Skip to content

BD Nano 2nd Gen 320550Recalls & reported problems

Becton Dickinson Pen Limited · BD Nano 2nd Gen Pen Needles, 4mm, 32G, 5bevel, UTW · Class 2

All models of BD Nano 2nd Gen (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BD Nano 2nd Gen pen needle, used with insulin pens, has 1,892 adverse event reports on record, with no deaths and 23 injuries. Most reports describe medication not flowing through the needle, with needles clogging during injection, not attaching to the pen, and sometimes breaking or leaking. There were no reports in the last 12 months, compared with 123 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,044
Deaths reported
0
Injuries reported
23
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,044 adverse event reports that list this exact device (DI 00382903205509), received between Sep 12, 2019 and Sep 30, 2026 — classified as 23 injuries, 2,021 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver” (1,340 reports).

2 other recalls cover devices cleared under the same FDA submission (K212015); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 23 reportsDec 2023: 18 reportsJan 2024: 27 reportsFeb 2024: 29 reportsMar 2024: 38 reportsApr 2024: 29 reportsMay 2024: 35 reportsJun 2024: 24 reportsJul 2024: 24 reportsAug 2024: 22 reportsSep 2024: 20 reportsOct 2024: 26 reportsNov 2024: 19 reportsDec 2024: 14 reportsJan 2025: 7 reportsFeb 2025: 7 reportsMar 2025: 26 reportsApr 2025: 7 reportsMay 2025: 19 reportsJun 2025: 9 reportsJul 2025: 4 reportsAug 2025: 5 reportsSep 2025: 6 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 152 reportsOct 2026: 0 reports202420252026152
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deliver1,340
  • Misconnection171
  • Complete Blockage129
  • Break122
  • Leak/Splash109
  • Defective Device91
  • Failure to Prime48
  • Defective Component39

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,764
  • No Consequences Or Impact To Patient158
  • No Known Impact Or Consequence To Patient51
  • Pain36
  • Hyperglycemia33
  • Foreign Body In Patient17
  • Bruise/Contusion8
  • Skin Inflammation/ Irritation8

Recent adverse event reports

The 25 most recent of 2,044 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954932 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954826 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954790 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954703 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954654 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954533 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954025 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953587 ↗
  • MalfunctionSep 30, 2026Misconnection; Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950723 ↗
  • MalfunctionSep 30, 2026Misconnection; Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950543 ↗
  • MalfunctionSep 30, 2026Misconnection; Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950094 ↗
  • MalfunctionSep 30, 2026Misconnection; Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949991 ↗
  • MalfunctionSep 30, 2026Misconnection; Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949900 ↗
  • MalfunctionSep 30, 2026Misconnection; Failure to Deliver

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949338 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Consumer Reported found about 2 dozen pen needles from two different lot number. Unknown amount on each lot #. Needles would not prime. noticed after removed from pen the non patient end needle to be bent. DC Lot # 3017179 Lot # 3025239Catalog# unknown - 32g 4mm 2nd gen Date of Event Unknown Sample Status Awaiting sample

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943940 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Consumer Reported found about 2 dozen pen needles from two different lot number. Unknown amount on each lot #. Needles would not prime. noticed after removed from pen the non patient end needle to be bent. DC.Lot # 3017179.Lot # 3025239.Catalog# unknown - 32g 4mm 2nd gen. Date of Event Unknown. Sample Status Awaiting sample.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943781 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Consumer Reported found about 2 dozen pen needles from two different lot number. Unknown amount on each lot #. Needles would not prime. noticed after removed from pen the non patient end needle to be bent. DC Lot # 3017179 Lot # 3025239Catalog# unknown - 32g 4mm 2nd gen Date of Event Unknown Sample Status Awaiting sample

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943671 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Consumer Reported found about 2 dozen pen needles from two different lot number. Unknown amount on each lot #. Needles would not prime. noticed after removed from pen the non patient end needle to be bent. DC Lot # 3017179 Lot # 3025239Catalog# unknown - 32g 4mm 2nd gen Date of Event Unknown Sample Status Awaiting sample.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943594 ↗
  • MalfunctionSep 30, 2026Defective Device

    Consumer left voicemail reporting defective needles. Rep returned his call and left voicemail requesting return call. Consumer returned call and reported that needle is protruding through the green inner shield. He stated that he has 17 pen needles where the needle came through the inner shield. Consumer stated that he stuck his finger with the needles a few times.I asked consumer about re-shielding after use and he said that he does not re-shield with the green cover but he prepares the insulin and puts it into the refrigerator for use when it is time to take his injection. Consumer is new to using pen needles and did not seem very sure of how to use. He stated that he does not re-use.Lot #: 3276406.Catalog #: 320550.Date of Event: unknown.Samples available: sending mail kit.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943420 ↗
  • MalfunctionSep 30, 2026Defective Device

    Consumer left voicemail reporting defective needles. Rep returned his call and left voicemail requesting return call. Consumer returned call and reported that needle is protruding through the green inner shield. He stated that he has 17 pen needles where the needle came through the inner shield. Consumer stated that he stuck his finger with the needles a few times.I asked consumer about re-shielding after use and he said that he does not re-shield with the green cover but he prepares the insulin and puts it into the refrigerator for use when it is time to take his injection. Consumer is new to using pen needles and did not seem very sure of how to use. He stated that he does not re-use.Lot #: 3276406Catalog #: 320550Date of Event: unknownSamples available: sending mail kit.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943419 ↗
  • MalfunctionSep 30, 2026Defective Device

    Consumer left voicemail reporting defective needles. Rep returned his call and left voicemail requesting return call. Consumer returned call and reported that needle is protruding through the green inner shield. He stated that he has 17 pen needles where the needle came through the inner shield. Consumer stated that he stuck his finger with the needles a few times.I asked consumer about re-shielding after use and he said that he does not re-shield with the green cover but he prepares the insulin and puts it into the refrigerator for use when it is time to take his injection. Consumer is new to using pen needles and did not seem very sure of how to use. He stated that he does not re-use.Lot #: 3276406.Catalog #: 320550.Date of Event: unknown.Samples available: sending mail kit.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943410 ↗
  • MalfunctionSep 30, 2026Break

    Consumer Reported found from this box only 30 pen needles that broke off into different injection sites. Stomach Leg and arm.Wife able to remove the needle with a tweezer each time.Denied reuse of needles. does not pinch up site. Before this box always used the BD 8mm pen needles. DCLot # 2333537Catalog# 320550Date of Event unknownDate of event 04.29.2029Sample Status Awaiting sample unused from this box

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943343 ↗
  • MalfunctionSep 30, 2026Break

    Consumer Reported found from this box only 30 pen needles that broke off into different injection sites. Stomach Leg and arm.Wife able to remove the needle with a tweezer each time.Denied reuse of needles. does not pinch up site. Before this box always used the BD 8mm pen needles. DCLot # 2333537Catalog# 320550Date of Event unknownDate of event 04.29.2029Sample Status Awaiting sample unused from this box

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943273 ↗
  • MalfunctionSep 30, 2026Failure to Deliver

    Consumer Reported found about 2 dozen pen needles from two different lot number. Unknown amount on each lot #. Needles would not prime. noticed after removed from pen the non patient end needle to be bent. DC Lot # 3017179 Lot # 3025239Catalog# unknown - 32g 4mm 2nd gen Date of Event Unknown Sample Status Awaiting sample.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943240 ↗
  • MalfunctionSep 30, 2026Break

    Consumer Reported found from this box only 30 pen needles that broke off into different injection sites. Stomach Leg and arm.Wife able to remove the needle with a tweezer each time.Denied reuse of needles. does not pinch up site. Before this box always used the BD 8mm pen needles. DCLot # 2333537Catalog# 320550Date of Event unknownDate of event 04.29.2029Sample Status Awaiting sample unused from this box.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943155 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K212015. One submission often covers many models, so check the product and lot details in each record.

  • Z-2424-2024 · Jun 18, 2024 · Open, Classified
    BD Nano 2nd Gen Pen Needles 32G X 4mm (5-count sample contained in Patient Counseling Sample Toolbox -Catalog 6484755/Lot Number: 3750144)-Sterile, single use devices designed for
  • Z-3179-2026 · Jul 26, 2023 · Open, Classified
    EWO_02512 (Catalog number 320550) For Investigational Use Only Nano Pro Un-Cannulated Demo Samples Quantity 100

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources and updates