Skip to content

BD Nokor 305214Recalls & reported problems

Becton, Dickinson And Company · NEEDLE VENTED NOKOR 18X1 TW · Class 1

All models of BD Nokor (5)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00382903052141), received between Jan 11, 2019 and May 10, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Defective Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Component3
  • Device Contamination with Chemical or Other Material2

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1
  • No Patient Involvement1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 10, 2024Defective Component; Defective Component

    NO ADDITIONAL INFORMATION RECEIVED MATERIAL# 305214 BATCH# UNKNOWN IT WAS REPORTED BY CUSTOMER THAT THEY ARE HAVING ISSUES WITH THE NEEDLE CORING OUT THE VIAL STOPPER. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. REQUESTING IF WE HAVE RECOMMENDATIONS FOR A CUSTOMER THAT IS HAVING ISSUES USING THE NOKOR VENTED NEEDLE (REF 305214). THEY ARE HAVING ISSUES WITH THE NEEDLE CORING OUT THE VIAL STOPPER. THEY HAVE TO THEN PLACE ANOTHER NEEDLE TO FILTER OUT THE MATERIAL. ALSO, HAS QUESTIONS ON HOW THE NOKOR VENTED NEEDLES WORK.

    Patient problems: Insufficient Information

    MAUDE report 19285354 ↗
  • MalfunctionMay 9, 2024Defective Component; Defective Component

    NO ADDITIONAL INFORMATION RECEIVED MATERIAL #: 305214 BATCH#: UNKNOWN IT WAS REPORTED BY CUSTOMER THAT THE CURRENT VENTED NEEDLE WE USE IS CAUSING A LOT OF WORK ON THE BACK END DUE TO IT CORING THE VIAL LEAVING THE PRODUCT CONTAMINATED AND THEREFORE IT HAS TO BE FILTERED. I THINK THERE IS A QUESTION ALSO ABOUT IF IT IS THE VIAL STOPPER, WE ARE HAVING ISSUES WITH. I NEED TO RULE OUT ONE. I THOUGHT WE COULD TRY ANOTHER NEEDLE TO SEE IF THAT HELPS. DO WE HAVE ANY OTHER VENTED NEEDLES LIKE THIS THAT WE USED TO HAVE? THAT SLIDE PART WAS SO HELPFUL IN COMPOUNDING AND USE WITH OUR REPEATER PUMP. I DON¿T THINK VERY MANY SITES USE THE REPEATER PUMPS LIKE WE DO. VERBATIM: COMPLAINT RECEIVED VIA EMAIL(S) ATTACHED. THE CURRENT VENTED NEEDLE WE USE IS CAUSING A LOT OF WORK ON THE BACK END DUE TO IT CORING THE VIAL LEAVING THE PRODUCT CONTAMINATED AND THEREFORE IT HAS TO BE FILTERED. I THINK THERE IS A QUESTION ALSO ABOUT IF IT IS THE VIAL STOPPER, WE ARE HAVING ISSUES WITH. I NEED TO RULE OUT ONE. I THOUGHT WE COULD TRY ANOTHER NEEDLE TO SEE IF THAT HELPS. DO WE HAVE ANY OTHER VENTED NEEDLES LIKE THIS THAT WE USED TO HAVE? THAT SLIDE PART WAS SO HELPFUL IN COMPOUNDING AND USE WITH OUR REPEATER

    Patient problems: Insufficient Information

    MAUDE report 19277173 ↗
  • MalfunctionDec 7, 2023Defective Component; Defective Component

    NO ADDITIONAL INFORMATION RECEIVED. MATERIALS#: 305214, BATCH#: 317334.1 IT WAS REPORTED BY THE CUSTOMER THAT THE NOKOR NON-CORING VENTED NEEDLE IS CLASSIFIED AS A BLUNT FILL NEEDLE. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. QUESTION: CALLER STATES THEY USE THE NOKOR NON-CORING VENTED NEEDLE 305214. CALLER WAS ASKING IF THERE ARE ANYINSTRUCTIONS FOR HOW TO USE THE NOKOR NEEDLES. CALLER STATES THEY RECENTLY PURCHASED THESE TO HELP ELIMINATECORING IN THEIR VIALS. THEY USED THE NOKOR NEEDLE THIS MORNING AND ENDED UP HAVING CORING IN A THIRD OF THEIRVIALS LEADING TO STOPPER MATERIAL IN THE VIALS. CALLER PROVIDED THE LOT NUMBER 3173341. CALLER ALSO ASKING FOR ACOPY OF THE COA. RESPONSE: INFORMED CALLER THE NOKOR NON-CORING VENTED NEEDLE IS CLASSIFIED AS A BLUNT FILL NEEDLE. PRODUCT CODE 305214IS VENTED, MEANING IT HAS A SIDE PORT TO HELP VENT AND REDUCE VIAL PRESSURE. EMAILED THE BLUNT FILL USAGEGUIDELINES FOR REFERENCE. INFORMED CALLER I WILL SUBMIT THIS TO OUR PRODUCT COMPLAINTS TEAM FOR FURTHERINVESTIGATION. INFORMED CALLER THE COA CAN BE LOCATED ON THE BD REGULATORY SITE. PRODUCT/MODELNUMBER: 305214.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18282611 ↗
  • MalfunctionJun 8, 2020Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT THE NEEDLES ARE CORING PRIOR TO USE WITH A BD NEEDLE VENTED NOKOR 18X1 TW. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED NEEDLES ARE CORING.

    Patient problems: No Patient Involvement

    MAUDE report 10131438 ↗
  • MalfunctionJan 11, 2019Device Contamination with Chemical or Other Material

    IT WAS REPORTED THAT FOUR OF THE BD NOKOR¿ FILTER AND ADMIX NEEDLES EXPERIENCED COMPROMISED STERILITY FROM FOREIGN MATTER.

    Patient problems: No Information

    MAUDE report 8240993 ↗

Other NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE devices

All NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page