Skip to content

Alinity 09P2635Recalls & reported problems

Abbott GmbH · Alinity i Active B-12 (Holotranscobalamin) Reagent Kit · Class 2

All models of Alinity (37)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00380740189815), received between May 11, 2026 and May 11, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026High Test Results

    THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I ACTIVE B12 RESULT GENERATED ON AN ALINITY I PROCESSING MODULE FOR ONE PATIENT SAMPLE. SID (B)(6) RESULT ON LOT 12569UP00 = 29.7 PMOL/L, RESULT ON OTHER LOT = 18.1 PMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25136873 ↗

Other Radioassay, vitamin b12 devices

All Radioassay, vitamin b12 devices →

Sources and updates