BD Ultra-Fine 320109Recalls & reported problems
Becton Dickinson Pen Limited · BD Ultra-Fine™ Pen Needles, 8mm 31G, 5bevel, TW · Class 2
All models of BD Ultra-Fine (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This BD Ultra-Fine pen needle is a needle used for injections with insulin pens. Many of its 208 reports describe needles that clog or fail to deliver medication, and some describe needle parts breaking off, with some leaving fragments in the body. The reports include 14 injuries and no deaths. No recalls name this device. No reports came in the last 12 months, versus 21 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 208
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 208 adverse event reports that list this exact device (DI 00382903201099), received between May 11, 2018 and Oct 15, 2025 — classified as 14 injuries, 194 malfunctions.
The most frequently reported specific device problem is “Failure to Deliver” (82 reports).
1 other recall cover devices cleared under the same FDA submission (K213478); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Deliver82
- Break39
- Misconnection20
- Defective Device18
- Leak/Splash14
- Complete Blockage8
- Device Markings/Labelling Problem8
- Missing Information7
Patient problems
- No Clinical Signs, Symptoms or Conditions165
- Foreign Body In Patient15
- No Consequences Or Impact To Patient14
- Needle Stick/Puncture6
- No Known Impact Or Consequence To Patient5
- Insufficient Information3
- No Information2
- Bruise/Contusion1
Recent adverse event reports
The 25 most recent of 208 reports, in the reporter's own words as published by the FDA.
InjuryOct 15, 2025Break; Break
CONSUMER REPORTED THAT THE NEEDLE BROKE OFF IN HIS ABDOMEN WHILE DOING HIS INJECTION. HE STATED THAT HIS DOG JUMPED ON HIM WHILE HE WAS INJECTING, AND CAUSED THE NEEDLE TO BREAK OFF. CONSUMER STATED THAT HE WENT TO THE ER, AND THEY DID AN X-RAY, CT-SCAN, AND ULTRASOUND, BUT COULD NOT LOCATE THE NEEDLE. HE ALSO STATED THAT HE COULD FEEL THE NEEDLE POLKING HIS SKIN WHEN HE TOUCHES THE AREA. CONSUMER DOES NOT RE-USE. LOT #: 2277264. CATALOG #: 320109. DATE OF EVENT: UNKNOWN. SAMPLES: DISCARDED.
Patient problems: Foreign Body In Patient
MAUDE report 23299704 ↗MalfunctionSep 2, 2025Off-Label Use; Failure to Deliver; Off-Label Use; Failure to Deliver
CONSUMER REPORTED FINDING THE NEEDLES CLOG DURING THE INJECTION. CALLER DOES NOT COMPLETE A FLOW CHECK. INFORMED CALLER OF PROPER PLACEMENT OF NON PATIENT END TO PEN AND TO COMPLETE A FLOW CHECK. DC LOT # 4275005 CATALOG# 320109 DATE OF EVENT UNKNOWN SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22946287 ↗MalfunctionJun 13, 2025Failure to Deliver; Failure to Deliver
THE CONSUMER REPORTED PEN NEEDLE CLOG WHEN TAKING INJECTIONS. STATED, SOMETIMES SHE GETS A CLOG WHEN PRIMING. LOT: 4177049, 4051832, 4128011, 4247011. CATALOG: 8MM PEN NEEDLES. DATE OF EVENT: UNKNOWN. SAMPLES: NO (B)(6).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22205195 ↗MalfunctionApr 28, 2025Failure to Deliver; Failure to Deliver
THE CONSUMER REPORTED PEN NEEDLE CLOG WHEN PRIMING HIS PEN NEEDLES. CONSUMER IS NEW TO INJECTIONS AND IS PRIMING THE 2 UNITS LIKE HE WAS TOLD. 12 PEN NEEDLES AFFECTED LOT: 4198054 CATALOG: 320109 DATE OF EVENT: UNKNOWN SAMPLES: NO CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21918946 ↗MalfunctionApr 25, 2025Failure to Deliver; Failure to Deliver
CONSUMER REPORTED FINDING 6 PEN NEEDLES FROM THIS BOX THAT CLOGGED DURING THE FLOW CHECK. INFORMED CALLER OF THE NON-PATIENT END PLACEMENT. DC LOT # 4205051. CATALOG# 320109. DATE OF EVENT. SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21915001 ↗MalfunctionMar 28, 2025Break; Break
CONSUMER REPORTED THE NON-PATIENT END BROKE OFF IN THE PEN AFTER INJECTION. 1 PEN NEEDLE AFFECTED LOT: 3353093 CATALOG: 320109 DATE OF EVENT: 2025-11-03 SAMPLES: YES CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21720631 ↗MalfunctionMar 21, 2025Failure to Deliver; Failure to Deliver
CONSUMER REPORTED PEN NEEDLE CLOG WHEN PRIMING BEFORE INJECTION. STATED, HER HANDS SHAKE A LOT DUE TO A MEDICAL CONDITION UNRELATED TO COMPLAINT. LOT: 4051832, 3011401. CATALOG: 8MM, 31G, SHORT PEN NEEDLES. DATE OF EVENT: UNKNOWN. SAMPLES: YES, (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21664683 ↗MalfunctionMar 7, 2025Failure to Deliver; Failure to Deliver
CONSUMER REPORTED NEEDLE CLOG DURING INJECTION. SHE STATED THAT THERE IS NO INSULIN FLOW. CONSUMER DID NOT HAVE THE CATALOG NUMBER BUT STATED THAT SHE IS USING THE 8MM PEN NEEDLES. LOT #: 3011401. CATALOG #: UNKNOWN. DATE OF EVENT: UNKNOWN. SAMPLES: AVAILABLE - SENDING MAIL KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21552929 ↗MalfunctionFeb 24, 2025Failure to Deliver; Failure to Deliver
CONSUMER REPORTED FOUND 1 PEN NEEDLE TO CLOG DURING INJECTION. TRIED INJECTING 4 TIMES WITH THIS PEN NEEDLE. INFORMED CALLER OF NON PATIENT END PLACEMENT AND TO COMPLETE A FLOW CHECK WITH ALL INJECTIONS LOT # 4030373 CATALOG# 320109 DATE OF EVENT JANUARY 30 2025 SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21446393 ↗MalfunctionFeb 18, 2025Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
CONSUMER REPORTED AFTER INJECTING 2 DIFFERENT NEEDLES NOTICED TWO DOTS AT EACH INJECTION SITE. USING FINGER TO SCRAPE SKIN, A VERY TINY PIECE OF METAL CAME OUT OF THE SITE. SHE SAVED ONE OF THE PIECE AND PLACED IT WITHIN SCOTCH TAPE FOR OUR REVIEW. DENIED REUSE. WILL ALSO SEND UNUSED SAMPLES FROM THIS BOX. DC LOT # 4093071 CATALOG# 320109 DATE OF EVENT UNKNOWN SAMPLE STATUS AWAITING SAMPLE.
Patient problems: Foreign Body In Patient
MAUDE report 21392846 ↗MalfunctionFeb 10, 2025Break; Misconnection; Break; Misconnection
CONSUMER REPORTED - PATIENT PRESENTED TO CLINIC TO REPORT PEN NEEDLES WERE MALFUNCTIONING. HE STATES HE WOULD PULL THE NEEDLE CAP OFF AND THE NEEDLE WAS STICKING IN THE CAP. HE TRIED TO SCREW THE NEEDLE ON WITHOUT SUCCESS. HE CHANGED TO A DIFFERENT BOX OF PEN NEEDLES AND HAD NO FURTHER PROBLEMS PER PATIENT HE WASTED APPROXIMATELY 12 NEEDLES THAT DIDN'T WORK. REF REPORTS: MW5164028, MW5164029, MW5164030, MW5164031, MW5164032, MW5164033, MW5164034, MW5164035, MW5164036, MW5164037, MW5164038, MW5164039. LOT#: 3312545. CATALOG#: UNKNOWN, PEN NEEDLE 31G 8MM, DATE OF EVENT: 25-NOV-2024, SAMPLE STATUS: DISCARD, INITIAL REPORTER: PHARMACIST, MDR REPORT KEY: 21036023, DATE RECEIVED: 27-DEC-2024, MDR DATE ENTERED: 31-DEC-2024, ALL 12 NEEDLES RECORDED ON DIFFERENT MW REPORT NUMBERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21344155 ↗MalfunctionJan 8, 2025Packaging Problem; Packaging Problem
CONSUMER REPORTED THAT THE TEAR DROP LABELS ARE MISSING FROM SEVERAL OF THE PEN NEEDLES. LOT#: 3290758, CATALOG#: 320109, DATE OF EVENT: UNKNOWN, SAMPLES: AVAILABLE - SENDING MAIL KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21103146 ↗MalfunctionDec 13, 2024Break; Break
CONSUMER REPORTED WHEN INJECTING WITH 2 DIFFERENT NEEDLES FROM THIS BOX, THE NEEDLE BROKE INTO SITE. CONSUMER REPORTED FIRST NEEDLE BREAK WAS ON (B)(6) 2024. THE SECOND NEEDLE BREAK WAS ON (B)(6) 2024. DENIED REUSE OF NEEDLES. XRAY COMPLETED (B)(6) 2024. DOCTORS VISIT (B)(6) 2024. LOT # 3353094. CATALOG# 320109 SAMPLE STATUS AWAITING SAMPLE PLASTIC HUB. NEEDLE STILL IN HER SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20929460 ↗MalfunctionDec 13, 2024Break; Break
CONSUMER REPORTED WHEN INJECTING WITH 2 DIFFERENT NEEDLES FROM THIS BOX, THE NEEDLE BROKE INTO SITE. CONSUMER REPORTED FIRST NEEDLE BREAK WAS ON 10.29.2024. THE SECOND NEEDLE BREAK WAS ON 11.01.2024. DENIED REUSE OF NEEDLES. XRAY COMPLETED 11.11.2024 DOCTORS VISIT 11.07.2024 LOT # 3353094 CATALOG # 320109 SAMPLE STATUS AWAITING SAMPLE PLASTIC HUB. NEEDLE STILL IN HER SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20929409 ↗MalfunctionDec 13, 2024Misconnection; Misconnection
RECEIVED VOICEMAIL REQUESTING CALL BACK REGARDING ISSUE CONSUMER IS HAVING WITH PEN NEEDLES. RETURNED CALL. CONSUMER STATED, THE NON PATIENT END IS BENT BEFORE ATTEMPTING INJECTION. STATED, SHE CANNOT ATTACH THE PEN NEEDLE TO THE PEN. STATED, HALF THE BOX WAS AFFECTED. LOT: 4072015. CATALOG: 320109. DATE OF EVENT: UNKNOWN. SAMPLES: YES CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20926632 ↗InjuryDec 12, 2024Break; Patient-Device Incompatibility; Break; Patient-Device Incompatibility
CONSUMER REPORTED THAT THE NEEDLES BEND DURING INJECTION. CONSUMER ALSO REPORTED THAT ONE NEEDLE BROKE OFF IN THE INJECTION SITE. CONSUMER STATED THAT SHE WENT TO THE ER TO HAVE THE NEEDLE REMOVED. NEEDLE WAS REMOVED BY THE DOCTOR. CONSUMER SAID SHE DID NOT HAVE ANY PAIN OR DISCOMFORT. CONSUMER DISCARDED THE BOX BUT WAS ABLE TO PROVIDE THE LOT # FROM THE TEAR DROP LABEL. LOT #: 9121953 CATALOG #: UNKNOWN DATE OF EVENT: 11-18-24 SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20918942 ↗MalfunctionDec 10, 2024Failure to Deliver; Failure to Deliver
CONSUMER LEFT VOICEMAIL REPORTING THAT SHE IS HAVING ISSUES WITH HER 8MM PEN NEEDLES. I RETURNED CONSUMER'S CALL AND LEFT VOICEMAIL FOR RETURN CALL. CONSUMER RETURNED MY CALL, STATED THAT THERE IS NO INSULIN FLOW DURING PRIMING. LOT #: 3157363, CATALOG #: 320109, DATE OF EVENT: UNKNOWN, SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20894202 ↗MalfunctionDec 6, 2024Off-Label Use; Failure to Deliver; Off-Label Use; Failure to Deliver
CONSUMER REPORTED WHEN REUSING AND STORES PEN NEEDLE ON THE PEN THE NEEDLE GETS CLOGGED. INFORMED CALLER NOT TO REUSE WITH REASONS. DC LOT # 4072015, CATALOG# 320109, DATE OF EVENT UNKNOWN, SAMPLE STATUS AWAITING SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20864518 ↗MalfunctionNov 29, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED NON PATIENT END NEEDLES BENT. CONSUMER ALSO REPORTED NEEDLE CLOG DURING PRIMING. SHE STATED THAT THERE WAS NO INSULIN FLOW, THEN SHE NOTICED THAT THE NON-PATIENT END NEEDLE WAS BENT. CONSUMER DOES NOT RE-USE. LOT #: 4038404. CATALOG #: 320109. DATE OF EVENT: UNKNOWN. SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20812336 ↗MalfunctionNov 29, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem
CONSUMER STATED, THERE WERE ONLY 15 PEN NEEDLES IN HER NEW BOX. ALSO STATED, THERE WERE "LANCETS" IN HER PEN NEEDLE BOX. STATED, THE BOX WAS SEALED WHEN SHE PURCHASED IT. LOT: 3108801, CATALOG: 320109, DATE OF EVENT: 2024-18-10, SAMPLES: NO CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20812321 ↗MalfunctionNov 13, 2024Defective Device; Defective Device
LOT: 0316718. FROM: SENT:(B)(6) , 2024 7:19 AM. TO: PRODUCTCOMPLAINTS <[email protected]> SUBJECT: BD PEN NEEDLES. HELLO, I AM A DIABETIC OF 11 YEARS (TYPE 1) AND I HAVE ALWAYS USED YOUR NEEDLES FOR MY PENS. MY MOST RECENT ALLOTMENT OF NEEDLES HAS SO FAR HAD ABOUT A 50% DEFECT RATE. THEY ARE MISSING THE NEEDLE PART. THE STORE WILL NOT RETURN THEM AND SAID I NEED TO ADDRESS IT WITH THE MANUFACTURER. THE ARE BD STERILE NEEDLE .25 X 8MM. 31G X 5/16. THEY ARE IN A BOX OF 100. THE LOT NUMBER IS 0316718 AND THE EXPIRATION DATE IS 2025-11-30. I WOULD LIKE THEM TO BE REPLACED AND SHIPPED TO ME. THE ADDRESS IS AS FOLLOWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20675442 ↗MalfunctionOct 30, 2024Failure to Deliver; Failure to Deliver
CONSUMER REPORTED FINDING THE NON PATIENT END TO CLOG DURING INJECTION BUT NOT THE FLOW CHECK. CONSUMER REPORTED FINDING THE NON PATIENT END NEEDLE TO BE MISSING ON 2 PEN NEEDLES LOT 4038404 LOT # 4038404. LOT # 3283933. CATALOG# 320109. DATE OF EVENT UNKNOWN. SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20568180 ↗MalfunctionOct 30, 2024Failure to Deliver; Defective Device; Failure to Deliver; Defective Device
CONSUMER REPORTED FINDING THE NON PATIENT END TO CLOG DURING INJECTION BUT NOT THE FLOW CHECK. CONSUMER REPORTED FINDING THE NON PATIENT END NEEDLE TO BE MISSING ON 2 PEN NEEDLES LOT 4038404. LOT # 4038404. LOT # 3283933. CATALOG# 320109. DATE OF EVENT UNKNOWN. SAMPLE STATUS DISCARD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20568127 ↗MalfunctionOct 24, 2024Leak/Splash; Failure to Deliver; Leak/Splash; Failure to Deliver
CONSUMER REPORTED INSULIN LEAKING THROUGH THE NEEDLE. CONSUMER STATED THAT HE IS NEW TO USING THE PEN NEEDLES. AFTER CAREFUL REVIEW OF HOW TO USE, THE CONSUMER STATED THAT HE DOES NOT REMOVE THE INNER COVER, SO THE LEAKAGE IS COMING FROM UNDER THE INNER COVER. HE SAID THAT HE WAS NOT INSTRUCTED ON HOW TO USE, SO HE JUST ASSUMED THAT HE ONLY NEEDED TO REMOVE THE CLEAR OUTER COVER. LOT#: 4016300, CATALOG#: 320109, DATE OF EVENT: UNKNOWN, SAMPLES: DISCARDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20525437 ↗MalfunctionOct 18, 2024Failure to Deliver; Failure to Deliver
SPOUSE OF CONSUMER REPORTED NEEDLE CLOG DURING PRIMING. HE STATED THAT THERE IS NO INSULIN FLOW. LOT #: 3347042. CATALOG #: 320109. DATE OF EVENT: UNKNOWN. SAMPLES: AVAILABLE - SENDING MAIL KIT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20482343 ↗
FDA clearance and classification
Cleared through 510(k) K213478 on Jun 28, 2022 (decision: Substantially Equivalent). Submitted by Becton, Dickinson and Company as “BD Pen Needle”. The FDA review took 242 days. FDA record ↗
The FDA regulates product code FMI (needle, hypodermic, single lumen) under 21 CFR 880.5570 as a Class II device. 21 CFR 880.5570 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K213478. One submission often covers many models, so check the product and lot details in each record.
- Z-3180-2026 · Jul 26, 2023 · Open, ClassifiedEWO_02513 (Catalog number 320122) For Investigational Use Only Nano Un-Cannulated Demo Samples Quantity 100
Common questions about the BD Ultra-Fine 320109
Has the BD Ultra-Fine 320109 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00382903201099), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the BD Ultra-Fine 320109?
The FDA's MAUDE database lists 208 adverse event reports for this device, received between May 11, 2018 and Oct 15, 2025. In those reports, the most common specific device problems are “Failure to Deliver” (82), “Break” (39) and “Misconnection” (20); the most common patient problems are “Foreign Body In Patient” (15), “Needle Stick/Puncture” (6) and “Bruise/Contusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD Ultra-Fine 320109?
None of the reports are classified as deaths, 14 as injuries and 194 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the BD Ultra-Fine 320109 FDA approved?
It was cleared through 510(k) K213478 on Jun 28, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the BD Ultra-Fine 320109, and is it still sold?
The FDA's device database lists Becton Dickinson Pen Limited as the company responsible for it (brand BD Ultra-Fine, model 320109). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Needle, hypodermic, single lumen devices
- BD Nano 2nd Gen 320550Becton Dickinson Pen Limited · Needle, hypodermic, single lumen1,892 reports
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,768 reports
- BD Nano Ultra-Fine 320122Becton Dickinson Pen Limited · Needle, hypodermic, single lumen566 reports
- N/A 305106Becton, Dickinson And Company · Needle, hypodermic, single lumen1 recall465 reports
- N/A 305110Becton, Dickinson And Company · Needle, hypodermic, single lumen418 reports
- BD Safe-Clip 328235Becton, Dickinson and Company Inc · Needle, hypodermic, single lumen268 reports
- BD AutoShield Duo 329515Becton, Dickinson And Company · Needle, hypodermic, single lumen200 reports
- BD Safety-Lok 305901Becton, Dickinson And Company · Needle, hypodermic, single lumen170 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.