ARCHITECT 04S9520Recalls & reported problems
Abbott Ireland · Creatinine2 3600 Tests · Class 2
All models of ARCHITECT (38)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740159535), received between May 2, 2024 and Jan 9, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Low Test Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2026High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED CREATININE2 RESULTS GENERATED BY THE ARCHITECT C8000 PROCESSING MODULE FOR A PATIENT. A NEW BLOOD SAMPLE WAS DRAWN AND YIELD A NORMAL RESULT. THE FOLLOWING DATA WAS PROVIDED: NORMAL REFERENCE RANGES: CREATININE2: 0.60 TO 1.30 MG/DL. SID (B)(6): 29DEC2025 CREATININE2 INITIAL RESULT = 7.84 MG/DL, REPEAT RESULTS ((B)(6) 2025) = 2.67 AND 2.69 MG/DL. SID (B)(6) (NEW BLOOD DRAW FOR THE SAME PATIENT): (B)(6) 2025 CREATININE2 RESULT = 0.68 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24019256 ↗MalfunctionSep 19, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CREATININE2 ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT. THE PATIENT WAS TAKING MEDICATIONS NOVALGINA AND DECADRON. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE, CREATININE 0.60 MG/DL - 1.10 MG/DL): (B)(6) 2024, SID (B)(6), INITIAL CREATININE RESULT (ANALYZER (B)(6) )= 10.67 MG/DL; REPEAT RESULT (ANALYZER (B)(6) )= 10.36 MG/DL; FROM A SECOND BLOOD COLLECTION, SID (B)(6), CREATININE RESULT (ANALYZER (B)(6) )= 0.69 MG/DL; FROM A THIRD BLOOD COLLECTION, SID (B)(6), CREATININE RESULT (ANALYZER (B)(6) )= 0.73 MG/DL. THE CUSTOMER REPORTED THAT THE INITIAL CREATININE RESULTS WERE FROM BLOOD DRAWN FROM THE SAME LIMB THAT IV MEDICATION WAS BEING ADMINISTERED. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20255977 ↗MalfunctionJun 19, 2024Low Test Results; Low Test Results
THE CUSTOMER REPORTED A DISCREPANT CREATININE2 RESULT GENERATED ON THE ARCHITECT C4000 ANALYZER ON A PATIENT. RESULTS PROVIDED: PT 4 (56-YR OLD FEMALE) = 1.88 / 5.02 / 4.02 / 4.96 / 4.98 / 5.09 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19565762 ↗MalfunctionMay 2, 2024Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DEPRESSED CREATININE2 RESULT GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE SAMPLE THAT WAS NOT REPORTED OUT OF THE LABORATORY. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6), INITIAL RESULT = 0.59 MG/D, REPEAT = 2.67 MG/DL . THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19231634 ↗
Other ALKALINE PICRATE, COLORIMETRY, CREATININE devices
- Dimension® Flex® reagent cartridge CRE2 10872079Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE1 recall9 reports
- ABX PENTRA CREATININE 120 CP 1220001933Horiba Abx Sas · Alkaline Picrate, Colorimetry, Creatinine1 recall0 reports
- ARCHITECT 3L81-22Abbott Laboratories · ALKALINE PICRATE, COLORIMETRY, CREATININE31 reports
- Alinity 07P9920Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE24 reports
- Atellica CH Crea_2 11097596Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE19 reports
- Creatinine Jaffe Gen.2 08057532190Roche Diagnostics GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE17 reports
- Alinity 04T9120Abbott Ireland · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
- Alinity 07P9930Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page