CAMINO® CAM02 (UDI 00382830049023)Recalls & reported problems

Natus Medical Incorporated · Intracranial Pressure and Temperature Monitor · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00382830049023), received between Jun 17, 2021 and Nov 4, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Sensing Problem1
  • Electrical /Electronic Property Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Patient Data Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 4, 2024

    PATIENT ON CAMINO ICP MONITOR. PATIENT WAS PLACED FLAT FOR CARE, CAMINO WITH BOLT INTACT. PATIENT HAD BEEN FOR ICP (INCREASED INTRACRANIAL PRESSURE) RANGE 3-5 JUMPED TO 18-19, NO CHANGE IN PATIENTS' ASSESSMENT. HOB (HEAD OF BED) ELEVATED WAVE CONSISTENT, THEN FLATTED. RESIDENT CALLED NATUS 24/7 CLINICAL SUPPORT. DURING TROUBLESHOOTING THE SCREEN PRESENTED TEMPERATURE NUMBERS ¿ [REDACTED NAME] DOES NOT USE TEMPERATURE INTEGRATION. THE SCREEN THEN FLASHED OUT OF THE TEMPERATURE RANGE. DUE TO THE CONCERN OF THE SCREEN AND INFLATED NUMBERS -BOLT REMOVED. BOTH THE CAMINO AND BOLT ARE BEING SENT TO THE COMPANY FOR FURTHER INSPECTION.

    Patient problems: Insufficient Information

    MAUDE report 20599744 ↗
  • MalfunctionJul 11, 2024Break; Patient Data Problem; Break; Patient Data Problem

    PART CAM02 - DISPLAY ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19724262 ↗
  • MalfunctionJul 12, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    PART CAM02 - SYSTEM ERROR DETECTING A SENSOR BOARD FAILURE. DETAILS FROM THE CUSTOMER AS BELOW: *-JUNE 15, SENSOR PLACEMENT IS CARRIED OUT, EQUIPMENT WORKING IN PERFECT CONDITION, PROPER FUNCTIONING IS CONFIRMED WITH THE DOCTOR. *-PATIENT HOSPITALIZED IN ICU, WITH PIC INTERVALS 15 TO 20 MM HG. *-5 DAYS THE EQUIPMENT WORKING NORMALLY. * -6 DAYS IN THE MORNING, THE SCREEN SHOWED A SYSTEM ERROR DETECTING A FAILURE IN THE SENSOR CARD. *- FROM THEN ON, NO MORE ICP INTERVALS WERE SHOWN, UNIT WAS SWAPPED OUT WITH ANOTHER TO CONTINUE PATIENT MONITORING. NO MEDICAL INTERVENTION REQUIRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17306796 ↗
  • MalfunctionJun 17, 2021Incorrect, Inadequate or Imprecise Result or Readings; Device Sensing Problem

    CUSTOMER DOCUMENTED INCONSISTENT ICP MEASUREMENTS INVOLVING THE CAM02. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12018138 ↗

Other Device, Monitoring, Intracranial Pressure devices

All Device, Monitoring, Intracranial Pressure devices →