ARCHITECT 7P32-21Recalls & reported problems
Abbott GmbH · Total Bilirubin · Class 1
All models of ARCHITECT (48)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380740129033), received between Nov 3, 2017 and Nov 26, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Low Test Results2
- High Readings1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 26, 2018High Readings; High Readings
THE CUSTOMER REPORTED THAT THEY ARE OBSERVING FALSELY ELEVATED TOTAL BILIRUBIN ON THEIR ARCHITECT COMPARED TO WHAT THEIR SISTER SITE IS GETTING ON BOTH PATIENT RESULTS AND CONTROLS. THE EXAMPLE RESULTS PROVIDED WERE FOR 2 ADULT PATIENTS: (B)(6) ON (B)(6) 2018 =1.9MG/DL (RANGE = 0-1.2MG/DL) / SAMPLE SENT TO ARUP = 1.1 / PREVIOUSLY ON (B)(6) 2017 BY ROCHE = 1.3; (B)(6) ON (B)(6) 2018 = 2.3MG/DL / SENT TO ARUP = 1.4 / PREVIOUSLY IN 2017 = 0.8 (UNKNOWN METHOD). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8101461 ↗MalfunctionJul 12, 2018Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED RESULTS WHILE USING TOTAL BILIRUBIN REAGENTS. THE FOLLOWING DATA WAS PROVIDED. SID (B)(6) INITIAL AND REPEAT RESULTS READ 0.481 MG/DL (REPORTED AS LESS THAN 1.50 MG/DL). REPEAT USING ANOTHER ANALYZER 2.67, 2.74 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7684141 ↗MalfunctionNov 3, 2017Low Test Results; Low Test Results
THE CUSTOMER STATED THAT A FALSELY DECREASED ARCHITECT TOTAL BILIRUBIN RESULT WAS GENERATED FOR A NEWBORN PATIENT (SAMPLE ID (B)(6)) ON (B)(6) 2017. INITIAL TESTING WAS 12.96 MG/DL. REPEAT TESTING WAS 0.57 MG/DL WHICH IS SUSPECTED TO BE FALSELY DECREASED. THE INITIAL HIGH RESULT WAS VERIFIED BY 1:1 DILUTION TESTING (12.10 MG/DL). THE CUSTOMER STATED THAT IT IS POSSIBLE THAT THE SECOND VALUE WAS GENERATED AFTER THE SAMPLE SAT FOR 30 MINUTES AT ROOM TEMPERATURE UNDER THE INFLUENCE OF LIGHT WHICH IS NOT IN ACCORDANCE WITH SAMPLE HANDLING INSTRUCTIONS IN THE REAGENT PACKAGE INSERT. THE CUSTOMER WILL ATTEMPT TO RETRACE THE HANDLING OF THIS SAMPLE. NO ADVERSE IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7001262 ↗
Other BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM devices
- ABL90 FLEX PLUS Analyzer 393-092Radiometer Medical ApS · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM1 recall1 death175 reports
- ADVIA® Chemistry Total Bilirubin_2 TBIL_2 Reagents 10341113Siemens Healthcare Diagnostics Inc. · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM1 recall0 reports
- Alinity 04V5121Abbott GmbH · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM36 reports
- Drager Jaundice Meter, model JM-105 MU20105Draeger Medical Systems, Inc. · Bilirubin (Total And Unbound) In The Neonate Test System1 death23 reports
- Alinity 04V5131Abbott GmbH · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM13 reports
- Alinity 04U0520Abbott Ireland · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM3 reports
- ARCHITECT 04T0920Abbott Ireland · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page