ARCHITECT 04S9520Recalls & reported problems

Abbott Ireland · Creatinine2 3600 Tests · Class 2

All models of ARCHITECT (38)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740159535), received between May 2, 2024 and Jan 9, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Low Test Results2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2026High Test Results

    THE CUSTOMER REPORTED FALSELY ELEVATED CREATININE2 RESULTS GENERATED BY THE ARCHITECT C8000 PROCESSING MODULE FOR A PATIENT. A NEW BLOOD SAMPLE WAS DRAWN AND YIELD A NORMAL RESULT. THE FOLLOWING DATA WAS PROVIDED: NORMAL REFERENCE RANGES: CREATININE2: 0.60 TO 1.30 MG/DL. SID (B)(6): 29DEC2025 CREATININE2 INITIAL RESULT = 7.84 MG/DL, REPEAT RESULTS ((B)(6) 2025) = 2.67 AND 2.69 MG/DL. SID (B)(6) (NEW BLOOD DRAW FOR THE SAME PATIENT): (B)(6) 2025 CREATININE2 RESULT = 0.68 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24019256 ↗
  • MalfunctionSep 19, 2024High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED CREATININE2 ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT. THE PATIENT WAS TAKING MEDICATIONS NOVALGINA AND DECADRON. THE FOLLOWING RESULTS WERE PROVIDED (REFERENCE RANGE, CREATININE 0.60 MG/DL - 1.10 MG/DL): (B)(6) 2024, SID (B)(6), INITIAL CREATININE RESULT (ANALYZER (B)(6) )= 10.67 MG/DL; REPEAT RESULT (ANALYZER (B)(6) )= 10.36 MG/DL; FROM A SECOND BLOOD COLLECTION, SID (B)(6), CREATININE RESULT (ANALYZER (B)(6) )= 0.69 MG/DL; FROM A THIRD BLOOD COLLECTION, SID (B)(6), CREATININE RESULT (ANALYZER (B)(6) )= 0.73 MG/DL. THE CUSTOMER REPORTED THAT THE INITIAL CREATININE RESULTS WERE FROM BLOOD DRAWN FROM THE SAME LIMB THAT IV MEDICATION WAS BEING ADMINISTERED. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20255977 ↗
  • MalfunctionJun 19, 2024Low Test Results; Low Test Results

    THE CUSTOMER REPORTED A DISCREPANT CREATININE2 RESULT GENERATED ON THE ARCHITECT C4000 ANALYZER ON A PATIENT. RESULTS PROVIDED: PT 4 (56-YR OLD FEMALE) = 1.88 / 5.02 / 4.02 / 4.96 / 4.98 / 5.09 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19565762 ↗
  • MalfunctionMay 2, 2024Low Test Results; Low Test Results

    THE CUSTOMER OBSERVED A FALSELY DEPRESSED CREATININE2 RESULT GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE SAMPLE THAT WAS NOT REPORTED OUT OF THE LABORATORY. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6), INITIAL RESULT = 0.59 MG/D, REPEAT = 2.67 MG/DL . THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19231634 ↗

Other ALKALINE PICRATE, COLORIMETRY, CREATININE devices

All ALKALINE PICRATE, COLORIMETRY, CREATININE devices →