Skip to content

Alinity 09P6901Recalls & reported problems

Abbott Laboratories · Alinity hs SLIDE MAKER STAINER MODULE · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00380740138868), received between Aug 27, 2024 and Sep 1, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Low Test Results1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 1, 2026Sparking

    The customer observed a spark from a wire behind the stain baths of the Alinity HS Processing Module while removing the baths for cleaning that occurred overnight. After replacing the baths, the customer observed the spark again from the same wire.No impact to patient management and no impact to user safety was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50541584 ↗
  • MalfunctionAug 27, 2024Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED NO BLAST FOUND FROM CELLAVISION FOR ONE KNOWN BLAST CELL PATIENT ON DIFFERENTIAL PREPARED BY THE ALINITY HS PROCESSING MODULE. THE CBC RESULTS GENERATED FROM ALINITY HQ PROCESSING MODULE WITH BLAST FLAGGING WHICH MATCHED THE PATIENT HISTORY. THE PATHOLOGIST QUESTIONED THE DIFFERENT RESULTS DUE TO NOT SEEN BLAST WITH KNOWN BLAST PATIENT. THE CUSTOMER STATED SMEAR HAS NO BLAST CELL DURING READING FROM CELLAVISION MICROSCOPE. THE CUSTOMER NOTICED SPECIMEN DROPPED INSIDE THE INSTRUMENT. THE ALINITY HS PROCESSING MODULE DID NOT GENERATE ANY FLAG FOR ERROR MESSAGE. THE CUSTOMER MADE MANUAL DIFFERENTIAL AND THAT CONFIRMED BLAST CELLS IN MANUAL PREPARED SMEAR. THE CUSTOMER AGREES WITH MANUAL DIFFERENTIAL RESULTS. THE BARCODE ID IS CORRECT ON SLIDE THAT MADE BY THE ALINITY HS PROCESSING MODULE. THE RESULTS PROVIDED WERE: (B)(6) 2024 SID (B)(6) WBC INITIAL=30.1 10E3/UL; HEMOGLOBIN INITIAL=7.56 G/DL; PLATELET INITIAL=21.2 10E3/UL; DIFFERENTIAL WITH LEFT SHIFT, VAR LYM AND BLAST FLAGS /NO REPEAT RESULTS PROVIDED. CELLAVISION= NO BLAST CELL IDENTIFIED. MANUAL DIFFERENTIAL=BLAST=37 THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20075336 ↗

Other Slide stainer, automated devices

All Slide stainer, automated devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.