ARCHITECT 3L80-32Recalls & reported problems
Abbott GmbH · Carbon Dioxide · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00380740161538), received between Sep 1, 2020 and Jul 21, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results3
- High Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 21, 2025Low Test Results; Low Test Results
THE CUSTOMER OBSERVED FALSELY DEPRESSED CARBON DIOXIDE RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2025, SID: (B)(6): INITIAL RESULT = 9.2 MMOL/L, REPEAT = 9.9 MMOL/L, VITROS RESULT = 18 MMOL/L, RADIOMETER RESULT = 23 MMOL/L. THE CUSTOMER INDICATED THE SAMPLE IS LIPEMIC. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22553539 ↗MalfunctionFeb 27, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CO2 RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE FOR ONE SAMPLE. THE CUSTOMER REPEATED THE SAMPLE ON ANOTHER ARCHITECT AND THE RESULTS WERE LOWER. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT = 32 MMOL/L REPEAT RESULTS= 25 AND 26 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18789983 ↗MalfunctionApr 28, 2023Low Test Results; Low Test Results
THE CUSTOMER OBSERVED A FALSELY DEPRESSED CARBON DIOXIDE RESULT GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE FOR ONE SAMPLE. RESULTS PROVIDED: INITIAL RESULT = 6.1 MMOL/L, REPEAT RESULT = 22.9 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16831177 ↗MalfunctionSep 1, 2020Low Test Results; Low Test Results
THE CUSTOMER REPORTED FALSELY DEPRESSED ARCHITECT CO2 RESULTS FOR FOUR PATIENTS. THE CUSTOMER ALLEGES HAVING TO ISSUE ABOUT 500 CORRECTED CO2 REPORTS. HOWEVER, NO DATA IS AVAILABLE FOR THIS ALLEGATION. THE CUSTOMER DID PROVIDE INFORMATION FOR FOUR PATIENTS: PT#1 (B)(6): INITIAL = 16 MMOL/L, REPEAT = 22 MMOL/L; PT# 2: (B)(6)INITIAL = 20 MMOL/L, REPEAT = 27 MMOL/L; PT#3: (B)(6): INITIAL = 14 MMOL/L, REPEAT = 25 MMOL/L; PT#4: (B)(6): INITIAL = 16 MMOL/L, REPEAT = 25 MMOL/L (NORMAL RANGE: 20-29 MMOL/L). THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10476215 ↗
Other ENZYMATIC, CARBON-DIOXIDE devices
- AU OSR6237Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls1 report
- AU OSR6137Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls0 reports
- AU OSR6637Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls0 reports
- Alinity 07P7220Abbott GmbH · ENZYMATIC, CARBON-DIOXIDE90 reports
- VITROS 8262396Ortho-Clinical Diagnostics, Inc. · ENZYMATIC, CARBON-DIOXIDE1 recall0 reports
- ARCHITECT 3L80-22Abbott GmbH · ENZYMATIC, CARBON-DIOXIDE31 reports
- Atellica CH CO2_c 11097521Siemens Healthcare Diagnostics Inc. · ENZYMATIC, CARBON-DIOXIDE21 reports
- Dimension Vista® Flex® reagent cartridge CO2 10445163Siemens Healthcare Diagnostics Inc. · ENZYMATIC, CARBON-DIOXIDE15 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.