Alinity 09P6324Recalls & reported problems

Abbott Ireland · Immunoglobulin M 360 Tests · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740169916), received between Apr 9, 2024 and Apr 9, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Installation-Related Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Installation-Related Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 9, 2024Installation-Related Problem; Installation-Related Problem

    THE CUSTOMER REPORTED A FALSELY DECREASED IMMUNOGLOBULIN M ON THE ALINITY C PROCESSING MODULE FOR ONE PATIENT. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6) INITIAL RUN (STANDARD 1:5 DILUTION) = >16.9 G/L, AUTOMATIC RERUN OF UNDILUTED SAMPLE = 1.045 G/L; RERUN WITH 1:20 MANUAL DILUTION = 40.724 G/L CONFIGURATION OF THE ASSAY IN THE ALINITY C SOFTWARE WAS PERFORMED IN AUGUST 2023. IF THE ANALYZER IS CONFIGURED AS G/L THE RANGE WOULD BE 0-0.25 AND IF CONFIGURED AS MG/DL THE RANGE WOULD BE 0-25. THE UNITS OF MEASURE WERE ENTERED AS G/L, BUT THE RANGE WAS NOT UPDATED TO 0-0.25 G/L. THE ANALYZER CONFIGURATION OF THE RERUN RULE WAS TO PERFORM AN AUTOMATIC RERUN OF AN UNDILUTED SAMPLE IF THE VALUE FELL WITHIN THE RANGE CONFIGURED IN THE SOFTWARE. SINCE THE ALINITY SOFTWARE WAS CONFIGURED WITH A RANGE OF 0-25 G/L, THE AUTOMATIC RERUN PROTOCOL WAS INCORRECT, AND THE SAMPLE WAS REPEATED AS AN UNDILUTED SAMPLE. THE SAMPLE NEEDED TO BE DILUTED PRIOR TO RERUNNING. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070813 ↗

Other IGM, ANTIGEN, ANTISERUM, CONTROL devices

All IGM, ANTIGEN, ANTISERUM, CONTROL devices →