BD Veritor Plus Analyzer - intended to provide rapid test… recallZ-0630-2022 · Becton Dickinson & Co. · initiated Jan 5, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.
FDA root cause category: Under Investigation by firm
What the company is doing
***UPDATE June 6, 2022**** BD issued a Medical Device Product Advisory letter to Distributors and Users dated July 6, 2022 as a follow-up notification to the voluntary product advisory dated January 5, 2022. In light of this recent complaint, BD is providing additional recommendations to customers to avoid a fire until BD can replace affected units in the field. Letter states: 1. Ensure the contents of this Product Advisory are read and understood by those within your organization. 2. To verify if your BD Veritor" Plus Analyzer(s) is impacted and to complete the acknowledgment distributor response form, you can visit the following website https://go.bd.com/Veritor-Analyzer-Notice.html and enter your serial number(s). 3. AT ALL TIMES USE ONLY the 5V 3A, 15W AC/DC power adapter provided with the BD Veritor" Plus Analyzers. If needed, a replacement AC/DC power adapter (catalog number 445201) can be purchased by calling BD Customer Service. 4. Affix the label(s) provided with this notification to your BD provided AC/DC power adapter(s) to ensure the correct power adapter is always connected to the analyzer(s). BD will contact your account to coordinate replacement of your affected analyzer(s) with expiration dates between December 2021 and August 2024. In the meantime, please implement the recommendations detailed above to mitigate risk of fire. 2. BD will provide credit or replacement for any affected analyzers that are disposed of in response to this notification. ------------------------------------------------------------------------------------------------------------------ BD issued Medical Device Advisory letter dated 1/05/22 to Lab Directors/Risk Managers and Distributors . Letter states reason for recall, health risk and action to take: 1. Ensure the contents of this Product Advisory are read and understood by those within your organization. 2. USE ONLY the AC power adapter provided with the BD Veritor" Plus Analyzers. An image of the AC power adapter provide
Recall details
- Recall number
- Z-0630-2022
- Status
- Open, Classified
- Initiated by the company
- Jan 5, 2022
- Posted by the FDA
- Feb 9, 2022
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 126,952 Analyzers Updated:129, 848 Analyzers
- Product code
- PSZ
- FDA submissions
- K180438
- Distribution
- Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Fiji, Guatemala, Hong Kong, Honduras, Haiti, Indonesia, India, Jamaica, Cambodia, South Korea, Macao, Mexico, New Zealand, Panama, Peru, Philippines, Singapore, El Salvador, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Vietnam.
Product description
BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066
Code information (models, lots, serial numbers)
All unexpired analyzers (Exp. Date 12/2021 - 8/2024) UDI: 00382902560661
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD Veritor™ Plus Analyzer 256066Becton, Dickinson And Company · COLORIMETER, PHOTOMETER, SPECTROPHOTOMETER FOR CLINICAL USE1 recall114 reports
Other devices under the same FDA submission
1 device cleared under K180438 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BD Veritor™ System For Rapid Detection of Flu A+B CLIA-waived kit 256045Becton, Dickinson And Company · Devices detecting influenza A, B, and C virus antigens251 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page