GENERATED WAS 15 NG/ML. THE NEXT SAMPLE GENERATED A RESULT ABOVE 60 NG/ML. THE PATIENT¿S TREATMENT DURING THIS TIME DID NOT CHANGE. THE CUSTOMER TREATED THE SAMPLE WITH PEG PRECIPITATION AND GENERATED A RESULT OF AROUND 5 NG/ML. THE CUSTOMER IS QUESTIONING IF THE PAST RESULTS WERE IMPACTED BY MACROPROLACTIN AS WELL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15353586 ↗MalfunctionJun 15, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS WHEN COMPARED TO ANOTHER METHOD FOR ONE PATIENT. THE SAMPLE RESULT ON ALINITY WAS 32 NG/ML, SAME SAMPLE TESTED ON SIEMEN ANALYZER AND THE RESULT WAS 10 (NORMAL RANGE 5.18 TO 26.53). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14700621 ↗MalfunctionApr 1, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS ON ONE PATIENT. THE RESULTS PROVIDED WERE: ON (B)(6) 2022 SID (B)(6) =36.72 NG/ML/REPEATED AT ARUP=12.5 NG/ML THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13978591 ↗MalfunctionMar 29, 2021High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I PROLACTIN RESULTS ON MULTIPLE PATIENTS EXAMPLE RESULTS PROVIDED (UOM = NG/ML): (B)(6). CUSTOMER¿S PROLACTIN REFERENCE RANGE: FEMALES (3.2 - 26.5 NG/ML), MALE (3.5 - 19.4 NG/ML). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11580805 ↗
FDA clearance and classification
The FDA regulates product code CFT (radioimmunoassay, prolactin (lactogen)) under 21 CFR 862.1625 as a Class I device. 21 CFR 862.1625 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Alinity 07P6620
Has the Alinity 07P6620 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00380740121570), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Alinity 07P6620?
The FDA's MAUDE database lists 25 adverse event reports for this device, received between Mar 29, 2021 and Mar 9, 2026. In those reports, the most common specific device problem is “High Test Results” (25). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Alinity 07P6620?
None of the reports are classified as deaths, 0 as injuries and 25 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Alinity 07P6620 FDA approved?
Devices of this type (product code CFT) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Abbott Ireland.
Who makes the Alinity 07P6620, and is it still sold?
The FDA's device database lists Abbott Ireland as the company responsible for it (brand Alinity, model 07P6620). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN) devices
- VITROS 1849793Ortho Clinical Diagnostics · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)1 recall0 reports
- ARCHITECT 7K76-25Abbott Ireland · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)38 reports
- Elecsys Prolactin II 07027737190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)18 reports
- Elecsys Prolactin II 03203093190Roche Diagnostics GmbH · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)13 reports
- ARCHITECT 7K76-35Abbott Ireland · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)12 reports
- Alinity 07P6630Abbott Ireland · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)10 reports
- Atellica IM PRL 10995656Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)5 reports
- ADVIA Centaur® PRL 10309976Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page