BD Veritor™ Plus Analyzer 256066Recalls & reported problems

Becton, Dickinson And Company · BD Veritor™ Plus Analyzer · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This BD Veritor Plus analyzer, a rapid test reader, has 117 adverse event reports, with 1 injury and no deaths. Many describe wrong results, such as false positives and false negatives. Reports fell from 19 in the prior 12 months to 9 in the last 12 months. 1 recall is open, from January 2022, over overheating risk from unauthorized power adapters.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
114
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 114 adverse event reports that list this exact device (DI 00382902560661), received between Sep 24, 2020 and Feb 9, 2026 — classified as 1 injury, 106 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (41 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 8 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 9 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 6 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 5, 2022Open, Classified

    May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.

    Z-0630-2022 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings41
  • False Positive Result34
  • False Negative Result28
  • Fire8
  • Non Reproducible Results2
  • Contamination of Device Ingredient or Reagent1
  • Patient Data Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions109
  • No Consequences Or Impact To Patient14
  • Test Result3
  • Dyspnea1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 114 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 2 OF 4. IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, A NEGATIVE SARS COVID RESULT FOR ONE (1) PATIENT SAMPLE WAS OBTAINED. HOWEVER, THE USER VISUALLY OBSERVED LINES ON THE TEST CARTRIDGE AND QUESTIONED THE NEGATIVE RESULT INTERPRETING THE LINES AS A POSITIVE RESULT. IT SHOULD BE NOTED THE ACTION OF VISUAL INTERPRETATION OF A TEST CARTRIDGE IS CONSIDERED OFF-LABEL USE OF THE PRODUCT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24294751 ↗
  • MalfunctionFeb 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 4 OF 4. IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, A NEGATIVE SARS COVID RESULT FOR ONE (1) PATIENT SAMPLE WAS OBTAINED. HOWEVER, THE USER VISUALLY OBSERVED LINES ON THE TEST CARTRIDGE AND QUESTIONED THE NEGATIVE RESULT INTERPRETING THE LINES AS A POSITIVE RESULT. IT SHOULD BE NOTED THE ACTION OF VISUAL INTERPRETATION OF A TEST CARTRIDGE IS CONSIDERED OFF-LABEL USE OF THE PRODUCT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24294743 ↗
  • MalfunctionFeb 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 1 OF 4. IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, A NEGATIVE SARS COVID RESULT FOR ONE (1) PATIENT SAMPLE WAS OBTAINED. HOWEVER, THE USER VISUALLY OBSERVED LINES ON THE TEST CARTRIDGE AND QUESTIONED THE NEGATIVE RESULT INTERPRETING THE LINES AS A POSITIVE RESULT. IT SHOULD BE NOTED THE ACTION OF VISUAL INTERPRETATION OF A TEST CARTRIDGE IS CONSIDERED OFF-LABEL USE OF THE PRODUCT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24294729 ↗
  • MalfunctionFeb 9, 2026Incorrect, Inadequate or Imprecise Result or Readings

    REPORT 3 OF 4. IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, A NEGATIVE SARS COVID RESULT FOR ONE (1) PATIENT SAMPLE WAS OBTAINED. HOWEVER, THE USER VISUALLY OBSERVED LINES ON THE TEST CARTRIDGE AND QUESTIONED THE NEGATIVE RESULT INTERPRETING THE LINES AS A POSITIVE RESULT. IT SHOULD BE NOTED THE ACTION OF VISUAL INTERPRETATION OF A TEST CARTRIDGE IS CONSIDERED OFF-LABEL USE OF THE PRODUCT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24294713 ↗
  • MalfunctionDec 8, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, AN UNSPECIFIED NUMBER OF FALSE POSITIVE FLU A+B PATIENT RESULTS WERE OBTAINED. THE SAME PATIENT SAMPLES WERE RETESTED USING DIFFERENT VERITOR ANALYZERS, WHICH WERE NOTED TO BE FUNCTIONING NORMALLY. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED FROM THE CUSTOMER AFTER SEVERAL FOLLOW UP ATTEMPTS BY BD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23750976 ↗
  • MalfunctionNov 7, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, A FALSE POSITIVE RESULT WAS OBTAINED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED FROM THE CUSTOMER AFTER SEVERAL FOLLOW UP ATTEMPTS BY BD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23496894 ↗
  • MalfunctionOct 27, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) FALSE POSITIVE FLU B PATIENT RESULT WAS OBTAINED, SIMULTANEOUSLY WITH A POSITIVE SARS-COV-2 RESULT. UPON CONFIRMATORY TESTING USING A STANDALONE FLU TEST CARTRIDGE, THE PATIENT SAMPLE WAS NEGATIVE FOR FLU B. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393597 ↗
  • MalfunctionOct 27, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) FALSE POSITIVE FLU B PATIENT RESULT WAS OBTAINED, SIMULTANEOUSLY WITH A POSITIVE SARS-COV-2 RESULT. UPON CONFIRMATORY TESTING USING A STANDALONE FLU TEST CARTRIDGE, THE PATIENT SAMPLE WAS NEGATIVE FOR FLU B. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393431 ↗
  • MalfunctionOct 27, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) FALSE POSITIVE FLU B PATIENT RESULT WAS OBTAINED, SIMULTANEOUSLY WITH A POSITIVE SARS-COV-2 RESULT. UPON CONFIRMATORY TESTING BY USING ANOTHER COMBINATION KIT, THE PATIENT SAMPLE WAS NEGATIVE FOR FLU AND POSITIVE FOR COVID. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393413 ↗
  • MalfunctionOct 27, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) FALSE POSITIVE FLU B PATIENT RESULT WAS OBTAINED, SIMULTANEOUSLY WITH A POSITIVE SARS-COV-2 RESULT. UPON CONFIRMATORY TESTING BY A LAB USING QUAD PCR, THE PATIENT SAMPLE WAS NEGATIVE FOR FLU AND POSITIVE FOR COVID. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393412 ↗
  • MalfunctionOct 27, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) FALSE POSITIVE FLU B PATIENT RESULT WAS OBTAINED, SIMULTANEOUSLY WITH A POSITIVE SARS-COV-2 RESULT. UPON CONFIRMATORY TESTING BY A LAB USING QUAD PCR, THE PATIENT SAMPLE WAS NEGATIVE FOR FLU B AND POSITIVE FOR COVID. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393410 ↗
  • MalfunctionOct 27, 2025False Positive Result; False Positive Result

    IT WAS REPORTED WHILE USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) FALSE POSITIVE FLU B PATIENT RESULT WAS OBTAINED, SIMULTANEOUSLY WITH A POSITIVE SARS-COV-2 RESULT. UPON CONFIRMATORY TESTING BY A LAB USING QUAD PCR, THE PATIENT SAMPLE WAS NEGATIVE FOR FLU AND POSITIVE FOR COVID. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393409 ↗
  • MalfunctionJul 31, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) PATIENT SAMPLE RESULTED AS NEGATIVE FOR GROUP A STREP. THE USER QUESTIONED THE NEGATIVE RESULT AS THE PATIENT WAS SYMPTOMATIC AND A LINE WAS VISUALLY OBSERVED ON THE TEST CARTRIDGE. IMMEDIATELY AFTER VISUALLY READING THE TEST CARTRIDGE, IT WAS INSERTED INTO A SECOND VERITOR ANALYZER AND A POSITIVE RESULT FOR GROUP A STREP WAS OBTAINED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22680259 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 5 OF 9. IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332965 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 8 OF 9. IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332457 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 9 OF 9. IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332456 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 6 OF 9 IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332455 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 3 OF 9 IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332454 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 4 OF 9 IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THE PATIENT WAS PRESCRIBED TAMIFLU; HOWEVER, IT WAS REPORTED THE PATIENT NEVER STARTED TAKING THE MEDICATION DUE TO NOTIFICATION FROM THE PROVIDER. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332453 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 7 OF 9. IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332451 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 2 OF 9. IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332446 ↗
  • MalfunctionJun 25, 2025False Positive Result; False Positive Result

    REPORT 1 OF 9 IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE USER ALLEGED ONE (1) FALSE POSITIVE FLU A PATIENT RESULT WAS OBTAINED. NO CONFIRMATORY TESTING WAS PERFORMED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332158 ↗
  • MalfunctionApr 2, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, ONE (1) PATIENT SAMPLE RESULTED AS NEGATIVE FROM A NEW ANALYZER; HOWEVER, THE USER VISUALLY OBSERVED POSITIVE LINES ON THE TEST CARTRIDGE AND EXPECTED A POSITIVE RESULT. THE SAME TEST CARTRIDGE WAS REINSERTED INTO THE NEW ANALYZER AND AGAIN A NEGATIVE RESULT WAS OBTAINED. THE USER THEN REINSERTED THE SAME TEST CARTRIDGE INTO A DIFFERENT ANALYZER, WHICH PROVIDED A POSITIVE RESULT. IT WAS NOT PROVIDED WHICH ANALYTE WAS DISCREPANT. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21754114 ↗
  • MalfunctionMar 7, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED AFTER USING THE BD VERITOR¿ PLUS ANALYZER, THE USER QUESTIONED ONE (1) POSITIVE FLU A PATIENT RESULT AFTER REMOVING AND VISUALLY READING THE TEST CARTRIDGE. THE SAME TEST CARTRIDGE WAS REINSERTED INTO A SECOND ANALYZER AND A NEGATIVE FLU A RESULT WAS OBTAINED. THERE WERE NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21552387 ↗
  • MalfunctionNov 19, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED WHEN USING THE BD VERITOR¿ PLUS ANALYZER, THE ANALYZER PROVIDED ONE (1) NEGATIVE PATIENT STREP RESULT; HOWEVER, THE CUSTOMER VISUALLY OBSERVED A LINE ON THE CARTRIDGE AFTER REMOVAL. THE SAME SAMPLE CARTRIDGE WAS THEN REINSERTED INTO A SECOND VERITOR ANALYZER AND A POSITIVE RESULT WAS OBTAINED. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20722901 ↗

Other COLORIMETER, PHOTOMETER, SPECTROPHOTOMETER FOR CLINICAL USE devices

All COLORIMETER, PHOTOMETER, SPECTROPHOTOMETER FOR CLINICAL USE devices →