Skip to content

UniCel DxH SMSRecalls & reported problems

Beckman Coulter, Inc. · UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System · Class 1

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15099590192679), received between Aug 23, 2019 and Mar 11, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Output Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Output Problem2
  • Device Handling Problem1

Patient problems

  • Test Result2
  • Exposure to Body Fluids1
  • Muscle/Tendon Damage1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2026Device Handling Problem

    DURING REMOVAL OF THE SAMPLE TRANSPORT MODULE (STM) ON THE UNICEL DXH SLIDEMAKER STAINER (PRODUCT NUMBER: 775222, SERIAL NUMBER: (B)(6) , THE MODULE BECAME LODGED ON THE GUIDE RAIL. THE BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) ATTEMPTED TO RESET THE STM, AT WHICH POINT IT UNEXPECTEDLY DISENGAGED AND BEGAN TO FALL. THE FSE INTERVENED TO PREVENT THE MODULE FROM DROPPING, RESULTING IN A MUSCLE STRAIN TO THE LEFT SHOULDER AND UPPER BACK. THE INJURY WAS NOT IMMEDIATELY APPARENT AT THE TIME OF THE INCIDENT BUT PROGRESSIVELY WORSENED DURING THE FSE¿S DRIVE HOME. THE FSE REPORTED THE INCIDENT TO THEIR SUPERVISOR AND CONTACTED THE MEDICAL CALL CENTER TO SEEK MEDICAL EVALUATION AND GUIDANCE.

    Patient problems: Muscle/Tendon Damage

    MAUDE report 24574846 ↗
  • InjuryOct 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE BEC EMPLOYEE (FIELD SERVICE ENGINEER, FSE) REPORTED THAT WHILE TROUBLESHOOTING ON DXH SMS, HE SUFFERED A LOWER BACK INJURY WHEN REINSTALLING THE DXH SMS SAMPLE TRANSPORT MODULE (STM). THE FSE STATED HE IMMEDIATELY NOTIFIED HIS MANAGER AND CALLED THE EMPLOYEE NURSE HOTLINE. THE HOTLINE SCHEDULED A DOCTOR'S APPOINTMENT FOR NEXT DAY ON ((B)(6) 2023). THE FSE STATED HE WAS ADMINISTERED A TORADOL INJECTION AND PRESCRIBED METHYL PREDNISOLONE, AND THE DOCTOR REQUESTED AN MRI TO EVALUATE THE SEVERITY OF THE CONDITION. THE FSE WAS OUT FOR TWO WEEKS ON WORKMAN'S COMPENSATION.

    Patient problems: Pain

    MAUDE report 17908912 ↗
  • InjurySep 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED SHE SCRATCHED HER RIGHT FOREARM WHILE TRYING TO REMOVE A SLIDE THAT FELL INSIDE THEIR DXH SLIDEMAKER STAINER (SMS) INSTRUMENT. THE WOUND ON THE FOREARM WAS A SUPERFICIAL SCRATCH THAT DID NOT REQUIRE MEDICAL INTERVENTION OR ANY WORK RESTRICTIONS. THE OPERATOR WAS ADMINISTERED RETROVIRALS AS A PREVENTATIVE MEASURE AND WAS TESTED FOR HIV AND OTHER VIRUSES ON THE DAY OF EXPOSURE. THE OPERATOR WAS ALSO SCHEDULED FOR A 30-DAY FOLLOW-UP.

    Patient problems: Exposure to Body Fluids

    MAUDE report 15459288 ↗
  • InjuryMar 5, 2020Output Problem; Output Problem

    THE CUSTOMER REPORTED THAT A PATIENT SAMPLE (PATIENT (B)(6)) SHOWED "VARIANTLY' ON THE RESULTS FROM THEIR UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM WAS RUN ON THE DXH SMS FOR SLIDE ASSESSMENT. THE SLIDE RESULT REPORTED 11% BLASTS. FIVE DAYS LATER (B)(6) 2020 A BONE MARROW BIOPSY WAS PERFORMED AND WAS NEGATIVE. THIS WAS FOR A FOLLOW UP FOR A PATIENT WITH MULTIPLE MYELOMA (TYPE IGG KAPPA) IN REMISSION. PATIENT WAS CLINICALLY ASYMPTOMATIC. PATIENT (B)(6) RESULTS FROM THE DXH INSTRUMENT SHOWED VARIANT LYMPHOCYTIC POPULATION WHICH TRIGGERED THE SLIDE ORDERED ON THE DXH SMS. THE SLIDE SHOWED 11% BLAST. ON THE SAME DAY, THE SAMPLE WAS RUN THROUGH A FLOW CYTOMETRY INSTRUMENT AND THE EXAMINATION OF THE SAMPLE WAS NEGATIVE FOR BLAST. AS A RESULT OF THE POSITIVE SLIDE RESULTS, THE DOCTOR PERFORMED A BONE MARROW BIOPSY WHICH WAS NEGATIVE FOR BLASTS.

    Patient problems: Test Result

    MAUDE report 9795964 ↗
  • InjuryAug 23, 2019Output Problem; Output Problem

    THE CUSTOMER REPORTED THAT THEIR UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM HAD GENERATED WBC CARRYOVER AND VARIABLE LYMPH MESSAGES WHILE RUNNING A PATIENT SAMPLE. THIS PROMPTED A MANUAL SLIDE ASSESSMENT FOR THE PATIENT AND AS A RESULT OF THE SLIDE, THE PATIENT WAS THOUGHT TO HAVE RELAPSED AND A BONE MARROW DRAWING WAS DONE. THE SAMPLE WAS FOR A (B)(6) YEAR OLD FEMALE PATIENT WITH ACUTE MYELOID LEUKEMIA (AML). IT WAS CONFIRMED THAT THE PATIENT HAD NOT RELAPSED AS INITIALLY SUSPECTED. THE PATIENT RECOVERED FROM THE PROCEDURE WITHOUT COMPLICATION. UPON FURTHER INVESTIGATION, IT WAS FOUND THAT THERE NO MALFUNCTION OF THE DXH800 AS ORIGINALLY REPORTED. THE MALFUNCTION WAS IDENTIFIED TO HAVE TAKEN PLACE ON THE DXH SLIDEMAKER STAINER. THE EVENT THAT TRIGGERED THE LAB¿S RECOGNITION OF THE ISSUE WAS NEITHER THE RECIPIENT BONE MARROW BIOPSY RESULTS NOR A SECOND CARRYOVER EVENT, BUT RATHER TESTING OF A NEW SAMPLE FROM THE RECIPIENT PATIENT ON (B)(6). HER RESULTS FROM THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM WERE CONSISTENT WITH RESULTS FROM (B)(6), BUT HER PERIPHERAL SMEAR PREPARED ON THE SMS DISCLOSED NO BLASTS. A SENIOR TECH AT THE LABORATORY RECOGNIZED THE SIGNIFICANCE OF

    Patient problems: Test Result

    MAUDE report 8927478 ↗

Other Slide Stainer, Automated devices

All Slide Stainer, Automated devices →