BD Luer-Lok 305617Recalls & reported problems
Becton, Dickinson And Company · SYRINGE 20ML LL TIP CONV PAK · Class 2
All models of BD Luer-Lok (48)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This BD Luer-Lok syringe is a 20 mL piston syringe. Reports mostly describe leaks and contamination, including particles and silicone lubricant in shipments, along with occasional breaks and labeling issues. There are 90 reports, no deaths, and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 90
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 90 adverse event reports that list this exact device (DI 00382903056170), received between Nov 27, 2017 and Jun 17, 2026 — classified as 90 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (36 reports).
10 other recalls cover devices cleared under the same FDA submission (K110771); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash36
- Device Contamination with Chemical or Other Material26
- Contamination /Decontamination Problem10
- Device Markings/Labelling Problem8
- Break4
- Contamination4
- Crack3
- Volume Accuracy Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions58
- No Known Impact Or Consequence To Patient20
- No Consequences Or Impact To Patient8
- No Information2
- No Patient Involvement2
Recent adverse event reports
The 25 most recent of 90 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2026Contamination /Decontamination Problem
IT WAS REPORTED BY THE CUSTOMER THAT DURING THE CHARACTERIZATION, IT TRACED BACK TO A MATERIAL, A PARTICLE WAS CONSISTENT WITH POLYSOME AND BUTADIENE. THEY USED IN TOTAL 80 SYRINGES (8 PACKS OF 10), THEY ONLY DISCOVERED IT IN 2 OF THEM BUT DON'T KNOW THE LOT NUMBER OF THE SPECIFIC ONES WITH THAT PARTICLE. VERBATIM: COMPLAINT VIA PHONE. CASE DESCRIPTION: THE SYRINGE WAS USED FOR A RECONSTITUTION, A PROCEDURE THEY PERFORM. DURING THE CHARACTERIZATION, IT TRACED BACK TO A MATERIAL, A PARTICLE WAS CONSISTENT WITH POLYSOME AND BUTADIENE. THEY USED IN TOTAL 80 SYRINGES (8 PACKS OF 10), THEY ONLY DISCOVERED IT IN 2 OF THEM BUT DON'T KNOW THE LOT NUMBER OF THE SPECIFIC ONES WITH THAT PARTICLE. THE CUSTOMER CALLED TO KNOW WHAT'S THE COMPOSITION OF THE SYRINGE AND WAS TRANSFERRED TO MED INFO, WHERE THEY REACHED THE ENGINEERING TEAM AND ARE GONNA CONTACT HER WITH THE INFORMATION. ADDITIONAL INFORMATION RECEIVED ON 5-JUN-2026: 1. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT DD-MMM-YYYY? THE PARTICULATE WAS CHARACTERIZED (FTIR) AND REPORTED ON (B)(6) 2026 AS CONSISTENT WITH POLYSTYRENE POSSIBLY WITH A SMALL AMOUNT OF BUTADIENE. 2. CAN YOU SHARE ANY PHYSICAL SAMPLE OR PHOTO IF AVAILAB
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25535312 ↗MalfunctionMar 23, 2026Contamination /Decontamination Problem
IT WAS REPORTED BY CUSTOMER THAT HAD 28 UNITS REJECTED FOR SUSPECTED PARTICULATES. VERBATIM: COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED DMR # 671 PO #: (B)(6)_DEFECT MATERIAL DESCRIPTION: SYRINGE TRAY, 20ML BD 10PK (REF. 305617) LOT NUMBER: 5181420 ITEM CODE: 305617 DEFECTIVE QUANTITY: 28 DEFECT DESCRIPTION: SYRINGE LOT (5181420) AND HAD28 UNITS REJECTED FOR SUSPECTED PARTICULATES. OBSERVED DATE: JANUARY 30, 2026 OBSERVED DURING WHICH PROCESS STAGE: COMPOUNDING IR/ DMI NUMBER IF LAUNCHED: DEV-04261 SAMPLE AVAILABILITY FOR SUPPLIER TO INVESTIGATE: NO IS DEFECTIVE EVIDENCE (I.E. PICTURES; TEST RESULTS ETC.) AVAILABLE? YES IS PATIENT IMPACTED? NO IF DEFECT HAS BEEN REPORTED TO THIRD PARTY? NO IF THE DEFECT REPORT FROM XXX CUSTOMER? NO IS THERE A TREND OBSERVED? NO SUPPLIER ACTION AWARENESS/ ATTENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24663545 ↗MalfunctionMar 13, 2026Break
IT WAS REPORTED BY THE CUSTOMER THAT ONE INCIDENT LEADING TO PATIENT ADVERSE REACTION. IN THE FIRST INCIDENT, A GROUP OF SYRINGES BROKE AT THE PHALANGE WHILE BEING STORED IN A FREEZER WITH RECONSTITUTED CEFEPIME 2 G IN 20 ML STERILE WATER. IN THE SECOND INCIDENT, A SYRINGE TIP BROKE OFF IN AN IV INFUSION LINE Y-SITE AFTER ADMINISTRATION OF CEFEPIME 2 G IN 20 ML STERILE WATER. RCC RECEIVED A COMPLAINT VIA WEB. PIR ATTACHED. THERE HAVE BEEN TWO INCIDENTS AT MY INSTITUTION XXXXXXXXXX) WHERE THESE 20 ML LUER LOCK SYRINGES BROKE, WITH ONE INCIDENT LEADING TO PATIENT ADVERSE REACTION. IN THE FIRST INCIDENT, A GROUP OF SYRINGES BROKE AT THE PHALANGE WHILE BEING STORED IN A FREEZER WITH RECONSTITUTED CEFEPIME 2 G IN 20 ML STERILE WATER. IN THE SECOND INCIDENT, A SYRINGE TIP BROKE OFF IN AN IV INFUSION LINE Y-SITE AFTER ADMINISTRATION OF CEFEPIME 2 G IN 20 ML STERILE WATER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24597967 ↗MalfunctionMar 13, 2026Contamination /Decontamination Problem
IT WAS REPORTED BY CUSTOMER THAT BD 20ML SYRINGE LOT 5181420, IR SPECTRUM DEMONSTRATED A 79.2 MATCH TO SILICONE LUBRICANT AND, WHEN COMPARED TO THE BROADER FTIR LIBRARY, A 93.7 MATCH TO SILASTIC Q7 4550, A PEROXIDE CURED MEDICAL GRADE SILICONE ELASTOMER CONSISTENT WITH STOPPER CONSTRUCTION. RCC RECEIVED A COMPLAINT VIA EMAIL. DMR # (B)(4) PO #: (B)(4) DEFECT MATERIAL DESCRIPTION: SYRINGE TRAY, 20ML BD 10PK (REF. 305617) LOT NUMBER: 5181420 ITEM CODE: 305617 DEFECTIVE QUANTITIES: 1 DEFECT DESCRIPTION: BD 20ML SYRINGE LOT 5181420, IR SPECTRUM DEMONSTRATED A 79.2 MATCH TO SILICONE LUBRICANT AND, WHEN COMPARED TO THE BROADER FTIR LIBRARY, A 93.7 MATCH TO SILASTIC Q7 4550, A PEROXIDE CURED MEDICAL GRADE SILICONE ELASTOMER CONSISTENT WITH STOPPER CONSTRUCTION. NO UNITS ARE AVAILABLE TO BE RETURNED. OBSERVED DATE: (B)(6) 2026 OBSERVED DURING WHICH PROCESS STAGE: ILP IR/ DMI NUMBER IF LAUNCHED: DEV-04209 SAMPLE AVAILABILITY FOR SUPPLIER TO INVESTIGATE: NO IS DEFECTIVE EVIDENCE (I.E. PICTURES; TEST RESULTS ETC.) AVAILABLE? YES, IS PATIENT IMPACTED? NO IF DEFECT HAS BEEN REPORTED TO THIRD PARTY? NO IF THE DEFECT REPORT FROM QUVA CUSTOMER? NO IS THERE A TREND OBSERVED? NO".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24597965 ↗MalfunctionMar 13, 2026Break
IT WAS REPORTED BY THE CUSTOMER THAT ONE INCIDENT LEADING TO PATIENT ADVERSE REACTION. IN THE FIRST INCIDENT, A GROUP OF SYRINGES BROKE AT THE PHALANGE WHILE BEING STORED IN A FREEZER WITH RECONSTITUTED CEFEPIME 2 G IN 20 ML STERILE WATER. IN THE SECOND INCIDENT, A SYRINGE TIP BROKE OFF IN AN IV INFUSION LINE Y-SITE AFTER ADMINISTRATION OF CEFEPIME 2 G IN 20 ML STERILE WATER. RCC RECEIVED A COMPLAINT VIA WEB. PIR ATTACHED. THERE HAVE BEEN TWO INCIDENTS AT MY INSTITUTION XXXXXXXXXX) WHERE THESE 20 ML LUER LOCK SYRINGES BROKE, WITH ONE INCIDENT LEADING TO PATIENT ADVERSE REACTION. IN THE FIRST INCIDENT, A GROUP OF SYRINGES BROKE AT THE PHALANGE WHILE BEING STORED IN A FREEZER WITH RECONSTITUTED CEFEPIME 2 G IN 20 ML STERILE WATER. IN THE SECOND INCIDENT, A SYRINGE TIP BROKE OFF IN AN IV INFUSION LINE Y-SITE AFTER ADMINISTRATION OF CEFEPIME 2 G IN 20 ML STERILE WATER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24594480 ↗MalfunctionFeb 13, 2026Contamination
IT WAS REPORTED BY CUSTOMER THAT THEY CONFIRMED SILICONE OIL PRESENCE ASSOCIATED WITH THE SUPPLIED COMPONENT LOT. RCC RECEIVED A COMPLAINT VIA EMAIL. DMR # (B)(4) PO #: (B)(4) DEFECT MATERIAL DESCRIPTION: SYRINGE TRAY, 20ML BD 10PK (REF. 305617) LOT NUMBER: 5195502, 5181420, ITEM CODE: 305617, DEFECTIVE QUANTITY: 2 , DEFECT DESCRIPTION: THIS DMR IS INITIATED TO FORMALLY NOTIFY THE SYRINGE SUPPLIER OF CONFIRMED (B)(4) WITH THE SUPPLIED COMPONENT LOT, IDENTIFIED DURING INTERNAL QUALITY INVESTIGATION. ALTHOUGH NO PARTICULATE MATTER WAS CONFIRMED BY FTIR ANALYSIS OR QC VISUAL ASSESSMENT, THE INVESTIGATION CONCLUDED THE OBSERVED DEFECT WAS SILICONE OIL. WE WISH TO REQUEST FOR SILICON OIL ADDITION SPECIFICATION AND THE SAFETY JUSTIFICATION REPORT/DATA (MEDICAL GRADE/ INERT, BIOCOMPATIBLE, ETC) OBSERVED DATE: (B)(6) 2026 OBSERVED DURING WHICH PROCESS STAGE: ILP IR/ DMI NUMBER IF LAUNCHED: (B)(4) SAMPLE AVAILABILITY FOR SUPPLIER TO INVESTIGATE: NO IS DEFECTIVE EVIDENCE (I.E. PICTURES; TEST RESULTS ETC.) AVAILABLE? YES, IS PATIENT IMPACTED? NO, IF DEFECT HAS BEEN REPORTED TO THIRD PARTY? NO , IF THE DEFECT REPORT FROM QUVA CUSTOMER? NO , IS THERE A TREND OBSERVED? NO , SUPPLIER ACTION AWARE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24334275 ↗MalfunctionFeb 13, 2026Contamination
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24334274 ↗MalfunctionFeb 10, 2026Device Contamination with Chemical or Other Material
DEFECT DESCRIPTION WAS INITIATED FOR AQL FAILURE FOR PARTICULATES. THE FTIR RESULTS CAME BACK WITH A 78 PERCENT CORRELATION TO SILICONE LUBRICANT. NO PATIENT HARM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24305091 ↗MalfunctionFeb 9, 2026Leak/Splash
IT WAS REPORTED BY CUSTOMER THAT BATCHING SYRINGES LEAKING DRUG AT THE PLUNGER AREA OF THE SYRINGE. RCC RECEIVED A COMPLAINT VIA EMAIL. DATE: (B)(6) 2026 COMPLAINT NUMBER: (B)(4) ACCOUNT NUMBER: (B)(4) ITEM NUMBER: 308-305617 ,LOT# 5129980, EXPIRY 2030-04-30 LOT# 5195506 , EXPIRY 2030-06-30 SAMPLE AVAILABLE: YES ITEM DESCRIPTION: 20ML LUER-LOK TIP STERILE SYRINGE PACK INCIDENT DESCRIPTION: AS PER ACCOUNT, BATCHING SYRINGES LEAKING DRUG AT THE PLUNGER AREA OF THE SYRINGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24297522 ↗MalfunctionFeb 9, 2026Leak/Splash
IT WAS REPORTED BY CUSTOMER THAT BATCHING SYRINGES LEAKING DRUG AT THE PLUNGER AREA OF THE SYRINGE. RCC RECEIVED A COMPLAINT VIA EMAIL. DATE: 22.01.2026 COMPLAINT NUMBER: (B)(4), ACCOUNT NUMBER: (B)(4), ITEM NUMBER: 308-305617, LOT # / SERIAL #: LOT# 5129980, EXPIRY 2030-04-30 LOT# 5195506 , EXPIRY 2030-06-30 SAMPLE AVAILABLE: YES ITEM DESCRIPTION: 20ML LUER-LOK TIP STERILE SYRINGE PACK INCIDENT DESCRIPTION: AS PER ACCOUNT, BATCHING SYRINGES LEAKING DRUG AT THE PLUNGER AREA OF THE SYRINGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24297521 ↗MalfunctionJan 30, 2026Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24222455 ↗MalfunctionJan 30, 2026Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24222453 ↗MalfunctionDec 13, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23798686 ↗MalfunctionDec 10, 2025Leak/Splash; Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770028 ↗MalfunctionDec 10, 2025Leak/Splash; Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23770027 ↗MalfunctionNov 10, 2025Leak/Splash; Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23518629 ↗MalfunctionOct 24, 2025Leak/Splash; Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23378644 ↗MalfunctionJun 15, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
NO ADDITIONAL INFORMATION
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22208681 ↗MalfunctionJun 3, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT BD SYRINGE 20ML LL TIP CONV PAK HAD FOREIGN MATTER. IT WAS REPORTED BY THE CUSTOMER THAT BLACK PARTICLE SEEN ON SOME SYRINGES. VERBATIM: BLACK PARTICLE SEEN ON SOME SYRINGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22133476 ↗MalfunctionMar 25, 2025Leak/Splash; Leak/Splash
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21688483 ↗MalfunctionFeb 26, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD SYRINGE 20ML LL TIP CONV PAK LEAKED. IT WAS REPORTED BY CUSTOMER THAT LIQUID APPEARED TO HAVE LEAKED BEHIND SYRINGE PLUNGER ON A NUMBER OF SYRINGES IN THIS LOT. 11 SYRINGES CONTAINING CEFAZOLIN 2 G/20 ML HAVE BEEN SEQUESTERED. VERBATIM: RCC RECEIVED A COMPLAINT VIA COMMUNITY. PIR ATTACHED. LIQUID APPEARED TO HAVE LEAKED BEHIND SYRINGE PLUNGER ON A NUMBER OF SYRINGES IN THIS LOT. 11. SYRINGES CONTAINING CEFAZOLIN 2 G/20 ML HAVE BEEN SEQUESTERED. PRODUCT CODE: 305617.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21472313 ↗MalfunctionFeb 3, 2025Break; Device Markings/Labelling Problem; Break; Device Markings/Labelling Problem
NO ADDITIONAL INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21293874 ↗MalfunctionJan 13, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD SYRINGE 20ML LL TIP CONV PAK LEAKED. IT WAS REPORTED BY THE CUSTOMER THAT FOUND TWO ADDITIONAL 20 ML LEAKING SYRINGES FROM THE SAME LOT (REF (B)(4), LOT: 4278835). VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. I FOUND TWO ADDITIONAL 20 ML LEAKING SYRINGES FROM THE SAME LOT (REF (B)(4) LOT: 4278835), SO I ALSO INCLUDED THESE SYRINGES IN THE TOTAL I AM SHIPPING TO YOU. I AM NOW SENDING TOTAL OF 19 SYRINGES FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21143082 ↗MalfunctionDec 31, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD SYRINGE 20ML LL TIP CONV PAK LEAKED BATCH#: 4278835 AND AN UNKNOWN BATCH. IT WAS REPORTED BY THE CUSTOMER THAT LEAKING AT THE BACK OF THE PLUNGER. VERBATIM: (B)(6). RCC RECEIVED A COMPLAINT VIA PHONE. PIR ATTACHED. LEAKING AT THE BACK OF THE PLUNGER. 309605 LOT 4222762 TOTAL OF 18, 305617 LOT 4278835 TOTAL 17, 309680 LOT 1242178 TOTAL 4, XXXX. HAS SAMPLES AVAILABLE FOR INVESTIGATION PLEASE SEND PACKAGE THE SYRINGES HAS MEDICATION IN THEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21048455 ↗MalfunctionDec 13, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD SYRINGE 20 ML LL TIP CONV PAK LEAKED. IT WAS REPORTED BY CUSTOMER THAT THANK YOU FOR THE LABEL. SINCE THE INITIAL REPORT WE HAVE AN ADDITIONAL LOT THAT ALSO HAS BEEN AFFECTED. LOT 3333318. LEAKAGE. VERBATIM: COMPLAINT RECEIVED VIA EMAIL(S) ATTACHED. THANK YOU FOR THE LABEL. SINCE THE INITIAL REPORT WE HAVE AN ADDITIONAL LOT THAT ALSO HAS BEEN AFFECTED. LOT 3333318. LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20926569 ↗
FDA clearance and classification
Cleared through Special 510(k) K110771 on May 13, 2011 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “Bd Single Use, Hypodermic Syringe”. The FDA review took 53 days. FDA record ↗
The FDA regulates product code FMF (syringe, piston) under 21 CFR 880.5860 as a Class II device. 21 CFR 880.5860 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K110771. One submission often covers many models, so check the product and lot details in each record.
- Z-2614-2026 · Apr 28, 2026 · Open, ClassifiedSYRINGE LUER LOK 10CC-200/BOX BD Product Code: 302995 Mentor Product Code: B9604 Syringe, Piston
- Z-2750-2017 · Sep 16, 2016 · TerminatedBD 3ml Syringe Luer-Lok" Tip with BD PrecisionGlide" Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #309580 The BD Single Use, Hypodermic Syringe is intended for use by health care prof
- Z-2747-2017 · Sep 16, 2016 · TerminatedBD 5ml Syringe Luer-Lok" Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 40mm) Catalog #305062
- Z-2753-2017 · Sep 16, 2016 · TerminatedBD 3ml Syringer Luer-Lok Tip Catalog #309604 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
- Z-2751-2017 · Sep 16, 2016 · TerminatedBD 3ml Syringe Luer-Lok" Tip Catalog 309657 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injection
- Z-2752-2017 · Sep 16, 2016 · TerminatedBD 10mL Syringe Luer-Lok Tip Catalog number 301997 The BD Single Use, Hypodermic Syringe is intended for use by health care professionals for general purpose fluid aspiration/injec
- Z-2749-2017 · Sep 16, 2016 · TerminatedBD 3ml Syringe Luer-Lok" Tip with BD PrecisionGlide" Needle 23G x 1 (0.6mm x 25mm) Catalog #309571 The BD Single Use, Hypodermic Syringe is intended for use by health care professi
- Z-1937-2016 · Mar 25, 2016 · TerminatedHenry Schein Single Use Syringe Luer-Lock, 10 mL, Intended for use by health care professionals for general purpose fluid aspiration/injection
- Z-1936-2016 · Mar 25, 2016 · TerminatedHenry Schein Single Use Syringe Luer-Lock, 10 mL, Intended for use by health care professionals for general purpose fluid aspiration/injection
- Z-1055-2015 · Dec 19, 2014 · TerminatedBD 10 mL Syringe BD Luer-Lok Tip with Blunt Plastic Cannula, Not for Skin Injection ; Catalog #303348, BD 10 mL Syringe Luer-Lok Tip with Blunt Fill Needle 18G x 1 1/2 (1.2mm x 4
Common questions about the BD Luer-Lok 305617
Has the BD Luer-Lok 305617 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00382903056170), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the BD Luer-Lok 305617?
The FDA's MAUDE database lists 90 adverse event reports for this device, received between Nov 27, 2017 and Jun 17, 2026. In those reports, the most common specific device problems are “Leak/Splash” (36), “Device Contamination with Chemical or Other Material” (26) and “Contamination /Decontamination Problem” (10). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD Luer-Lok 305617?
None of the reports are classified as deaths, 0 as injuries and 90 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BD Luer-Lok 305617 FDA approved?
It was cleared through Special 510(k) K110771 on May 13, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the BD Luer-Lok 305617, and is it still sold?
The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD Luer-Lok, model 305617). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Syringe, piston devices
- BD Luer-Lok 309657Becton, Dickinson And Company · Syringe, piston1,254 reports
- BD Luer-Lok 309653Becton, Dickinson And Company · Syringe, piston928 reports
- BD Luer-Lok 309628Becton, Dickinson And Company · Syringe, piston852 reports
- N/A 302995Becton, Dickinson And Company · Syringe, piston1 recall527 reports
- BD Plastipak Syringe 300865Becton, Dickinson And Company · Syringe, piston505 reports
- N/A 302830Becton, Dickinson And Company · Syringe, piston455 reports
- N/A 302832Becton, Dickinson And Company · Syringe, piston416 reports
- BD Ultra-Fine 328468Becton, Dickinson and Company Inc · Syringe, piston319 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page