LDL-Cholesterol Gen.3 08057966190Recalls & reported problems
Roche Diagnostics GmbH · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336121375), received between Oct 24, 2023 and Sep 28, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results3
- Low Test Results2
- Chemical Problem1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Non Reproducible Results
There was an allegation of questionable LDL-Cholesterol Gen.3 results from the cobas c 503 analytical unit. One example was provided of an initial result of 11.3 mg/dL, and a repeat result of 122 mg/dL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50914631 ↗MalfunctionNov 26, 2025Low Test Results; Low Test Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE LDL-CHOLESTEROL GEN.3 RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS C 503 ANALYTICAL UNIT. ON (B)(6) 2025: THE INITIAL RESULT FROM THE MODULE WAS 1.39 MMOL/L. THE REPEAT RESULT AT 1:2 DILUTION WAS 0.956 MMOL/L WITH A FINAL CALCULATED DILUTION RESULT OF 1.912 MMOL/L. THE REPEAT RESULT AT 1:10 DILUTION WAS 0.173 MMOL/L WITH A FINAL CALCULATED DILUTION RESULT OF 1.73 MMOL/L. ON (B)(6) 2025: THE REPEAT RESULT FROM THE MODULE WAS 1.96 MMOL/L. ON (B)(6) 2025: THE REPEAT RESULT FROM AN ARCHITECT ANALYZER WAS 4.44 MMOL/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23655738 ↗MalfunctionOct 29, 2025Non Reproducible Results; Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH LDL-CHOLESTEROL GEN. 3 ON A COBAS C 503 ANALYTICAL UNIT. THE SAMPLE INITIALLY RESULTED IN AN LDL VALUE OF 5.00 MMOL/L WITH A DATA FLAG. THE SAMPLE WAS AUTOMATICALLY REPEATED, RESULTING IN AN LDL VALUE OF 5.10 MMOL/L WITH A DATA FLAG. THESE VALUES WERE HELD AND NOT RELEASED TO THE PATIENT. THE SAMPLE WAS REPEATED AGAIN, RESULTING IN AN LDL VALUE OF 1.49 MMOL/L WHICH WAS THEN RELEASED TO THE PATIENT. THE SAMPLE WAS REPEATED AFTERWARDS AND RESULTED IN A VALUE OF 5.07 MMOL/L WITH A DATA FLAG. AN AMENDED REPORT WAS ISSUED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23416805 ↗MalfunctionNov 26, 2024Chemical Problem; Non Reproducible Results; Chemical Problem; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE LDLC3- (LOW DENSITY, LIPOPROTEIN) RESULTS FROM THE COBAS C 503 ANALYTICAL UNIT. PATIENT 1: THE INITIAL RESULT WAS 26 MG/DL. THE SAME SAMPLE WAS REPEATED ON ANOTHER PRO ANALYZER WITH A RESULT OF 27 MG/DL AND ON A C311 ANALYZER WITH A RESULT OF 22 MG/DL. THE SAME SAMPLE WAS REPEATED ON AN ABBOTT ANALYZER AND THE RESULT WAS 87 MG/DL. PATIENT 2: ((B)(6) 2024): THE INITIAL RESULT WAS 10 MG/DL. THE SAME SAMPLE WAS REPEATED ON ANOTHER PRO ANALYZER WITH A RESULT OF 11 MG/DL. THE SAME SAMPLE WAS REPEATED ON AN ABBOTT ANALYZER AND THE RESULT WAS 119 MG/DL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20783786 ↗MalfunctionOct 24, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results
THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR LDLC3 ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 19.4 MG/DL. THE SAMPLE WAS REPEATED DUE TO A REQUEST FROM THE CLINICIAN. THE REPEAT RESULT WAS 127 MG/DL. THE REPEAT RESULT WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998131 ↗
Other LDL & VLDL PRECIPITATION, HDL devices
- CardioChek® Plus Professional Analyzer 2700Polymer Technology Systems, Inc. · LDL & VLDL PRECIPITATION, HDL1 recall114 reports
- CardioChek® Plus Professional Analyzer, Refurbished 2702Polymer Technology Systems, Inc. · LDL & VLDL PRECIPITATION, HDL1 recall5 reports
- Pointe Scientific, Inc. H7545-CALHoriba Instruments Incorporated · Ldl & Vldl Precipitation, Hdl1 recall0 reports
- CardioChek 5700414Henry Schein, Inc. · LDL & VLDL PRECIPITATION, HDL1 recall0 reports
- LDL-Cholesterol Gen.3 07005768190Roche Diagnostics GmbH · LDL & VLDL PRECIPITATION, HDL6 reports
- LDL-Cholesterol Gen.3 07005717190Roche Diagnostics GmbH · LDL & VLDL PRECIPITATION, HDL5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗