AlinityRecalls & reported problems across all models
Abbott Ireland · 22 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 643
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls
- 7
We found 22 models of Alinity by Abbott Ireland in the FDA's device registry (GUDID). Together they are named in 643 adverse event reports in the FDA's MAUDE database, received between Aug 20, 2018 and Sep 29, 2026 — classified as 4 injuries, 639 malfunctions.
The most frequently reported specific device problem is “High Test Results” (424 reports).
7 FDA recall records name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Alinity recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Nov 21, 2018Terminated
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Z-2208-2019 · Root cause: Device Design · Recall details · FDA record ↗
- Nov 21, 2018Terminated
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Z-0669-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Nov 21, 2018Terminated
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Z-0668-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Nov 21, 2018Terminated
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Z-0667-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Nov 21, 2018Terminated
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Z-0666-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Nov 21, 2018Terminated
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Z-0665-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Nov 21, 2018Terminated
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Z-0664-2019 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported Alinity problems
Device problems
- High Test Results424
- Low Test Results150
- Device Damaged Prior to Use24
- Unable to Obtain Readings24
- Incorrect, Inadequate or Imprecise Result or Readings22
- False Positive Result16
- Imprecise or Erratic Results6
- False Negative Result3
Patient problems
- No Clinical Signs, Symptoms or Conditions589
- No Patient Involvement21
- No Known Impact Or Consequence To Patient20
- No Consequences Or Impact To Patient6
- Insufficient Information5
- Cardiac Enzyme Elevation2
Alinity models
22 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Alinity 07P7030RADIOIMMUNOASSAY, FREE THYROXINE1 recall168 reports
- Alinity 07P4830RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE2 recalls91 reports
- Alinity 07P7020RADIOIMMUNOASSAY, FREE THYROXINE79 reports
- Alinity 04Z2125IMMUNOASSAY METHOD, TROPONIN SUBUNIT74 reports
- Alinity 07P5121SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN1 recall15 reports
- Alinity 07P6530FERRITIN, ANTIGEN, ANTISERUM, CONTROL1 recall7 reports
- Alinity 08P4532SYSTEM, TEST, VITAMIN D1 recall3 reports
- Alinity 08P3620RADIOIMMUNOASSAY, PROGESTERONE1 recall1 report
- Alinity 07P4820RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE44 reports
- Alinity 07P6620RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)31 reports
- Alinity 08P1934PHOTOMETRIC METHOD, MAGNESIUM30 reports
- Alinity 04Z2135IMMUNOASSAY METHOD, TROPONIN SUBUNIT26 reports
- Alinity 08P1925PHOTOMETRIC METHOD, MAGNESIUM20 reports
- Alinity 04T9120ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
- Alinity 07P5020RADIOIMMUNOASSAY, ESTRADIOL12 reports
- Alinity 07P6630RADIOIMMUNOASSAY, PROLACTIN (LACTOGEN)10 reports
- Alinity 07P5030RADIOIMMUNOASSAY, ESTRADIOL6 reports
- Alinity 07P5131SYSTEM, TEST, HUMAN CHORIONIC GONADOTROPIN5 reports
- Alinity 04U0520BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM3 reports
- Alinity 04T9820Photometric method, iron (non-heme)1 report
- Alinity 09P6324IGM, ANTIGEN, ANTISERUM, CONTROL1 report
- Alinity 08P0424TURBIDIMETRIC METHOD, PROTEIN OR ALBUMIN (URINARY, NON-QUANT.)1 report
Recent Alinity adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932189 ↗MalfunctionSep 28, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923280 ↗MalfunctionSep 28, 2026High Test Results
The customer reported falsely elevated Alinity I Free T4 generated from Alinity i Processing Module (SN:(b)(6)) and provided the following data for a 74 year old female patient:On (b)(6)2026, Sample ID (b)(6) initial result was 24.4 pmol/L and when repeated on another Alinity i Processing Module, the result was 13.7 pmol/ L (Customer Normal range: 9-19.9 pmol/L).Other laboratory data provided:TSH was 2.29 mIU/L (Reference range: 0.35-4.94)Per the customer discrepant results were not reported out to the medical provider. There was no reported impact to patient management.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50916380 ↗MalfunctionSep 21, 2026High Test Results
The customer reported falsely elevated Alinity i Free T4 results for multiple patients. The following data were provided:SID (b)(6)(female, born 1947)(7SEP2026)Initial: >64.35 pmol/L,Repeat: 19.13 pmol/L, 17.70 pmol/L.SID (b)(6)(male, born 1979)((b)(6) 2026)Initial: >64.35 pmol/L.Repeat: 15.31 pmol/L. No impact to patient management was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50846171 ↗MalfunctionSep 21, 2026Low Test Results
The customer reported falsely depressed Alinity i TSH results for three patients. The samples were repeated with higher results. The following data was provided:Initial Results:Sample id: (b)(6) : < 0.0083 mIU/mL,Sample id: (b)(6) : <0.0131 mIU/mL,Sample id: (b)(6) : < 0.0083 mIU/mL.Rerun results:Sample id:(b)(6): 1.3704 mIU/mL,Sample id:(b)(6) : 3.7146 mIU/mL,Sample id:(b)(6) : 0.8114 mIU/mL There was no impact to patient management reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50842562 ↗MalfunctionSep 16, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50794333 ↗MalfunctionSep 16, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50794332 ↗MalfunctionSep 16, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50794211 ↗MalfunctionSep 16, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50794208 ↗MalfunctionSep 16, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50791568 ↗MalfunctionSep 15, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50780513 ↗MalfunctionSep 15, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50778619 ↗MalfunctionSep 15, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50778053 ↗MalfunctionSep 15, 2026High Test Results
The customer observed elevated Alinity i STAT High Sensitivity Troponin-I results for a 75-year-old patient diagnosed with lung cancer status post left lobectomy in (b)(6) 2026. The patient presented with increased shortness of breath with activities of daily living (ADLs) and sharp pain to left upper quadrant. The patient had an EKG, chest x-ray and pulmonary CT (contrast) completed prior to transferring to the hospital. The following data was provided: (b)(6) 2026, SID: (b)(6). Alinity i STAT High Sensitivity Troponin-I results: 512.2 pg/mL, 548.8 pg/mL, 492.3 pg/mL, 506.8 pg/mL, 506.5 pg/mL,528.3 pg/mL, 521.1 pg/mL. Repeat testing on all instruments remained consistent with the initial findings. The specimen was also tested on the Beckman Access 2 and the result was 11 ng/L (Beckman reference range 16-49 ng/L is Abnormal High). The patient was transferred to ESJ and the Beckman Troponin results at ESJ were 14 ng/L and 15 ng/L. Other results provided: BNP result 607 and 632 pg/mL. There was no negative impact to patient management reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50771676 ↗MalfunctionSep 15, 2026High Test Results
The customer reported a falsely elevated Alinity c Creatinine2 result generated on the Alinity c Processing Module. The questioned creatinine result did not match previous results, and the patient was admitted to the hospital. The following data was provided:(b)(6)2026, SID (b)(6): Initial Creatinine2 result = 2.37 mg/dL (Normal range: 0.72 to 1.25 mg/dL)(b)(6)2026, Repeat Creatinine2 result = 0.70 mg/dL (measured from original sample stored in the refrigerator) There was no impact to patient management reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50771439 ↗
Other Abbott Ireland products
All Abbott Ireland devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page