CONFIDENCE SPINAL CEMENT SYSTEM 283903611Recalls & reported problems
Medos International Sàrl · CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G x 6 Inches · Class 1
All models of CONFIDENCE SPINAL CEMENT SYSTEM (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10705034209494), received between Jun 15, 2018 and Jan 30, 2025 — classified as 10 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break17
- Entrapment of Device1
- Material Deformation1
Patient problems
- Foreign Body In Patient8
- No Clinical Signs, Symptoms or Conditions3
- Local Reaction2
- Unspecified Tissue Injury2
- Insufficient Information1
- No Consequences Or Impact To Patient1
- Not Applicable1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 30, 2025Break; Break
IT WAS REPORTED THAT THE SURGEON PERFORMING AN L5-S1 FUSION, WANTED TO USE JAMSHEDI/PROBE TO TAP INTO BONE BEFORE PLACING A GUIDEWIRE OVER WHICH HE WOULD PLACE CORTICAL FIX SCREW. HOSPITAL HAD NONE OF THEIR USUAL KITS AS THEY ARE NOW DISCONTINUED, SO NOW THE ONLY OPTION WAS TO USE THE JAMSHEDI/PROBE. SURGEON USED TOFFEE HAMMER TO TAP IN THE CANNULA, THE T-HANDLE AT TOP OF CANNULA SNAPPED OFF SO REMOVED THE INNER PART TO PLACE GUIDEWIRE. THE OUTER TUBE OF CANNULA COULD THEN NOT BE REMOVED AND WAS CRUSHED WHEN USING A ¿BONE NIBBLER¿ TO TRY AND PULL IT OUT. SURGEON WAS SNAPPING OFF SMALL PARTS IN AN ATTEMPT TO REMOVE IT. THE TUBE WAS TOO CRUSHED TO PASS THE INNER PART BACK DOWN TO GIVE IT SOME RIGIDITY. AFTER A FURTHER 40 MINUTES AND THE ASSISTANCE OF A BURR, THE SURGEON WAS ABLE TO DRILL DOWN THE SIDE OF THE BROKEN PIECE AND REMOVE THE LAST BIT OF CANNULA. SCRUB TEAM AND SALES REPRESENTATIVE CONFIRMED THAT IT WAS THE FULL TIP AS COMPARED SIDE BY SIDE TO ANOTHER CANNULA. AN ADDITIONAL HOUR WAS ADDED TO THE CASE DUE TO THE EVENT. HOSPITAL HAS SINCE SOURCED NEW REPLACEMENT OF THEIR DISPOSABLE PROBES. THEY HAVE ALSO ADDED VIPER 2 CANNULATED PEDICLE PROBES WITH STYLETS TO HAVE THE REUSABL
Patient problems: Insufficient Information
MAUDE report 21274947 ↗InjuryJul 26, 2023Break; Break
THERE WERE NO ADVERSE CONSEQUENCE/S THAT AFFECTED THE PATIENT BECAUSE OF THE REPORTED EVENT. PATIENT IS OK.
Patient problems: Unspecified Tissue Injury
MAUDE report 17398008 ↗InjuryApr 25, 2023Break; Entrapment of Device; Break; Entrapment of Device
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT PATIENT UNDERWENT A PPS PROCEDURE ON (B)(6) 2023. DURING THE PROCEDURE, THE CONFIDENCE NEEDLE BROKE WHEN IT WAS INSERTED INTO A VERTEBRA. THE SURGEON FELT SOMETHING WRONG WITH TAPPING AND CHECKED THE AFFECTED AREA BY IMAGE INTENSIFIER. A FRAGMENT OF THE BROKEN CONFIDENCE NEEDLE WAS OBSERVED. THE SURGEON DECIDED TO LEAVE THE FRAGMENT IN THE PATIENT¿S BODY, AS IT WAS CONSIDERED RISKIER TO REMOVE THE FRAGMENT. ABOUT 2 CM OF THE BROKEN CONFIDENCE NEEDLE REMAINED IN THE PATIENT¿S BODY. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. PATIENT WAS STABLE. THE SURGEON STATED THAT THE PATIENT'S VERTEBRA WAS HARD. THIS REPORT IS FOR A CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES.
Patient problems: Foreign Body In Patient
MAUDE report 16811237 ↗InjuryJun 3, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, THE CONFIDENCE 11G NEEDLE TIP BROKE OFF IN THE PEDICLE DURING MIS SURGERY AND THE TIP WAS UNABLE TO BE REMOVED. THE SURGEON PLACED A SCREW BEHIND THE BROKEN TIP AND IS COMFORTABLE THAT THE TIP WILL NOT MOVE OR CAUSE INJURY TO THE PATIENT. THERE WAS A SURGICAL DELAY OF 10 MINUTES. PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT OUTCOME WAS UNKNOWN. THIS REPORT IS FOR ONE CONF INTRO NDLE, DIA, 11G 6". THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 14595244 ↗MalfunctionJan 20, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, DURING A PERCUTANEOUS PROCEDURE, AFTER CENTERING THE PEDICLE WITH THE TROCAR AND INSERTING THE KIRSHNER WIRE, THE DOCTOR WAS UNABLE TO REMOVE THE OUTER PART OF THE TROCAR. AFTER REPEATED ATTEMPTS TO REMOVE THE TROCAR, IT BROKE. A METAL PART REMAINED ATTACHED TO THE HANDLE THAT THE SURGEON FOUND IN HIS HAND, A SECOND PART CAME OFF WITH THE USE OF A FORCEPS AND A LAST PART REMAINED INSIDE THE VERTEBRAL PEDUNCLE. THE DOCTOR HAD TO SWITCH FROM A PERCUTANEOUS PROCEDURE TO AN OPEN PROCEDURE. TO REMOVE WHAT WAS IN THE BONE HE HAD TO WORK WITH A MOTORIZED BUR AND REMOVE THE FRAGMENTS THAT WERE GRADUALLY BEING CREATED. THERE WAS A TWO (2) HOUR SURGICAL DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO CONSEQUENCE TO THE PATIENT. THIS REPORT INVOLVES ONE (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13308168 ↗MalfunctionJan 20, 2022Break; Material Deformation; Break; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 DURING A LUMBAR ARTHRODESIS FIXATION SURGERY, THE CONFIDENCE SPINAL CEMENT NEEDLE BROKE. PART OF THE NEEDLE REMAINED INSIDE THE BONE (VERTEBRA) WITHOUT CAUSING DAMAGE OR INJURY. THERE IS NO FURTHER INFORMATION AVAILABLE. THIS REPORT IS FOR ONE (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13306761 ↗InjuryOct 22, 2021Break; Break
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021 THAT THE NEEDLE BROKE IN THE BONE OF THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A DIFFERENT ACCESS IN THE PEDICLE IN ORDER TO INTRODUCE THE SCREW. THERE WAS A SURGICAL DELAY. THIS REPORT IS FOR (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 12682701 ↗MalfunctionAug 24, 2021Break; Break
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT AS FOLLOWS: THE SURGEON WAS PERFORMING A L4-S1 ALIF + MIS INSTRUMENTED FUSION ON PATIENT ON THE (B)(6) 2021. WHILE INSERTING A CONFIDENCE NEEDLE TO ASSIST WITH THE GUIDING OF THE SACRAL ALAR ILIAC SCREW (SAI), A GUIDEWIRE WAS PLACED DOWN THE CONFIDENCE NEEDLE BUT UPON REMOVING THE NEEDLE IT WAS NOTED THAT A LARGE SECTION OF THE NEEDLE HAD BROKEN OFF AND REMAINED IN THE PATIENT. A SECOND NEEDLE WAS USED TO PLACE THE SAI SCREW AND WHEN THE GUIDEWIRE WAS REMOVED FROM THE FIRST ATTEMPT THE BROKEN SECTION REMOVED WITH THE GUIDEWIRE. THE SURGEON WAS SATISFIED THAT NO FRAGMENTS REMAINED. THIS REPORT IS FOR A CONFIDENCE NEEDLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12357919 ↗InjuryApr 13, 2021Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, TWO INTRODUCER NEEDLES BROKE INTRAOPERATIVELY DURING STABILIZATION OF LUMBAR VERTEBRAE 3 TO 5. THE NEEDLE WAS BROUGHT INTO THE PEDICLE, THEN THE INSIDES OF THE PEDICLE WERE TAKEN OUT AND A VIPER 2 GUIDEWIRE WAS INSERTED. DURING THE UNSCREWING OF THE OUTER SLEEVE OF THE NEEDLE, A CRACK WAS HEARD AND THE SLEEVE BROKE. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 11657195 ↗InjuryApr 13, 2021Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, TWO INTRODUCER NEEDLES BROKE INTRAOPERATIVELY DURING STABILIZATION OF LUMBAR VERTEBRAE 3 TO 5. THE NEEDLE WAS BROUGHT INTO THE PEDICLE, THEN THE INSIDES OF THE PEDICLE WERE TAKEN OUT AND A VIPER 2 GUIDEWIRE WAS INSERTED. DURING THE UNSCREWING OF THE OUTER SLEEVE OF THE NEEDLE, A CRACK WAS HEARD AND THE SLEEVE BROKE. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS FOR ONE (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 11657170 ↗InjuryDec 14, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, DURING THE REMOVAL OF THE CONFIDENCE INTRODUCER NEEDLE FROM THE SCLEROTIC BONE, THE HANDLE BROKE OFF, WHICH LEFT THE NEEDLE IN-SITU. THE NEEDLE WAS REMOVED BY INCREASING THE EXPOSURE OF TISSUE. FRAGMENTS WERE GENERATED AND REMOVED EASILY. THERE WAS A SURGICAL DELAY OF FIVE (5) TO TEN (10) MINUTES. AN X-RAY WAS TAKEN TO CONFIRM THAT NO FRAGMENTS WERE LEFT IN-SITU. THERE WAS NO REPORTED CONSEQUENCE TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT INVOLVES ONE (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 11002658 ↗MalfunctionJun 23, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 DURING A PERCUTANEOUS PEDICLE SCREW FIXATION, THE CONFIDENCE NEEDLE WAS USED WITH ANOTHER UNKNOWN MANUFACTURE SYSTEM. THE T-SHAPED COMPONENT BROKE AND CAME OFF THE UNIT WHILE IN USE. THE PROCEDURE WAS DELAYED FOR THIRTY (30) MINUTES. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) CONFIDENCE SPINAL CEMENT SYSTEM INTRODUCER NEEDLE, DIAMOND TIP 11G X 6 INCHES. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10184951 ↗InjuryNov 11, 2019Break; Break
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A VERTEBROPLASTY SURGERY WITH THE CONFIDENCE SYSTEM ON (B)(6) 2019, WHEN THE NEEDLES WERE REMOVED EIGHT (8) MINUTES AFTER MIXING, ONE OF THEM BROKE AT THE PEDICLE AND THE FRAGMENT WAS NOT RECOVERED. TWO NEEDLES OF DIFFERENT LOT WERE USED. THE SURGEON REPORTED THAT THERE IS NO PROBLEM BUT OBSERVED THE WEAKNESS OF THE CANNULAS TO DENSE BONE ROTATION, SINCE THESE WERE INSERTED WITH A LOT OF IMPACT. THIS REPORT IS FOR A CONFIDENCE INTRODUCER NEEDLE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Foreign Body In Patient
MAUDE report 9305512 ↗InjurySep 26, 2019Break; Break
I ATTEND VERTEBROPLASTY SURGERY WITH THE CONFIDENCE SYSTEM. WHEN THE NEEDLES REFERENCES 283903611 WERE REMOVED, AFTER 8 MIN AFTER MIXING, ONE OF THEM FRACTURES AT THE RATE OF THE PEDICLE AND THE FRAGMENT WAS NOT RECOVERED. SURGEON REPORTS THAT THERE IS NO PROBLEM BUT OBSERVED THE WEAKNESS OF THE CANNULAS TO DENSE BONE ROTATION, SINCE THESE WERE INSERTED WITH A LOT OF IMPACT. TWO NEEDLES OF DIFFERENT LOT WERE USED AND WERE NOT MARKED, I RETURN BOTH AND THEIR BOXES. ATTACHED PHOTO OF HOW IT IS IN THE PEDICLE AND THE BOXES THAT REMAIN IN CONTAINER. PROCEDURE: VERTEBROPLASTY. THIS COMPLAINT INVOLVES ONE (1) DEVICE.
Patient problems: Foreign Body In Patient
MAUDE report 9124011 ↗InjuryOct 4, 2018Break; Break
IT WAS REPORTED THAT THE REPORTED DEVISE WAS USED TO ILIAC TO CREATE THE BONE HOLE FOR THE INVASIVE OSTEOSYNTHESIS ON (B)(6) 2018. DURING THE SURGERY, THE SURGEON FELT DIFFICULTY TO REMOVE THE CANNULA FROM THE PATIENT¿S BONE HOLE WHEN THE SURGEON REMOVED THE CANNULA AFTER PASSED THE GUIDE WIRE. THE SURGEON USED THE HUMMER TO REMOVE THE CANNULA, BUT THE SURGEON COULD NOT REMOVE. THEN, THE SURGEON USED THE PLIERS, HOWEVER, THE CANNULA WAS BROKEN AND REMAINED IN THE PATIENT¿S BONE. THE SURGEON DECIDED TO KEEP THE BROKEN FRAGMENT INTO THE PATIENT¿S BODY, AND WILL BE REMOVED FROM THE PATIENT¿S BODY DURING THE SCHEDULED SURGERY (2 OR 3 MONTH LATER, SURGERY DATE IS NOT FIXED YET). HERE WAS 60 MIN. SURGICAL DELAY. NO FURTHER INFORMATION WAS PROVIDED BY THE HOSPITAL.
Patient problems: Foreign Body In Patient; Not Applicable
MAUDE report 7935027 ↗MalfunctionJun 15, 2018Break; Break
TWO SEPARATE SHAFTS OF INTRODUCER NEEDLES BROKE OFF INSIDE THE PATIENTS BILATERAL PEDICLES. THE INTRODUCER NEEDLE JUST CRUMPLED AND FURTHER BROKE APART WHEN THE SURGEON TRIED TO DISLODGE THE BROKEN PIECES FROM THE PATIENTS SPINE (BILATERAL L5 PEDICLES).
Patient problems: Local Reaction
MAUDE report 7606870 ↗MalfunctionJun 15, 2018Break; Break
TWO SEPARATE SHAFTS OF INTRODUCER NEEDLES BROKE OFF INSIDE THE PATIENTS BILATERAL PEDICLES. THE INTRODUCER NEEDLE JUST CRUMPLED AND FURTHER BROKE APART WHEN THE SURGEON TRIED TO DISLODGE THE BROKEN PIECES FROM THE PATIENTS SPINE (BILATERAL L5 PEDICLES).
Patient problems: Local Reaction
MAUDE report 7606779 ↗
Other NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE devices
- N/A 305211Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE425 reports
- N/A 305180Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE300 reports
- N/A 305200Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE99 reports
- PneumoDart TM-317TyTek Medical Inc. · Needle, Aspiration And Injection, Disposable1 recall1 report
- TPAK TM-303TyTek Medical Inc. · Needle, Aspiration And Injection, Disposable1 recall0 reports
- N/A 305181Becton, Dickinson And Company · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE32 reports
- SOL-M™ Blunt Fill Needle with 5 Micron Filter 110022Sol-Millennium Medical, Inc. · Needle, Aspiration And Injection, Disposable17 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283903411Medos International Sàrl · NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page