Alinity 07P8421Recalls & reported problems

Abbott GmbH · Alinity i Anti-HBc Reagent Kit 200 Tests · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740136659), received between Apr 2, 2026 and Apr 2, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “False Negative Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2026False Negative Result

    THE CUSTOMER OBSERVED FALSE NEGATIVE ALINITY I ANTI-HBC RESULTS. ON (B)(6) 2026, SID (B)(6) (MALE, 69 YEARS OLD) WAS 0.95 S/CO GRAYZONE, REPEATED 0.30 S/CO NONREACTIVE, 0.23 S/CO NONREACTIVE, 0.23 S/CO NONREACTIVE. THE SAME SAMPLE WAS REPEATED ON SERIAL (B)(6) WITH THE SAME REAGENT LOT 76110BE00. ON (B)(6) 2026 RESULT WAS 1.65 S/CO REACTIVE, 1.44 S/CO REACTIVE, 1.51 S/CO REACTIVE. IT IS UNKNOWN WHAT RESULT WAS REPORTED OUT OF LAB. REFERENCE LAB RESULT CAME BACK AS NONREACTIVE. ADDITIONALLY, THE CUSTOMER PULLED SAMPLES THAT ALREADY RESULTED FROM THIS INSTRUMENT AND RAN THEM ON SERIAL (B)(6), WITH DIFFERENT RESULTS FROM 2 SAMPLES. ON (B)(6) 2026, SID (B)(6) WAS 2.58 S/CO REACTIVE, REPEAT 0.69 NONREACTIVE, 2ND REPEAT 4.87 REACTIVE. SID (B)(6) WAS REPORTED AS REACTIVE, BUT ON (B)(6) 2026 REPEATED ON SERIAL (B)(6) RESULT OF 0.22 S/CO NONREACTIVE. ON (B)(6) 2026, SID (B)(6) WAS 1.43, 1.99 S/CO REACTIVE. SID (B)(6) WAS REPORTED AS REACTIVE BUT ON (B)(6) 2026 REPEATED ON SERIAL (B)(6) RESULT OF 0.74 S/CO NONREACTIVE. NORMAL REFERENCE RANGE OF <0.80 S/CO NONREACTIVE, GRAYZONE OF 0.80 TO <1.21 S/CO, REACTIVE OF = 1.21 S/CO. NO ADDITIONAL PATIENT INFORMATION WAS PROVIDED. NO IMPACT TO PATIENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24772037 ↗