Skip to content

CardioChek® Plus Professional Analyzer, Refurbished 2702Recalls & reported problems

Polymer Technology Systems, Inc. · The CardioChek® Plus professional analyzer is intended for in vitro diagnostic use to test whole blood. The specific analyte measured is determined by the type of test strip chosen. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00381932702010), received between Aug 25, 2020 and Jun 14, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (5 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 2, 2019Class IITerminated

    One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

    Z-0833-2020 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Electrical /Electronic Property Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 14, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED THAT THE BATTERIES IN THEIR CARDIOCHEK PLUS ANALYZER "GOT VERY HOT AND MELTED". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19536414 ↗
  • MalfunctionApr 7, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER PROVIDED A LOG SHEET OF 6 ANALYZERS WITH THE SAME COMPLAINT DESCRIPTION. ALL 6 ANALYZERS HAVE THE SAME ALLEGATION, "THE ANALYZER GOT VERY HOT AND NOW THERE IS CORROSION". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM FOR ANY OF THE REPORTED INCIDENTS. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS. EACH ANALYZER WILL RECEIVE ITS OWN MEDWATCH FORM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16704322 ↗
  • MalfunctionApr 7, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER PROVIDED A LOG SHEET OF 6 ANALYZERS WITH THE SAME COMPLAINT DESCRIPTION. ALL 6 ANALYZERS HAVE THE SAME ALLEGATION, "THE ANALYZER GOT VERY HOT AND NOW THERE IS CORROSION". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM FOR ANY OF THE REPORTED INCIDENTS. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS. EACH ANALYZER WILL RECEIVE ITS OWN MEDWATCH FORM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16704317 ↗
  • MalfunctionMar 11, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED THAT THE ANALYZER STARTED OVERHEATING UPON REPLACING THE BATTERIES. THERE WAS NO ALLEGATION OF PATIENT/USER HARM. THERE WAS NO ALLEGATION OF INCORRECT RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11461895 ↗
  • MalfunctionAug 25, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED WHEN RUNNING CONTROLS THAT THE ANALYZER STARTED TO "GET HOT". THE CUSTOMER OPENED THE BATTERY COMPARTMENT TO REMOVE THE BATTERIES AND STATED THEY WERE "REALLY HOT". THERE WAS NO ALLEGATION OF PATIENT/USER HARM. THERE WAS NO ALLEGATION OF INCORRECT RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10447976 ↗

Other LDL & VLDL PRECIPITATION, HDL devices

All LDL & VLDL PRECIPITATION, HDL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page