Alinity 07P7030Recalls & reported problems

Abbott Ireland · Alinity i Free T4 Reagent Kit 1200 Tests · Class 2

Quick summary

This Abbott Alinity i free thyroxine test kit has 166 reports, with no deaths and no injuries. Most reports describe results that read too high, with some reading too low and repeat tests giving different values, and none list a patient symptom. The kit had 1 recall, now closed, from November 2018, over reagent cartridges with a damaged bottle neck that could fail to load on the instrument. Reports fell from 59 in the prior 12 months to 28 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
168
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 168 adverse event reports that list this exact device (DI 00380740121822), received between Nov 21, 2018 and Sep 28, 2026 — classified as 168 malfunctions.

The most frequently reported specific device problem is “High Test Results” (142 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 11 reportsApr 2024: 9 reportsMay 2024: 15 reportsJun 2024: 5 reportsJul 2024: 6 reportsAug 2024: 11 reportsSep 2024: 6 reportsOct 2024: 12 reportsNov 2024: 4 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 4 reportsMar 2025: 5 reportsApr 2025: 8 reportsMay 2025: 3 reportsJun 2025: 7 reportsJul 2025: 1 reportsAug 2025: 6 reportsSep 2025: 8 reportsOct 2025: 5 reportsNov 2025: 3 reportsDec 2025: 3 reportsJan 2026: 6 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 3 reportsMay 2026: 3 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 4 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202615
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 21, 2018Terminated

    Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

    Z-0665-2019 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • High Test Results142
  • Low Test Results19
  • Incorrect, Inadequate or Imprecise Result or Readings7
  • Device Damaged Prior to Use1
  • High Readings1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions167
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 168 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026High Test Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923280 ↗
  • MalfunctionSep 28, 2026High Test Results

    The customer reported falsely elevated Alinity I Free T4 generated from Alinity i Processing Module (SN:(b)(6)) and provided the following data for a 74 year old female patient:On (b)(6)2026, Sample ID (b)(6) initial result was 24.4 pmol/L and when repeated on another Alinity i Processing Module, the result was 13.7 pmol/ L (Customer Normal range: 9-19.9 pmol/L).Other laboratory data provided:TSH was 2.29 mIU/L (Reference range: 0.35-4.94)Per the customer discrepant results were not reported out to the medical provider. There was no reported impact to patient management.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916380 ↗
  • MalfunctionSep 21, 2026High Test Results

    The customer reported falsely elevated Alinity i Free T4 results for multiple patients. The following data were provided:SID (b)(6)(female, born 1947)(7SEP2026)Initial: >64.35 pmol/L,Repeat: 19.13 pmol/L, 17.70 pmol/L.SID (b)(6)(male, born 1979)((b)(6) 2026)Initial: >64.35 pmol/L.Repeat: 15.31 pmol/L. No impact to patient management was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846171 ↗
  • MalfunctionAug 18, 2026Low Test Results

    The customer observed falsely depressed Alinty i Free T4 results for one sample. The following data was provided (Free T4 normal range 9-19 pmol/L):(b)(6) 2026 (30-year-old female)SID (b)(6) initial result 8.58 pmol/L (Lot 87264UD00), repeated on 31 July 2026 8.08 / 8.81 / 9.10 pmol/L.No impact to patient management was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50299869 ↗
  • MalfunctionAug 18, 2026Low Test Results

    The customer observed falsely depressed Alinty i Free T4 results for one sample. The following data was provided (Free T4 normal range 9-19 pmol/L):(female, 68-year-old)(b)(6) 2026 SID (b)(6) initial result 10.27 pmol/L (Lot 86129UD00), repeated 8.77 and 9.63 pmol/L (Lot 86129UD00). No impact to patient management was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50299805 ↗
  • MalfunctionAug 17, 2026Low Test Results

    The customer observed falsely depressed Alinity i Free T4 results generated for patient samples. The following data was provided (Customer¿s reference range 0.75-1.22 ng/dL):Sample ID (b)(6), (41-year-old Female) Initial result = 0.69 ng/dL, repeat = 0.84 ng/dL.Sample ID (b)(6), (39-year-old Female) Initial result = 0.68 ng/dL, repeat = 0.76 ng/dL.Sample ID (b)(6), (36-year-old Female) Initial result = 0.70 ng/dL, repeat = 0.79 ng/dL.Sample ID (b)(6), (b)(6) Initial result = 0.74 ng/dL, repeat = 0.88 ng/dL.Sample ID (b)(6), (62-year-old Female) Initial result = 0.73 ng/dL, repeat = 0.87 ng/dL.Sample ID (b)(6), (50-year-old Female) Initial result = 0.67 ng/dL, repeat = 0.76 ng/dL.Sample ID (b)(6), (45-year-old Female) Initial result = 0.73 ng/dL, repeat = 0.84 ng/dL.Sample ID (b)(6), (20-year-old Female) Initial result = 0.66 ng/dL, repeat = 0.75 ng/dL .No impact to patient management was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50275427 ↗
  • MalfunctionAug 12, 2026High Test Results

    The customer observed falsely elevated Alinity i Free T4 results for two samples. The following data was provided (normal range 9.01 to 19.05 pmol/L):(b)(6) 2026 Patient 1 (79-year-old female)Initial result was 19.31 pmol/L, specimen was retested again within 45 min interval, and result was 16.58 / 18.07 / 17.35 pmol/L,Patient 2 (53-year-old female)Initial result was 22.96 pmol/L, specimen was retested again within 45 min interval, and result was 13.50 pmol/L. The customer recentrifuged specimen and results were 13.82 and 14.06 pmol/L. No impact to patient management was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50224329 ↗
  • MalfunctionJun 8, 2026Low Test Results

    THE CUSTOMER OBSERVED A FALSELY DEPRESSED ALINITY I FREE T4 RESULT FOR 2-MONTH-OLD MALE PEDIATRIC PATIENT. THE PATIENT HAS A HISTORY OF CONGENITAL CMV AND COMPLETED TREATMENT WITH VALGANCICLOVIR. (CUSTOMER PROVIDED REFERENCE RANGE FOR CHILD: 11.40¿21.90 PMOL/L). (B)(6) 2026 SID (B)(6) INITIAL RESULT = 10.7 PMOL/L. ADDITIONAL LAB RESULT PROVIDED: TSH RESULT = 1.26 MIU/L WITHIN NORMAL RANGE (0.35¿4.94 MIU/L). (B)(6) 2026 SID (B)(6) PATIENT TESTED AT ANOTHER LABORATORY FREE T4 RESULT = 14.54 PMOL/L, NORMAL TSH RESULT (NO SPECIFIC RESULT PROVIDED). ON (B)(6) 2026 NEW SAMPLE FREE T4 RESULT WITH SID (B)(6) = 10.97 PMOL/L, (NEW REFERENCE RANGE. DUE TO PATIENT AGE FOR FREE T4: 9.36¿14.49 PMOL/L) WITHIN NORMAL RANGE. ADDITIONAL LAB RESULT PROVIDED: TSH 1.60 MIU/ML WITHIN NORMAL RANGE (0.35¿4.94 MIU/L). SAMPLE WITH SID (B)(6) WAS FROZEN AND RETESTED ON (B)(6) 2026 ON (B)(6) WITH A RESULT = 12.34 PMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25428171 ↗
  • MalfunctionJun 4, 2026High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 FOR A 57-YEAR-OLD WHITE FEMALE. PATIENT HISTORY INCLUDES: RHEUMATOID ARTHRITIS, FIBROMYALGIA, IBS, ASTHMA, TYPE 2 DIABETES, HAY FEVER, ANXIETY / PANIC DISORDER AND TAKING MEDICATIONS BUDESONIDE + FORMOTEROL INHALER, CETIRIZINE, FEXOFENIDINE, FOLIC ACID, HYOSCINE BUTYLBROMIDE, IBUPROFEN, LEFLUNOMIDE, METHOTREXATE, OMEPRAZOLE, PARACETAMOL, PREDNISOLONE, SALBUTAMOL, AND SERTRALINE. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 9.0-19.1 PMOL/L): ON (B)(6) 2026, SID (B)(6), INITIAL FREE T4 RESULT= 26.10 PMOL/L; REPEAT RESULTS= 18.54 PMOL/L, 15.95 PMOL/L; (B)(6) 2026, REPEAT RESULTS= 12.31 PMOL/L, 11.32 PMOL/L ADDITIONAL RESULTS PROVIDED: PT RESULT= 9.6 (REFERENCE RANGE 9.7-12.0), INR RESULT= 0.5, APTT RESULT= 22.9. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25400385 ↗
  • MalfunctionMay 12, 2026High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR TWO PATIENTS. THE RESULTS WERE NOT REPORTED OUT. THE SAMPLES WERE CENTRIFUGED AND REPEATED WITH LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) 26 YEAR OLD MALE PATIENT DIAGNOSED WITH DECOMPENSATED DILATED CARDIOMYOPATHY. PROCESSED ON (B)(6) 2026 INITIAL RESULT = >64.3 REPEAT RESULTS = 13.89 AND 14.56 PMOL/L. HISTORICAL RESULT FROM (B)(6) 2026 = 17.37 PMOL/L. THE PATIENT IS ON THE FOLLOWING MEDICATIONS: CASPOFUNGIN, ZOPICLONE, CEFAZOLIN, MORPHINE, MYCOPHENOLATE, METHYLPREDNISOLONE, TEICOPLANIN, PIPERACILLIN, TAZOBACTAM, FUROSEMIDE, LEVETIRACETAM, HYDRALAZINE, NICARDIPINE, LANSOPRAZOLE, TRAMADOL, CHLORAMPHENAMINE, CYCLIZINE, NALOXONE, MELATONIN, HEPARIN, ACTRAPID INSULIN, ONDANSETRON. SID (B)(6) 80 YEAR OLD MALE PATIENT, ELECTIVE ANGIOGRAM FOR EXERTIONAL ANGINA AND SHORTNESS OF BREATH ON EXERTION, PROCESSED ON (B)(6) 2026 INITIAL RESULT = 20.21 REPEAT RESULTS = 15.26 AND 15.51 PMOL/L THE PATIENT IS ON THE FOLLOWING MEDICATIONS: CLOPIDOGREL, BUDESONIDE + FORMOTEROL INHALER, DARBEPOETIN ALFA SC, NITRATES SUBLINGUAL, HYDROXOCOBALAMIN IM, ASPIRIN, LEVOTHYROXINE, RAMIPRIL, SIMVASTATIN. REFERENCE RANGE 9-19.1

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150768 ↗
  • MalfunctionMay 11, 2026High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR A 32-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS 9.0-19.0 PMOL/L): SAMPLE ID (B)(6) INITIAL RESULT, ON (B)(6) 2026, WAS >64.35, REPEAT RESULT WAS 19.13, REPEAT RESULTS, ON ANOTHER ANALYZER, WERE 18.29, 18.17, AND 18.39 PMOL/L. THE TSH RESULTS FOR THIS PATIENT WERE ALSO PROVIDED: 2.521, REPEAT WAS 2.631 MIU/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25136271 ↗
  • MalfunctionMay 8, 2026High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 AND PROVIDED THE FOLLOWING DATA: CUSTOMER NORMAL RANGE : 0.70 - 1.48 NG/DL SAMPLE ID (B)(6) - MALE , 1 YEAR OLD. FREE T4 INITIAL RESULT WAS 2.88 NG/DL AND REPEAT RESULT WAS 0.76 NG/DL OTHER LABORATORY VALUES PROVIDED: FT3 : 2.46 PG/ML TSH : 1.8672 UIU/ML SAMPLE ID :(B)(6) - FEMALE , 16 YEAR OLD. FREE T4 INITIAL RESULT WAS > 5.00 NG/DL AND REPEAT RESULT WAS 1.16 NG/DL PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25125868 ↗
  • MalfunctionApr 30, 2026High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I FREE T4 GENERATED FROM ALINITY I PROCESSING MODULE AND PROVIDED THE FOLLOWING DATA: SAMPLE ID (B)(6), THE INITIAL RESULT WAS > 5.00 NG/DL AND REPEAT RESULT WAS 1.36 NG/DL. THE CUSTOMER PROVIDED ADDITIONAL LABORATORY DATA: TSH: WAS 0.4590 UIU/ML & 0.4560 UIU/ML FREE T3: WAS 3.13 PG/ML & 2.73 PG/ML TRAB: WAS 1.10 IU/L TG WAS 14.24 NG/ML. HISTORICAL TSH VALUES: THE CUSTOMER COMMUNICATED THAT THE PATIENT'S TSH LEVELS ARE IMPROVING OVER THE PAST SIX MONTHS: 0.2 , 0.4 , 0.5. PATIENT INFORMATION: 80-YEAR-OLD MALE WITH A MEDICAL HISTORY OF PSVT (PAROXYSMAL SUPRAVENTRICULAR TACHYCARDIA) AND HYPERTHYROIDISM. MEDICATIONS: THIAMAZOLE & WARFARIN THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25046373 ↗
  • MalfunctionApr 3, 2026High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE 9 - 23 PMOL/L): (B)(6) 2026, SID (B)(6), INITIAL FREE T4 RESULT= 48.6 PMOL/L; REPEAT RESULTS= 17.4 AND 16.6 PMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24777815 ↗
  • MalfunctionApr 3, 2026High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 (FT4) RESULTS ON A 88YRS OLD FEMALE PATIENT. THE RESULTS PROVIDED WERE: ON (B)(6) 2025 SID (B)(6) INITIAL=> 5.00 NG/ML /REPEATED=1.40 NG/ML. LABORATORY REFERENCE RANGE FOR FT4=0.7 TO 1.48 NG/DL. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24775034 ↗
  • MalfunctionMar 16, 2026High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR A 58-YEAR-OLD MALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S NORMAL RANGE IS 9-19 PMOL/L): SAMPLE ID (B)(6), INITIAL RESULT WAS 42, REPEAT RESULT WAS 15 PMOL/L. ADDITIONAL LAB RESULTS WERE PROVIDED WITH TSH RESULT BEING 1.8 MIU/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24606540 ↗
  • MalfunctionMar 11, 2026High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR A 3-YEAR-OLD FEMALE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: (CUSTOMER PROVIDED REFERENCE RANGE: 0.7 - 1.48 NG/DL). (B)(6) 2026 SID (B)(6) INITIAL RESULT = 1.60 NG/DL, REPEAT RESULT = 1.05 NG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24567732 ↗
  • MalfunctionFeb 11, 2026High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE 75-YEAR-OLD FEMALE PATIENT. THE SAMPLE WAS REPEATED WITH A LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID(B)(6): ALINITY SERIAL NUMBER (B)(6) RESULT = 27.48 PMOL/L REPEAT RESULT = 17.94 PMOL/L. ALINITY SERIAL NUMBER (B)(6) RESULT = 17.84 PMOL/L. ALINITY SERIAL NUMBER (B)(6) RESULT = 17.16 PMOL/L. REFERENCE RANGE IS 9-19 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24316193 ↗
  • MalfunctionJan 30, 2026High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 RESULTS FOR ONE PATIENT. THE SAMPLE WAS FROZEN BETWEEN EACH MEASUREMENT. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE 12-22 PMOL/L): FIRST SAMPLE (THE PATIENT IS POSITIVE FOR RHEUMATOID FACTOR (RF)), SID (B)(6), INITIAL FREE T4 RESULT 35.3 PMOL/L ((B)(6) 2025), REPEATED 18.1 ((B)(6) 2025), 23.2 PMOL/L ((B)(6) 2025), AND 26.9 PMOL/L ((B)(6) 2025). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24220469 ↗
  • MalfunctionJan 27, 2026High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I FREE T4 AND PROVIDED THE FOLLOWING DATA FOR SAMPLE ID (B)(6): ON (B)(6) THE RESULT WAS 1.33 NG/DL, ON (B)(6), THE RESULT WAS 1.75 NG/DL, ON (B)(6) THE RESULTS WERE 1.12 NG/DL & 1.10 NG/DL, CUSTOMER REFERENCE RANGE: 0.81 TO 1.32 NG/DL. PATIENT INFORMATION: 16 YEAR OLD FEMALE. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT. UPDATE: THE CUSTOMER REPORTED MORE PATIENTS WITH FALSELY ELEVATED ALINITY I FREE T4 AND PROVIDED THE FOLLOWING DATA: SAMPLE ID (B)(6) (MALE, 47 YEARS OLD): ON (B)(6) RESULTS WERE 1.09 NG/DL (FROM SN: (B)(6)) AND 2.28 NG/DL (FROM SN: (B)(6)) ON (B)(6) RESULTS WERE 1.05 NG/DL (FROM SN: (B)(6)) AND 1.03 NG/DL (FROM SN: (B)(6)). SAMPLE ID (B)(6) (FEMALE, 37 YEARS OLD): ON (B)(6) RESULTS WERE 1.14 NG/DL (FROM SN: (B)(6)) AND 1.42 NG/DL (FROM SN: (B)(6)) ON (B)(6) RESULTS WERE 1.08 NG/DL (FROM SN: (B)(6)) AND 1.09 NG/DL (FROM SN: (B)(6)). SAMPLE ID (B)(6) (GENDER AND AGE NOT PROVIDED): ON (B)(6) THE RESULT WAS 4.06 NG/DL (FROM SN: (B)(6)). ON (B)(6) THE RESULTS WERE 1.69 NG/DL (FROM SN: (B)(6)), 0.97 NG/DL (FROM SN: (B)(6)), 1.29 NG/DL (FROM SN: (B)(6)), 1.82 NG/DL (FROM SN: (B)(6)) & 0.96 NG/DL (FROM SN: (B)(6)). THE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24192490 ↗
  • MalfunctionJan 26, 2026High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S NORMAL RANGE IS 10.8-20.2 PMOL/L): SAMPLE ID (B)(6) INITIAL RESULT WAS >64.35, REPEAT RESULT WAS 15.9, REPEAT RESULT, ON ANOTHER ANALYZER, WAS 17.3 PMOL/L. ADDITIONAL LABORATORY RESULTS WERE PROVIDED: TSH RESULT WAS 3.03, REPEAT RESULT WAS 3.02 UIU/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24180121 ↗
  • MalfunctionJan 12, 2026High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 (FT4) RESULT FOR A FEMALE PATIENT IN HER 30S WITH A MEDICAL HISTORY OF GRAVE¿S DISEASE AND IS CURRENTLY ON THYRADIN (THYROID HORMONE REPLACEMENT MEDICATION) GENERATED ON THE ALINITY I PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED: PREVIOUS FT4 RESULT (TESTED IN (B)(6) 2024) = 1.30 NG/DL. ON (B)(6) 2025, SID (B)(6): INITIAL FT4 RESULT = 2.16 NG/DL (DUE TO THE PREVIOUS VALUE, THE CUSTOMER RETESTED THE SAMPLE). REPEAT FT4 RESULTS: FIRST REPEAT (SID 84) = 1.41 NG/DL; SECOND REPEAT (SID 84-2) = 1.35 NG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24042854 ↗
  • MalfunctionJan 12, 2026High Test Results

    THE CUSTOMER OBSERVED ELEVATED ALINITY I FREE T4 AND PROVIDED THE FOLLOWING DATA: ON (B)(6) 2025, SAMPLE ID (B)(6) RESULT WAS > 64 (CUSTOMER REFERENCE RANGE 9-19 PMOL/L). THE CUSTOMER REPORTED THAT THERE WAS NOT ENOUGH SAMPLE TO RETEST. THE CUSTOMER PROVIDED THE FOLLOWING HISTORICAL AND ADDITIONAL THYROID TEST RESULTS: ON (B)(6) 2024 (SAMPLE ID (B)(6), FREE T4 WAS 15 PMOL/L WAS AND TSH WAS 3.42 MU/L, ON (B)(6) 2025 (SAMPLE ID (B)(6), FREE T4 WAS 15 PMOL/L WAS AND TSH WAS 3.40 MU/L, ON (B)(6) 2024 (SAMPLE ID (B)(6), FREE T4 WAS 14 PMOL/L WAS AND TSH WAS 5.68 MU/L, ON (B)(6) 2025, SAMPLE ID (B)(6), TSH WAS 1.29 MU/L. PER THE CUSTOMER DISCREPANT RESULTS WERE NOT REPORTED OUT TO THE MEDICAL PROVIDER. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24040641 ↗
  • MalfunctionJan 8, 2026Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER OBSERVED ERRATIC ALINITY I FREE T4 FOR MULTIPLE PATIENTS. THE CUSTOMER NOTES THAT THE RESULTS WERE WITHIN NORMAL RANGE, BUT LATER THE RESULTS WERE OUT OF REFERENCE RANGE. THE FOLLOWING DATA WAS PROVIDED: REFERENCE RANGES: ADULTS: 0.75 TO 1.22 NG/DL. PEDIATRICS: < 1 MONTH NOT AVAILABLE 1 TO 23 MONTHS: 0.86 TO 1.32 NG/DL. 2 TO 12 YEARS: 0.84 TO 1.32 NG/DL. 13 TO 18 YEARS: 0.81 TO 1.32 NG/DL. SID (B)(6), 9-YEAR-OLD MALE, INITIAL FREE T4 RESULT= 1.30 NG/DL; REPEAT RESULT= 1.93 NG/DL. SID (B)(6), 39-YEAR-OLD MALE, INITIAL FREE T4 RESULT= 1.06 NG/DL; REPEAT RESULT= 1.30 NG/DL. SID (B)(6), 73-YEAR-OLD MALE, INITIAL FREE T4 RESULT= 1.32 NG/DL; REPEAT RESULT= 1.55 NG/DL. FURTHER TESTING OF THE SAMPLES WAS PERFORMED: SID (B)(6) (B)(6) 2025, ANALYZER (B)(6), FREE T4 RESULT= 1.30 NG/DL (B)(6) 2025, AFTER HOMOGENIZING THE SAMPLE, ANALYZER (B)(6), RESULT= 1.93 NG/DL; ANALYZER (B)(6), RESULT= 1.96 NG/DL, 1.88 NG/DL, 1.87 NG/DL (B)(6) 2025, AFTER RECENTRIFUGING THE SAMPLE, ANALYZER (B)(6) RESULT= 1.02 NG/DL; ANALYZER (B)(6) RESULT= 1.06 NG/DL; (B)(6) 2025, ANALYZER (B)(6) RESULT= 1.06 NG/DL SID (B)(6) (B)(6) 2025, ANALYZER (B)(6), FREE T4 RESULT= 1.06 NG/DL (B)(6) 2025, AFTER HOMOGENI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24005213 ↗
  • MalfunctionDec 24, 2025High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 RESULTS FOR TWO SAMPLES. THE FOLLOWING DATA WAS PROVIDED: (B)(6) 2025 SAMPLE 1 SID (B)(6) INITIAL RESULT >64.350 PMOL/L, REPEATED 10.341 AND 12.188 PMOL/L, TSH 0.845 MIU/L SERUM LIPAEMIC RESULT 61.00 INDEX (1+). SAMPLE 2 SID (B)(6) INITIAL RESULT >64.350 PMOL/L, REPEATED 49.596 / 55.342 / 43.883 PMOL/L, TSH RESULT <0.008 MIU/L SERUM HEMOLYSE RESULT 166 INDEX (2+). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23888423 ↗

Other RADIOIMMUNOASSAY, FREE THYROXINE devices