Skip to content

BD BBL™ Mueller Hinton II Agar 212257Recalls & reported problems

Becton, Dickinson And Company · BD BBL™ Mueller Hinton II Agar, 25 lb · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00382902122579), received between Aug 12, 2022 and Aug 12, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED BY THE CUSTOMER USES PRODUCT TO PRODUCE AGAR IN PETRI DISHES TO ANALYSE ANTIBIOTIC RESISTANCE. BD BBL¿ MUELLER HINTON II AGAR. REPEAT TESTING WAS PERFORMED. NO PATIENT IMPACT WAS REPORTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PRODUCT DOES NOT MEET EUCAST SPECIFICATIONS ESPECIALLY WHEN USING AMPICILLI, I.E.VERY SMALL INHIBITION AREA LEADING TO PRODUCTION STOP. OTHER ANTIBIOTICS ALSO SHOW SMALLER INHIBITION AREAS.- IN ADDITION TO VARIOUS OTHER ANTIBIOTICS, THE INHIBITORY YARD IS PERMANENTLY TOO SMALL, ESPECIALLY IN AMPICILLIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15218474 ↗

Other CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH devices

All CULTURE MEDIA, ANTIMICROBIAL SUSCEPTIBILITY TEST, MUELLER HINTON AGAR/BROTH devices →