Skip to content

BD Luer-Lok 303310Recalls & reported problems

Becton, Dickinson And Company · SYRINGE 20ML LL S/C 50 · Class 2

All models of BD Luer-Lok (48)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This BD Luer-Lok syringe is a piston syringe used to draw up and inject fluids. The 82 reports describe foreign particles inside the syringe, along with leaks and cracked barrels; no deaths or injuries are reported. No recalls name this device. Reports dropped from 24 to 9 over the past year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
82
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 82 adverse event reports that list this exact device (DI 00382903033102), received between Mar 27, 2021 and Apr 7, 2026 — classified as 82 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 4 reportsJan 2024: 7 reportsFeb 2024: 3 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 4 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 7 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material24
  • Contamination15
  • Leak/Splash10
  • Contamination /Decontamination Problem9
  • Crack7
  • Defective Component5
  • Defective Device5
  • Material Puncture/Hole5

Patient problems

  • No Clinical Signs, Symptoms or Conditions79
  • Insufficient Information3

Recent adverse event reports

The 25 most recent of 82 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD VISIBLE DROPLETS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL #:303310 BATCH#:5253925. PLEASE SEE THE INFORMATION BELOW FROM REGARDING AN ISSUE THEY ARE REPORTING WITH CAN YOU PLEASE START A QI AND WORK WITH TO GATHER ANY ADDITIONAL INFORMATION YOU MAY NEED? PLEASE INCLUDE THE DEFECTIVE TRACKING TEAM IN ANY CORRESPONDENCE FOR THE TRACKING/DOCUMENTATION. THIS REPORT'S UNIQUE IDENTIFIER IS 1497 SHOULD YOU NEED TO REFERENCE IT. REPORT DATE: (B)(6) 2026 PRODUCT DESCRIPTION: BD 20CC LUER-LOK SYRINGE MANUFACTURER PRODUCT NUMBER: REF303310 LOT NUMBER: 5253925 IS PRODUCT IN A KIT/CUSTOM PACK: YES DATE OF EVENT: 2026-03-31 DESCRIPTION OF DEFECT: MOISTURE INSIDE BARREL OF SYRINGE, DROPLET IN GAP BETWEEN RUBBER ON PLUNGER AND HUB OF BARREL. USED ON PATIENT: NO LOCATION OF PRODUCT: INPATIENT PHARMACY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24807881 ↗
  • MalfunctionFeb 23, 2026Contamination

    IT WAS REPORTED THAT BD SYRINGE 20ML LL S/C 50 CONTAINED FOREIGN MATTER. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WE ARE FINDING BLACK FLOATIES IN YOUR BD 20ML SYRINGE. WE HAVE PULLED ALL 20 ML SYRINGES FROM MY AREA. THE ONES THAT ARE IN THE MYELOGRAM TRAY HAVE THEM AS WELL. I HAVE A SYRINGE AND THE WRAPPER FOR THE INCIDENT ON 2/4 THAT I SENT THE PICTURE OF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24415540 ↗
  • MalfunctionFeb 4, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT BD SYRINGE 20ML LL S/C 50 CONTAINED FOREIGN MATTER. RCC RECEIVED A COMPLAINT VIA EMAIL. WE ARE FINDING BLACK FLOATIES IN YOUR BD 20ML SYRINGE REF 303310. WE HAVE PULLED ALL 20 ML SYRINGES FROM MY AREA. THE ONES THAT ARE IN THE MYELOGRAM TRAY HAVE THEM AS WELL. DO YOU HAVE A FORM THAT CAN BE FILLED OUT ABOUT THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24260701 ↗
  • MalfunctionJan 7, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL#: 303310, BATCH#: 5163348. VERBATIM: GOOD AFTERNOON, WE RECEIVED A REPORT FROM OUR XXXX CLINICIANS CONCERNING AN EVENT THEY HAVE EXPERIENCED WHILE USING A SYRINGE 20ML LL STERILE LATEX FREE (303310). LOT# 5163348. THE EVENT DATE IS 12.20.2025. NO HARM TO THE PATIENT HAS BEEN REPORTED. THE COMPLAINT IS DESCRIBED: PARTICULATE FOUND IN THE GASKET OF 20ML BD LUER-LOK TIP SYRINGE. REF 303310. LOT 5163348, EXP 2030-06-30. SEQUESTERED IN WP11 PHARMACY SATELLITE BY THE STERILE PRODUCTS PHARMACIST DESK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23999141 ↗
  • MalfunctionJan 6, 2026Contamination

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23977766 ↗
  • MalfunctionJan 4, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL: 303310; BATCH#: 5132719. VERBATIM: 2 DIFFERENT SYRINGES HAD PARTICULATE IN THE SYRINGE PRIOR. THE PARTICLES WEREN'T PRESENT IN THE SOLUTION PRIOR TO USE AND WAS DETERMINED TO BE IN THE SYRINGE. IMAGES ARE ATTACHED BUT THE 2 FAULTY ITEMS ARE STILL IN THE CUSTOMER'S OFFICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23953930 ↗
  • MalfunctionNov 27, 2025Crack; Leak/Splash; Crack; Leak/Splash

    ADDITIONAL INFORMATION: WHERE WAS THE CRACK? ¿ THE BARREL AND BARREL FLANGE. WHICH CLOSED SYSTEM WAS THE SYRINGE ATTACHED TO? - CHEMOLOCK. DID THE SYRINGE LEAK WHERE THE CRACK WAS - YES SOME LIQUID LEAKED OUT THE SYRINGE BARREL WHERE THE SYRINGE WAS CRACKED AND THE CLOSED SYSTEM ALSO HAD A LEAK ¿ NO THE CHEMOLOCK PIECES BEHAVING AS USUAL. CHEMO LIKELY LEAKING OUT THE SIDE OF THE CRACKED SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23665876 ↗
  • MalfunctionNov 26, 2025Material Integrity Problem; Material Integrity Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23656145 ↗
  • MalfunctionNov 7, 2025Crack; Leak/Splash; Crack; Leak/Splash

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 BARREL CRACKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 303310 LOT # 5069180   VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. MY PHARMACY TECHNICIAN WAS DRAWING UP A DOSE OF INJECTABLE INTO A 20CC BD SYRINGE AND FLUID STARTED LEAKING OUT THE SIDE OF THE SYRINGE. UPON CLOSE INSPECTION, IT WAS DISCOVERED THAT THERE WAS A SMALL CRACK IN THE BARREL OF THE SYRINGE. THE PREPARATION WAS DISCARDED AND NEW SUPPLIES AND DRUG WERE OBTAINED. THERE WAS NO HARM TO THE PATIENT, ONLY WASTED DRUG FOR THE PHARMACY. ITEM# 303310 BATCH# 5069180 CUSTOMER RESPONSE ON 24-OCT-2025: THE SAMPLE IS NO LONGER AVAILABLE. IT WAS INADVERTENTLY DISCARDED BY OUR CLEANING STAFF AFTER SEVERAL WEEKS OF SITTING ON MY DESK. I DO HAVE A PHOTO - PLEASE SEE THE ATTACHED IMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23501576 ↗
  • MalfunctionOct 22, 2025Leak/Splash; Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23361163 ↗
  • MalfunctionSep 5, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 303310. BATCH # 5069190. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. ITEM: 303310. QUANTITY AFFECTED: (B)(4) EACH. SERIAL/LOT NUMBER: (B)(6). ARE ANY SAMPLES AVAILABLE FOR RETURN? YES. REPORTED ISSUE: CUSTOMER REPORTING SOME SYRINGES HAVE MOISTURE IN THEM. CUSTOMER RESPONSE ON 21-AUG-2025. PICTURE BELOW. THE HOSPITAL REPORTED THIS TO ME ON THURSDAY, AUGUST 14TH. THE CUSTOMER NO LONGER HAS THE SAMPLE, THEY DISCARDED IT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22984042 ↗
  • MalfunctionAug 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22821987 ↗
  • MalfunctionJul 14, 2025Contamination; Contamination

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 303310. BATCH # UNKNOWN. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. PHARMACY STAFF IDENTIFIED MULTIPLE QUALITY ISSUES WITH THE BD STERILE LUER-LOCK 20 ML SYRINGES, INCLUDING PARTICULATE MATTER INSIDE THE BARREL. IN SOME CASES, THE PARTICULATE MATTER WAS NOT IDENTIFIED UNTIL THE FINAL STERILE PRODUCT WAS BEING CHECKED. ONE OF THE SYRINGES FROM THE AFFECTED LOT WAS MELTED AT THE TOP AND HAD A MISSING LUER LOCK. ANOTHER CONCERN IDENTIFIED WAS INCONSISTENT VOLUME MARKINGS ON THE SYRINGE. THE AFFECTED LOTS INCLUDE 303310. ADDITIONALLY, THERE HAVE BEEN MULTIPLE REPORTS OF QUALITY AND MANUFACTURING ISSUES WITH BD SYRINGES. SOME PACKAGES WERE MISSING SYRINGES, AND OTHERS CONTAINED DAMAGED SYRINGES. REFERENCE REPORTS#: MW5171495, MW5171496.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22502704 ↗
  • MalfunctionJul 14, 2025Contamination; Contamination

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22502692 ↗
  • MalfunctionJul 10, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22468737 ↗
  • MalfunctionJun 26, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 303310, BATCH # 5015724. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WE HAVE FOUND MULTI 20ML SYRINGES THAT HAVE ISSUES. WE FOUND TWO SO FAR WITH DEFECTIVE HOLE IN THE TOP AND ONE WITH WHAT APPEARS TO BE PLASTIC SHARDS IN THE SYRINGE ITSELF. THIS IS OF GREAT CONCERN FOR OUR HOSPITAL PLEASE CONTACT (B)(4) PLEASE. CUSTOMER RESPONDED ON 18-JUN. 1. THE EVENT OCCURRED ON 06-12-2025. 2- WE DO STILL HAVE 2 DEFECTIVE SYRINGES AND I AM INCLUDING (B)(4) TO HELP WITH WHERE TO MAIL THE LABEL AND WHAT DEPARTMENT IT SHOULD GO TO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22342267 ↗
  • MalfunctionJun 2, 2025Contamination /Decontamination Problem; Device Contaminated During Manufacture or Shipping; Contamination /Decontamination Problem; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL #303310 LOT #4319964. IV TECH NOTICED A PARTICULATE ON THE PLUNGER OF AN UNOPENED BD 20ML IV SYRINGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22126509 ↗
  • MalfunctionApr 29, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21929398 ↗
  • MalfunctionApr 18, 2025Contamination; Contamination

    BASED ON INVESTIGATION RESULTS, THE REPORTED FOREIGN MATTER WAS SILICONE WHICH IS PART OF THE MANUFACTURING PROCESS, AND THE AMOUNT OF SILICONE FOUND WAS WITHIN SPECIFICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21865817 ↗
  • MalfunctionApr 18, 2025Contamination; Contamination

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL#: 303310, BATCH#: 4023256 AND 4073843, 4044308. IT WAS REPORTED BY CUSTOMER THAT THEY HAVE TWO 20 ML SYRINGE WITH SMALL PLASTIC PIECES SEALED INSIDE OF THE PACKAGE. ADDITIONALLY, LOOKING THROUGH THE SEQUESTERED STOCK, THEY NOTICED VERY INCONSISTENT MARKINGS. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. WE HAVE AN ISSUE WITH PRODUCT # 303310 (STERILE LUER-LOK 20 ML SYRINGES), LOT #4044308. EVERY SYRINGE HAS HAD PARTICULATE IN IT, WITH PARTICULATE NOT SEEN IN SOME CASES UNTIL THE FINAL STERILE PRODUCT IS EVALUATED. ONE OF THE SYRINGES WE OPENED IS MELTED ON THE TOP AND IS COMPLETELY MISSING THE LUER-LOK TIP. I HAVE A SAMPLE OF A SYRINGE WITH PARTICULATE AND THE MELTED SYRINGE. I AM VERY CONCERNED THAT PARTICULATE PRODUCTS COULD BE ADMINISTERED TO PATIENTS IF THE LOT IS NOT RECALLED. WE ARE ALSO EXPERIENCING PARTICULATE ISSUES WITH LOT # 4023256 TO A LESSER DEGREE. PLEASE SHARE THE DATE OF EVENT. ISSUE DISCOVERED DURING THE COURSE OF THE WEEKEND, (B)(6) 2025. PLEASE CONFIRM WHICH LOT# IS MELTED ON THE TOP. 4044308 - PLEASE SHARE THE QUANTITY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21865597 ↗
  • MalfunctionApr 17, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 BARREL CRACKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 303310 BATCH #4 261568. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED . GOOD MORNING BD TEAM, WE HAVE A TREND OF 20ML SYRINGES WITH HOLES IN THEM. PLEASE SEE BELOW. SAFER # PS-2025-059698 LOCATION **** DATE 04/01/25 REF 303310 LOT 4261568 --------------------------------------------------- CUSTOMER RESPONDED ON 09-APR DATE OF EVENT - 04/01/2025. ISSUE OBSERVED BEFORE USE / DURING USE - PINHOLE IN 20CC SYRINGE, CAUSING SODIUM CHLORIDE TO LEAK OUT OF SYRINGE. STATUS OF SAMPLE (YES/ NO) - YES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21861652 ↗
  • MalfunctionApr 10, 2025Contamination; Contamination

    IT WAS REPORTED THAT THE BD SYRINGE 20ML LL S/C 50 HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL#: 303310, BATCH#: 4044308 & 4023256. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. WE HAVE AN ISSUE WITH PRODUCT #303310 (STERILE LUER-LOK 20 ML SYRINGES), LOT #4044308. EVERY SYRINGE HAS HAD PARTICULATE IN IT, WITH PARTICULATE NOT SEEN IN SOME CASES UNTIL THE FINAL STERILE PRODUCT IS EVALUATED. ONE OF THE SYRINGES WE OPENED IS MELTED ON THE TOP AND IS COMPLETELY MISSING THE LUER-LOK TIP. I HAVE A SAMPLE OF A SYRINGE WITH PARTICULATE AND THE MELTED SYRINGE. I AM VERY CONCERNED THAT PARTICULATE PRODUCTS COULD BE ADMINISTERED TO PATIENTS IF THE LOT IS NOT RECALLED. WE ARE ALSO EXPERIENCING PARTICULATE ISSUES WITH LOT #4023256 TO A LESSER DEGREE. ADDITIONAL INFORMATION PROVIDED: PLEASE CONFIRM DID THE PARTICULATE PRESENT IN FLUID OR NON-FLUID PATH. A. IN FLUID PATH FOR THE PRODUCTS IN LOT 4044308 AND 4023256.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21814761 ↗
  • MalfunctionApr 10, 2025Contamination; Contamination

    BASED ON INVESTIGATION RESULTS, THE REPORTED FOREIGN MATTER WAS SILICONE WHICH IS PART OF THE MANUFACTURING PROCESS, AND THE AMOUNT OF SILICONE FOUND WAS WITHIN SPECIFICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21814759 ↗
  • MalfunctionApr 8, 2025Crack; Leak/Splash

    PLASTIC SYRINGE CRACKED AS TECHNICIAN WAS DRAWING UP CHEMO MEDICATION. LEADING TO DRUG LEAKING OUT OF CLOSED SYSTEM TRANSFER SYSTEM.

    Patient problems: Insufficient Information

    MAUDE report 21798358 ↗
  • MalfunctionMar 13, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    MATERIAL #303310, LOT #4114045, 4261566. IT WAS REPORTED THAT THE BD LUER-LOK HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. WE HAVE A 20ML SYRINGE THAT HAS SPOTS INSIDE OF IT. FURTHER INFORMATION PROVIDED: ¿ 1. IN THE MEDSUN FORM MENTIONED THAT THE DATE OF THE EVENT AS ¿JAN-2025¿. COULD YOU PLEASE SPECIFY THE EXACT DATE WHEN THE REPORTED ISSUE HAPPENED? 1.9.2025 2. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. NONE 3.THE MEDSUN FORM INDICATES THAT THE SAMPLE HAS BEEN ¿RETURNED TO MANUFACTURER (YES).¿ COULD YOU PLEASE CONFIRM WHETHER THE SAMPLE HAS ALREADY BEEN RETURNED TO BD? IF SO, PLEASE SHARE THE TRACKING NUMBER. 1.13 EMAILED [email protected]. 01.17 FEDEX TRK# (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21599722 ↗

FDA clearance and classification

Cleared through Special 510(k) K151766 on Sep 2, 2015 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “BD Single Use, Hypodermic Syringe”. The FDA review took 64 days. FDA record ↗

The FDA regulates product code FMF (syringe, piston) under 21 CFR 880.5860 as a Class II device. 21 CFR 880.5860 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the BD Luer-Lok 303310

Has the BD Luer-Lok 303310 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00382903033102), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the BD Luer-Lok 303310?

The FDA's MAUDE database lists 82 adverse event reports for this device, received between Mar 27, 2021 and Apr 7, 2026. In those reports, the most common specific device problems are “Device Contamination with Chemical or Other Material” (24), “Contamination” (15) and “Leak/Splash” (10). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BD Luer-Lok 303310?

None of the reports are classified as deaths, 0 as injuries and 82 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BD Luer-Lok 303310 FDA approved?

It was cleared through Special 510(k) K151766 on Sep 2, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the BD Luer-Lok 303310, and is it still sold?

The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD Luer-Lok, model 303310). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Syringe, piston devices

All Syringe, piston devices →

Sources and updates