Skip to content

BD BBL™ Mueller Hinton II Agar 211438Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Mueller Hinton II Agar, 500 g · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00382902114383), received between Mar 2, 2022 and Feb 20, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Illegible Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL¿ MUELLER HINTON II AGAR THAT INSUFFICIENT AND HETEROGENOUS GROWTH OF AN UNSPECIFIED NUMBER OF PATIENT SAMPLES HAS BEEN OBSERVED ACROSS MULTIPLE BATCHES SINCE JULY 2024. THE CUSTOMER PREPARED THE DEHYDRATED CULTURE MEDIA BY STERILIZING THE BASE, ADDING 100ML OF HORSE BLOOD, ADDING 900 ML OF WATER, AND ADJUSTING THE PH TO 7.20±0.2. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21427239 ↗
  • MalfunctionSep 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHILE USING BD BBL MUELLER HINTON II AGAR THAT FALSE RESISTANCE WAS OBTAINED AN UNSPECIFIED NUMBER OF TIMES. HEAVY GROWTH UP TO THE SENSI DISK WAS OBSERVED WITH TRIMETHOPRIM-SULFAMETHOXAZOL AND A THIN HAZE OF GROWTH WAS OBSERVED WITH TRIMETHOPRIM. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20250804 ↗
  • MalfunctionMar 2, 2022Illegible Information; Illegible Information

    IT WAS REPORTED WHILE TESTING WITH BD BBL¿ MUELLER HINTON II AGAR TWO LABEL WERE FOUND TO BE ILLEGIBLE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ON NOVEMBER 25, THE SUPROM CLIENT NOTIFIED A COMPLAINT, ABOUT THE RECEPTION OF THE PRODUCT WITH THE ILLEGIBLE LABEL, CATALOG: 211438, LOT: 1131296, 02 BOTTLES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13654756 ↗

Other CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH devices

All CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page