BD MicroProbe™ Processor 250100Recalls & reported problems
Becton, Dickinson And Company · BD MicroProbe™ Processor (120 V) · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00382902501008), received between Jul 13, 2021 and Nov 16, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- False Positive Result2
- False Negative Result1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 16, 2022False Positive Result; False Positive Result
IT WAS REPORTED THAT WHILE USING THE BD MICROPROBE¿ PROCESSOR (120 V) THAT THERE WAS FALSE POSITIVES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MICROPROBE - 250100 - ALL POSITIVE RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15811208 ↗MalfunctionMay 20, 2022False Negative Result; False Negative Result
IT WAS REPORTED THAT A BD MICROPROBE¿ PROCESSOR (120 V) HAD A FALSE NEGATIVE RESULT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED BY THE CUSTOMER THAT FALSE NEGATIVE RESULTS COMPARED TO PAP RESULTS. CUSTOMER REPORTING FALSE NEGATIVE RESULTS COMPARED TO PAP RESULTS. PAP SHOWING POSITIVE BUT THE AFFIRM IS SHOWING NEGATIVE RESULTS. CONFIRMATORY TESTING WAS COMPLETE BY WET MOUNT. THERE WAS NO PATIENT IMPACT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14457627 ↗MalfunctionJul 13, 2021False Positive Result; False Positive Result
IT WAS REPORTED THAT WHILE USING BD MICROPROBE¿ PROCESSOR (120 V) FALSE POSITIVE RESULTS WERE OBTAINED. SAMPLES WERE RETESTED ON 2ND INSTRUMENT AND RESULTS WERE NEGATIVE. THERE WAS NO PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FALSE POSITIVES CUSTOMER STATES THEY HAVE A 2ND INSTRUMENT AND ALL SPECIMEN RESULTS WERE ACCEPTABLE UPON REPEAT. THE INSTRUMENT IN QUESTION SEEMS TO BE GETTING STUCK ON CERTAIN WELLS AND DOES NOT WASH THE CARD USING ALL THE WELLS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12163284 ↗
FDA clearance and classification
Cleared through 510(k) K923268 on Jul 30, 1992 (decision: Substantially Equivalent). Submitted by Microprobe Corp. as “Affirm Processor”. The FDA review took 24 days. FDA record ↗
The FDA regulates product code JQW (station, pipetting and diluting, for clinical use) under 21 CFR 862.2750 as a Class I device. 21 CFR 862.2750 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the BD MicroProbe™ Processor 250100
Has the BD MicroProbe™ Processor 250100 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00382902501008), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the BD MicroProbe™ Processor 250100?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Jul 13, 2021 and Nov 16, 2022. In those reports, the most common specific device problems are “False Positive Result” (2) and “False Negative Result” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD MicroProbe™ Processor 250100?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BD MicroProbe™ Processor 250100 FDA approved?
It was cleared through 510(k) K923268 on Jul 30, 1992. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the BD MicroProbe™ Processor 250100, and is it still sold?
The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD MicroProbe™ Processor, model 250100). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STATION, PIPETTING AND DILUTING, FOR CLINICAL USE devices
- FACS 337146Becton, Dickinson And Company · STATION, PIPETTING AND DILUTING, FOR CLINICAL USE102 reports
- Tomcat® TOMCAT SYSTEMHologic, Inc. · STATION, PIPETTING AND DILUTING, FOR CLINICAL USE8 reports
- 405LSR-SIAgilent Technologies, Inc. · STATION, PIPETTING AND DILUTING, FOR CLINICAL USE2 reports
- STAR COMPL. WITH ACCESS. 185080Hamilton Bonaduz AG · Station, Pipetting And Diluting, For Clinical Use2 reports
- Fluent® INSTRUMENT FLUENT 780 BASE UNITTecan Schweiz AG · Station, Pipetting And Diluting, For Clinical Use1 report
- 405TSRVS-SIAgilent Technologies, Inc. · STATION, PIPETTING AND DILUTING, FOR CLINICAL USE1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page