Skip to content

Alinity 04U0520Recalls & reported problems

Abbott Ireland · Total Bilirubin2 3500 Tests · Class 1

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00380740167585), received between Dec 4, 2025 and Jul 20, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 20, 2026High Test Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25879944 ↗
  • MalfunctionJan 12, 2026High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED TOTAL BILIRUBIN2 RESULTS GENERATED BY THE ALINITY C PROCESSING MODULE FOR MULTIPLE PATIENTS. PHYSICIANS QUESTIONED THESE RESULTS AS THEY WERE INCONSISTENT WITH PRIOR PATIENT RESULTS. THE FOLLOWING DATA WERE PROVIDED: CUSTOMER¿S REFERENCE RANGE (ADULTS): 0.3 ¿ 1.2 MG/DL PATIENT 1 (72-YEAR-OLD FEMALE): (B)(6) 2025: 1.5 MG/DL. PREVIOUS ((B)(6) 2025): 0.5 MG/DL. PATIENT 2 (46-YEAR-OLD FEMALE): (B)(6) 2025: 2.1 MG/DL. PREVIOUS (B)(6) 2025): 0.3 MG/DL. PATIENT 3 (44-YEAR-OLD FEMALE): (B)(6) 2025: 2.6 MG/DL. PREVIOUS ((B)(6) 2025): 0.5 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24038890 ↗
  • MalfunctionDec 4, 2025Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED FALSELY DEPRESSED ALINITY C TOTAL BILIRUBIN RESULTS FOR ONE 83-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6), (B)(6) 2025 BILI T 0.38; REPEATED BILI T 0.45, (B)(6) 2025 BILI T 0.40. THE SAMPLE WAS SENT TO ANOTHER CAC LABORATORY: BILI T 0.50 (COLORIMETRIC METHOD). NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23715758 ↗

Other BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM devices

All BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.