Skip to content

Alinity 08P1021Recalls & reported problems

Abbott Ireland Diagnostics · Alinity i HBsAg Qualitative II Reagent Kit 200 Tests · Class 2

All models of Alinity (14)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740136581), received between May 20, 2022 and Jan 3, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result4

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2024False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED A FALSE REACTIVE ALINITY I HBSAG QUALITATIVE II RESULT FOR A PATIENT IN THE DIALYSIS UNIT. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT A WEEK AGO = NEGATIVE, HEP B VACCINE ADMINISTERED. (B)(6) 2023 SAME PATIENT, NEW SAMPLE. RESULT = POSITIVE FOR BOTH HEP B ANTIGEN AND ANTIBODY. SAMPLE ID (B)(6) INITIAL RESULT = REACTIVE. REPEAT IN DUPLICATE AND BOTH RESULTS = REACTIVE. CONFIRMATORY TEST PERFORMED AND RESULT = REACTIVE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18440583 ↗
  • MalfunctionFeb 20, 2023False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY I HBSAG QUALITATIVE II AND ALINITY I HBSAG QUALITATIVE II CONFIRMATORY RESULTS FOR A PREGNANT FEMALE PATIENT. THE SAMPLE WAS PULLED AND REPEATED 4 DAYS LATER WITH A NONREACTIVE RESULT. A SECOND SAMPLE WAS DRAWN WHICH GENERATED A NONREACTIVE RESULT. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER STATED THEY HAD A SAMPLE THAT WAS DRAWN ON (B)(6) 2023 THAT RAN ON (B)(6) 2023 AND GAVE THE FOLLOWING RESULTS ACCORDING TO CUSTOMER: SID (B)(6) HBSAG=1.04 HBSAG CONFIRMATORY = REACTIVE. REPEAT OF SAMPLE ON (B)(6) 2023 SAMPLE WAS RESPUN AND RERAN IN DUPLICATE ACCORDING TO CUSTOMER: HBSAG=1.19, HBSAG=1.27, HBSAG CONFIRMATORY = 1.11 CONFIRMED POSITIVE. SAMPLE WAS RERUN ON (B)(6) 2023, RESULT = SID (B)(6) HBSAG=0.33 NONREACTIVE. NEW SAMPLE DRAWN (B)(6) 2023 SID= (B)(6) HBSAG = NONREACTIVE HBSAG CONFIRMATORY = NEGATIVE. PER ALINITY I HBSAG QUALITATIVE II PACKAGE INSERT < 1.00 NONREACTIVE NO RETEST REQUIRED. = 1.00 REACTIVE RETEST IN DUPLICATE NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16403034 ↗
  • MalfunctionFeb 20, 2023False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY I HBSAG QUALITATIVE II AND ALINITY I HBSAG QUALITATIVE II CONFIRMATORY RESULTS FOR A PREGNANT FEMALE PATIENT. THE SAMPLE WAS PULLED AND REPEATED 4 DAYS LATER WITH A NONREACTIVE RESULT. A SECOND SAMPLE WAS DRAWN WHICH GENERATED A NONREACTIVE RESULT. THE FOLLOWING DATA WAS PROVIDED: CUSTOMER STATED THEY HAD A SAMPLE THAT WAS DRAWN ON (B)(6) 2023 THAT RAN ON (B)(6) 2023 AND GAVE THE FOLLOWING RESULTS ACCORDING TO CUSTOMER: SID (B)(6) HBSAG=1.04 HBSAG CONFIRMATORY = REACTIVE. REPEAT OF SAMPLE ON (B)(6) 2023 SAMPLE WAS RESPUN AND RERAN IN DUPLICATE ACCORDING TO CUSTOMER: HBSAG=1.19 ; HBSAG=1.27 ; HBSAG CONFIRMATORY = 1.11 CONFIRMED POSITIVE. SAMPLE WAS RERUN ON (B)(6) 2023 RESULT = SID (B)(6) HBSAG=0.33 NONREACTIVE. NEW SAMPLE DRAWN (B)(6) 2023 SID= (B)(6) HBSAG = NONREACTIVE HBSAG CONFIRMATORY = NEGATIVE PER ALINITY I HBSAG QUALITATIVE II PACKAGE INSERT < 1.00 NONREACTIVE NO RETEST REQUIRED. = 1.00 REACTIVE RETEST IN DUPLICATE NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16402197 ↗
  • MalfunctionMay 20, 2022False Positive Result; False Positive Result

    THE CUSTOMER OBSERVED FALSE REACTIVE ALINITY I HBSAG RESULTS GENERATED ON AN ALINITY I PROCESSING MODULE FOR TWO PATIENTS THAT CONFIRMED POSITIVE AND WERE NEGATIVE ON AN OTHER METHOD. THE FOLLOWING INFORMATION WAS PROVIDED: SID (B)(6) INITIAL RESULT = 1.35 S/CO, REPEAT RESULTS = 1.25 S/CO AND 1.33 S/CO; HBSAG CONFIRMATORY RESULTS: C1 = 0.35 S/CO, C2=1.86 S/CO, NEUT% = 94%. SAMPLE WAS SENT TO REFERENCE LAB AND CAME BACK NEGATIVE ON UNKNOWN METHODOLOGY. SID (B)(6) INITIAL RESULT = 2.63 S/CO, REPEAT RESULTS = 2.74 S/CO AND 2.72 S/CO; HBSAG CONFIRMATORY RESULTS: C2 = 2.31 S/CO, NEUT% = 97% NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14458054 ↗

FDA clearance and classification

The FDA regulates product code LOM (test, hepatitis b (b core, be antigen, be antibody, b core igm)) under 21 CFR 866.3172 as a Class II device. 21 CFR 866.3172 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: In vitro diagnostic device intended for aid in the diagnosis of chronic or acute HBV infection. HBV surface antigen (HbsAg) is also used for screening of HBV infection in pregnant women to identify neonates who are at risk of acquiring hepatitis B during perinatal period.

Common questions about the Alinity 08P1021

Has the Alinity 08P1021 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00380740136581), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Alinity 08P1021?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between May 20, 2022 and Jan 3, 2024. In those reports, the most common specific device problem is “False Positive Result” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Alinity 08P1021?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Alinity 08P1021, and is it still sold?

The FDA's device database lists Abbott Ireland Diagnostics as the company responsible for it (brand Alinity, model 08P1021). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices

All Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page