ARCHITECT 03P6834Recalls & reported problems

Abbott Ireland · CC MAGNESIUM 3750 Tests · Class 1

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00380740169862), received between Dec 28, 2022 and Jun 2, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “High Test Results” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results5
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 2, 2025High Test Results; High Test Results

    THE CUSTOMER REPORTED A FALSELY ELEVATED MAGNESIUM RESULT GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE ON A PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6) = 3.21 / 0.86 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22121915 ↗
  • MalfunctionMay 19, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER OBSERVED IMPRECISE MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE: 0.61-1.03 MMOL/L): INITIAL RESULT = 1.02 MMOL/L; REPEAT RESULT = 0.52 MMOL/L THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22047642 ↗
  • MalfunctionNov 4, 2024High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED MAGNESIUM RESULT GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR ONE SAMPLE. THE FOLLOWING RESULTS WERE PROVIDED: (CUSTOMER PROVIDED NORMAL RANGE FOR MG+ 1.6-2.6 MG/DL) (B)(6) 2024 INITIAL RESULT = 6.2 MG/DL, REPEAT RESULT ON ANOTHER INSTRUMENT = 1.8 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20599226 ↗
  • MalfunctionMay 22, 2024High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED MAGNESIUM RESULT FOR ONE PATIENT ON AN ARCHITECT C16000. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS 1.6-2.6 MG/DL): SAMPLE ID (B)(6) INITIAL RESULT WAS 8.2, REPEAT WAS 2.1 MG/DL. THE SAMPLE WAS REPEATED FIVE TIMES FOR A PRECISION STUDY AND THE RESULTS WERE ALL 2.1 MG/DL. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19373172 ↗
  • MalfunctionJul 21, 2023High Test Results; High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C16000 PROCESSING MODULE FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: (B)(6) INITIAL RESULT = 3.165 MMOL/L, REPEAT = 0.924 MMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17373628 ↗
  • MalfunctionDec 28, 2022High Test Results; High Test Results

    THE CUSTOMER OBSERVED FALSELY ELEVATED MAGNESIUM RESULTS GENERATED ON THE ARCHITECT C4000 PROCESSING MODULE FOR MULTIPLE SAMPLES. THE FOLLOWING EXAMPLE RESULTS WERE PROVIDED: ON (B)(6) 2022. INITIAL RESULT = 7.9MGDL, REPEAT RESULT ON ANOTHER ANALYZER= 1.3MG/DL. ON (B)(6) 2022. INITIAL RESULT = 6MG/DL, REPEAT RESULT ON ANOTHER ANALYZER= 2.1MG/DL REFERENCE (NORMAL) RANGE: 1.6 TO 2.6 MG/DL. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16059207 ↗

Other PHOTOMETRIC METHOD, MAGNESIUM devices

All PHOTOMETRIC METHOD, MAGNESIUM devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.