Skip to content

Grass® MR Conditional/CT Cup Electrodes MRARR27-59Recalls & reported problems

Natus Manufacturing Limited · Grass® MR Conditional/CT Cup Electrodes Array, 27 leads, 3 arrays, 1.5M · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382830051255), received between Feb 27, 2026 and Feb 27, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Superficial (First Degree) Burn1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 27, 2026Insufficient Device Problem Information

    MRARR27-59 (MR ARRAY D/C ELECTRODE 27 LEAD W/1.5M EXT) - A PATIENT WAS BURNED BY ONE OF THE MR ELECTRODES (OF A 27-ARRAY). THEY WERE USING THIS PARTICULAR ELECTRODE AS AN EKG ON THE PATIENT'S SHOULDER AND IT BURNED THE PATIENT. IT WAS CONFIRMED THAT THE PATIENT DID NOT EXPERIENCE ANY OTHER BURNS FROM ANY OTHER ELECTRODES.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 24474681 ↗

FDA clearance and classification

Cleared through 510(k) K242346 on Aug 29, 2024 (decision: Substantially Equivalent). Submitted by Natus Manufacturing Limited as “Grass® MR Conditional/CT Cup Electrodes”. The FDA review took 22 days. FDA record ↗

The FDA regulates product code GXY (electrode, cutaneous) under 21 CFR 882.1320 as a Class II device. 21 CFR 882.1320 ↗

Common questions about the Grass® MR Conditional/CT Cup Electrodes MRARR27-59

Has the Grass® MR Conditional/CT Cup Electrodes MRARR27-59 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00382830051255), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Grass® MR Conditional/CT Cup Electrodes MRARR27-59?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Feb 27, 2026. That report lists patient problem “Superficial (First Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Grass® MR Conditional/CT Cup Electrodes MRARR27-59?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Grass® MR Conditional/CT Cup Electrodes MRARR27-59 FDA approved?

It was cleared through 510(k) K242346 on Aug 29, 2024. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Grass® MR Conditional/CT Cup Electrodes MRARR27-59 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Grass® MR Conditional/CT Cup Electrodes MRARR27-59, and is it still sold?

The FDA's device database lists Natus Manufacturing Limited as the company responsible for it (brand Grass® MR Conditional/CT Cup Electrodes, model MRARR27-59). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Electrode, Cutaneous devices

All Electrode, Cutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page