ARCHITECT 3L81-23Recalls & reported problems
Abbott GmbH · Creatinine · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00380740161552), received between Apr 29, 2021 and May 14, 2024 — classified as 8 malfunctions.
The most frequently reported specific device problem is “High Test Results” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results8
Patient problems
- No Clinical Signs, Symptoms or Conditions8
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2024High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CREATININE RESULTS GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT SAMPLE. THE FOLLOWING DATA WAS PROVIDED: TESTED ON (B)(6) 2024 (CREATININE NORMAL RANGE: 0.5-1.5 MG/DL). SID (B)(6): CREATININE INITIAL RESULT = 2.125 MG/DL; CREATININE REPEAT RESULT = 0.800 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19315569 ↗MalfunctionNov 16, 2022High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY ELEVATED CREATININE RESULT FOR A PATIENT SAMPLE WHILE RUNNING ON THE ARCHTIECT C4000 PROCESSING MODULE. THE PATIENT WAS TRANSPORTED TO THE EMERGENCY ROOM DUE TO THE ELEVATED CREATININE RESULT, HOWEVER NO KNOWN TREATMENT WAS PERFORMED BASED ON THE RESULT PROVIDED. LAB¿S REFERENCE RANGE FOR CREATININE: 0.7-1.25 MG/DL. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2022, SID (B)(6): INITIAL RESULT: 7.26 MMG/DL. RERUN ON THE SAME INSTRUMENT: 0.79, 0.80 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15810879 ↗MalfunctionMay 20, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CREATININE RESULTS GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) INITIAL AND REPEAT RESULTS = 7.76 MG/DL, A NEW SAMPLE WAS COLLECTED WHICH GENERATED AN INITIAL RESULT OF 0.854 MG/DL AND REPEAT RESULT OF 0.887 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14463356 ↗MalfunctionMay 11, 2022High Test Results; High Test Results
THE CUSTOMER REPORTED A FALSELY INCREASED CREATININE RESULT GENERATED ON THE ARCHITECT C8000 ANALYZER ON A PATIENT. RESULTS PROVIDED: 15APR2022 SID (B)(4) (NORMAL RANGE: 0.6-1.3 MG/DL), LOT# 09277UN21. SERUM SAMPLE: SN# (B)(4) = 2.9 / 2.9 MG/DL, SN# (B)(4) = 2.9 / 2.9 MG/DL, SN# (B)(4) = 2.8 MG/DL, SN# (B)(4) = 2.8 MG/DL. PREVIOUS RESULT FOR THIS PATIENT = 0.9 MG/DL. MAYO LABS (ROCHE PLATFORM) = 0.9 MG/DL NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14358009 ↗MalfunctionMay 5, 2022High Test Results; High Test Results
THE CUSTOMER QUESTIONED CREATININE RESULTS GENERATED WITH THE ARCHITECT C4000 PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING WAS PROVIDED BY THE CUSTOMER. SID (B)(6) RESULT 3.18 RERUN 0.84 MG/DL. CUSTOMER HAD A RENAL PANEL UNDER A DIFFERENT SID (SID (B)(6)) AND RESULT WAS CLOSE TO THE REPEATED RESULT. CUSTOMER STATED THE RESULT WAS CORRECTED TO 0.84 WHICH MATCHES PATIENT HISTORY. CUSTOMER NORMAL RANGE 0.70-7.20 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14290919 ↗MalfunctionApr 1, 2022High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED ARCHITECT CREATININE RESULTS WHICH QUESTIONED BY PHYSICIAN ON ONE FEMALE PATIENT. THE RESULTS PROVIDED WERE: ON (B)(6) 2022 SID (B)(6) INITIAL=2.99 MG/DL /REPEATED=0.77 MG/DL /REDRAWN AND REPEATED SID (B)(6) = 0.73 MG/DL . LABORATORY REFERENCE RANGE FOR CREATININE=0.57 TO 1.11 MG/DL THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13978761 ↗MalfunctionJul 20, 2021High Test Results; High Test Results
THE CUSTOMER OBSERVED FALSELY ELEVATED CREATININE RESULTS GENERATED ON THE ARCHITECT C8000 PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED (REFERENCE RANGE: 53 TO 100 UMOL/L): (B)(6) 2021 SID (B)(6) INITIAL RESULT = 263.8 UMOL/L, REPEAT = 263.8 UMOL/L, REPEAT ON ANOTHER ARCHITECT = 264.1 UMOL/L, ROCHE = 112.6 UMOL/L, NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12195815 ↗MalfunctionApr 29, 2021High Test Results; High Test Results
THE CUSTOMER REPORTED FALSELY ELEVATED CREATININE RESULTS ON ONE PATIENT GENERATED ON THE ARCHITECT ANALYZER. THE RESULTS PROVIDED WERE: (B)(6). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11747992 ↗
Other ALKALINE PICRATE, COLORIMETRY, CREATININE devices
- Dimension® Flex® reagent cartridge CRE2 10872079Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE1 recall9 reports
- ABX PENTRA CREATININE 120 CP 1220001933Horiba Abx Sas · Alkaline Picrate, Colorimetry, Creatinine1 recall0 reports
- ARCHITECT 3L81-22Abbott Laboratories · ALKALINE PICRATE, COLORIMETRY, CREATININE31 reports
- Alinity 07P9920Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE24 reports
- Atellica CH Crea_2 11097596Siemens Healthcare Diagnostics Inc. · ALKALINE PICRATE, COLORIMETRY, CREATININE19 reports
- Creatinine Jaffe Gen.2 08057532190Roche Diagnostics GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE17 reports
- Alinity 07P9930Abbott GmbH · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
- Alinity 04T9120Abbott Ireland · ALKALINE PICRATE, COLORIMETRY, CREATININE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.