BD Pet Syringes, 0.3mL x 12MM U-40 323000Recalls & reported problems
Becton, Dickinson and Company Inc · SYRINGE 0.3ML 29GA 12MM U40 VET 10BAG 500CS · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This BD Pet Syringe is a small 0.3 mL veterinary insulin syringe used with pets. The 34 reports mostly describe barrels with missing, faint or crooked scale markings, plus syringes breaking, leaking, or having a loose or bent rubber plunger. Two injuries were reported, including needle sticks and bleeding, and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00382903230006), received between Apr 9, 2018 and Oct 7, 2024 — classified as 2 injuries, 32 malfunctions.
The most frequently reported specific device problem is “Volume Accuracy Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Volume Accuracy Problem6
- Leak/Splash5
- Break4
- Defective Component4
- Defective Device4
- Detachment of Device or Device Component3
- Failure to Deliver3
- Material Separation3
Patient problems
- No Clinical Signs, Symptoms or Conditions20
- Needle Stick/Puncture4
- No Consequences Or Impact To Patient4
- No Known Impact Or Consequence To Patient4
- No Information2
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 7, 2024Volume Accuracy Problem; Volume Accuracy Problem
PET OWNER REPORTED FINDING TWO SYRINGES THAT WERE MISSING THE "SCALE MARKINGS" BEFORE USE. LOT: 3346004, CATALOG: 323000, DATE OF EVENT: UNKNOWN, SAMPLES: YES CL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20384347 ↗MalfunctionMay 7, 2024Leak/Splash; Leak/Splash
VETERINARIAN REPORTED ON BEHALF OF CONSUMER THAT THE RUBBER STOPPERS ARE CROOKED IN THE BARREL. SHE STATED THAT THERE ARE ABOUT 5 SYRINGES WITH THE PROBLEM. VET DID NOT HAVE ALL OF THE INFORMATION SO SHE SAID SHE WILL ASK THE CONSUMER TO CONTACT US. SHE WAS ABLE TO PROVIDE THE LOT #. LOT #: 3065610 CATALOG #: UNKNOWN DATE OF EVENT: UNKNOWN SAMPLES: AVAILABILITY UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19263122 ↗MalfunctionJun 28, 2023Defective Component; Defective Component
IT WAS REPORTED WHILE USING BD PET SYRINGES VETERINARY INSULIN SYRINGES THE PLUNGER WAS LOOSE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER STATED, TWO SYRINGES ARE USED PER DAY. STATED, THE PLUNGER ROD IS REALLY LOOSE STATED, SCALE MARKING FONTS ARE LARGER AT TOP OF BARREL AND MUCH SMALLER AT THE BOTTOM OF BARREL STATED, SOMETIMES FONT IS SO SMALL THAT HE CANNOT SEE THE NUMBERS. LOT: 2206995 CATALOG: DATE OF EVENT: UNKNOWN SAMPLES: YES CL U40 SYRINGE COMPLAINT RECEIVED VOICEMAIL FROM PET OWNER STATING HE'S FINDING "BENT NEEDLES" PRIOR TO INJECTION STATED, BARREL SIZES SEEM DIFFERENT, "SOME ARE LARGE AND SAME ARE SMALL".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17220252 ↗MalfunctionJun 28, 2023Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED WHILE USING BD PET SYRINGES VETERINARY INSULIN SYRINGES THERE WERE SCALE MARKING ISSUES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER STATED, TWO SYRINGES ARE USED PER DAY. STATED, THE PLUNGER ROD IS REALLY LOOSE STATED, SCALE MARKING FONTS ARE LARGER AT TOP OF BARREL AND MUCH SMALLER AT THE BOTTOM OF BARREL. STATED, SOMETIMES FONT IS SO SMALL THAT HE CANNOT SEE THE NUMBERS. LOT: 2206995. CATALOG: DATE OF EVENT: UNKNOWN. SAMPLES: YES CL. U40 SYRINGE COMPLAINT. RECEIVED VOICEMAIL FROM PET OWNER STATING HE'S FINDING "BENT NEEDLES" PRIOR TO INJECTION STATED, BARREL SIZES SEEM DIFFERENT, "SOME ARE LARGE AND SAME ARE SMALL".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17220220 ↗MalfunctionApr 25, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT THE BD¿ PET SYRINGES VETERINARY INSULIN SYRINGE EXPERIENCED SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED FOUND 1 SYRINGE WITHOUT ANY SCALE MARKING ON THE BARREL OF SYRINGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16812463 ↗MalfunctionMar 29, 2023Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD¿ PET SYRINGES VETERINARY INSULIN SYRINGES FROM LOTS 2332719 AND 2003427 HAD SLANTED SCALE MARKINGS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "PET OWNER REPORTED "SCALE MARKINGS" ON SOME OF THE BARRELS ARE NOT STRAIGHT. STATED, THE MARKINGS ARE "SLANTED"? STATED, TWO BOXES AFFECTED"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16640347 ↗MalfunctionMar 15, 2023Defective Component; Failure to Deliver; Defective Component; Failure to Deliver
IT WAS REPORTED THAT THE BD PET SYRINGES VETERINARY INSULIN SYRINGE WAS STUCK IN VIAL WHILE DRAWING THE INSULIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: HE ALSO STATED THAT THE NEEDLE OF ANOTHER SYRINGE WAS STUCK IN THE VIAL WHILE DRAWING THE INSULIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16548103 ↗MalfunctionMar 13, 2023Break; Break
IT WAS REPORTED THAT THE BD PET SYRINGES VETERINARY INSULIN SYRINGE BROKE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WHEN HE TRIED TO REMOVE THE SYRINGE, THE VIAL FELL ON THE FLOOR AND BROKE WITH THE NEEDLE/SYRINGE IN IT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16534889 ↗MalfunctionMar 13, 2023Break; Break
IT WAS REPORTED THAT THE BD PET SYRINGES VETERINARY INSULIN SYRINGE BROKE WHEN REMOVING CAP. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: NEEDLE FELL APART UPON REMOVING CAP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16531504 ↗MalfunctionFeb 10, 2023Defective Device; Defective Device
IT WAS REPORTED THAT THE BD PET SYRINGES VETERINARY INSULIN SYRINGE WAS DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WANTED TO REPORT THIS BENT SYRINGE FROM A LOT OF U-40, SYRINGES I PURCHASED FOR MY DOG.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16358213 ↗MalfunctionApr 15, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE BD 0.3ML VET SYRINGE EXPERIENCED HUB SEPARATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER REPORTED NEEDLE HUB SEPARATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14120779 ↗MalfunctionApr 14, 2022Material Separation; Material Separation
IT WAS REPORTED THAT THE BD PET SYRINGES VETERINARY INSULIN SYRINGE EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER REPORTED THAT THE NEEDLE HUB SEPARATED AND STAYED INSIDE OF THE SHIELD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14116374 ↗MalfunctionFeb 7, 2022Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED WHILE USING BD PET SYRINGES 3/10ML 12.7MM X 29G THE SCALE MARKINGS WERE MISALIGNED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 323000 BATCH NO. 1102439. IT WAS REPORTED THAT SCALE MARKINGS ARE CROOKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13472333 ↗MalfunctionJan 26, 2022Break; Delivered as Unsterile Product; Break; Delivered as Unsterile Product
IT WAS REPORTED WHILE USING BD PET SYRINGES 3/10ML 12.7MM X 29G A PLUNGER WAS FOUND BROKEN AND NON STERILE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BATCH#: 1102439, MATERIAL#: 323000. IT WAS REPORTED THAT THERE WAS A BROKEN PLUNGER ROD AND CAP. THE SYRINGE WAS NOT USED AND WAS RECEIVED BROKEN OUT OF THE PACKAGE. THE INNER PACKAGE (BAG) AND OUTER PACKAGING (BOX) SHOWED NO SIGNS OF DAMAGE. PRODUCT HAS BEEN SAVED IF NEEDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13359759 ↗MalfunctionDec 22, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE BD VETERINARY INSULIN SYRINGE EXPERIENCED HUB SEPARATION FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER REPORTED NEEDLE HUB SEPARATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13065542 ↗MalfunctionDec 1, 2021Defective Component; Detachment of Device or Device Component; Defective Component; Detachment of Device or Device Component
IT WAS REPORTED THAT BD SYRINGE 0.3ML VET U40 29GA 12.7MM EXPERIENCED 2 CASES OF HUB SEPARATION FROM THE DEVICE, AND 2 CASES OF STOPPER DEFORMATION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: I HAVE BEEN USING THIS PRODUCT FOR MORE THAN 3 1/2 YEARS. I HAVE NOTICED AN INCREASE IN DEFECTIVE SYRINGES. THE DEFECTS INCLUDE BUT ARE NOT LIMITED TO: -NEEDLE BREAKS OFF AND IS STUCK IN THE RED CAP -NEEDLE IS BENT IN THE RED CAP -BLACK INDICATOR IS DISTORTED /DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12906364 ↗MalfunctionJul 19, 2021Leak/Splash; Leak/Splash
IT WAS REPORTED THAT BD PET SYRINGES, 0.3ML X 12MM U-40 THE PLUNGER ROD WAS DIFFICULT TO MOVE AND THE HUB SEPARATED FROM DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: PET OWNER REPORTED THAT THE NEEDLE HUB SEPARATED AND WAS STUCK INSIDE OF THE SHIELD. ALSO REPORTED THAT THE PLUNGER ROD IS DIFFICULT TO MOVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12193571 ↗MalfunctionJul 6, 2021Material Separation; Material Separation
IT WAS REPORTED THAT 3 VET SYRINGE 0.3ML X 29G X 12.7MM HUBS SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CONSUMER REPORTED THAT WHEN NEEDLE SHIELDS WERE REMOVED THE NEEDLE HUBS SEPARATED. DATE OF EVENT: UNKNOWN. SAMPLES: YES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12125820 ↗MalfunctionJun 14, 2021Defective Device; Defective Device
IT WAS REPORTED THAT THE SYRINGE 0.3ML VET U40 29GA 12.7MM 10BAG EXPERIENCED DEVICE DAMAGE/DEFORMATION WHILE STILL CONSIDERED OPERABLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO:323000 BATCH NO: 0174596. PET OWNER REPORTED NEEDLES CROOKED/BENT. INCONSISTENT NEEDLE COLOR, STATED THAT SOME OF THE NEEDLES APPEAR TO HAVE A DULL RATHER THAN SHINY. SOME OF THE SYRINGES HAVE TINY CRACKS ON BARREL TOWARDS THE TOP. CONSUMER DOES NOT RE-USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11996411 ↗MalfunctionJun 8, 2021Material Separation; Material Separation
IT WAS REPORTED THAT 1 BD VET SYRINGE 0.3ML X 29G X 12.7MM HUB SEPARATED FROM THE DEVICE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CONSUMER REPORTED THE NEEDLES ARE DULL, WILL NOT PENETRATE THE SKIN AND THE NEEDLE HUB SEPARATED. DATE OF EVENT: UNKNOWN, SAMPLES: AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11963237 ↗MalfunctionDec 22, 2020Volume Accuracy Problem; Volume Accuracy Problem
IT WAS REPORTED THAT A VET SYRINGE 0.3ML X 29G X 12.7MM 10 BAG HAD SCALE MARKING ISSUES DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT THE U-40 SYRINGES ARE CAUSING THE PET OWNER TO USE THE INSULIN QUICKER AND SHE BELIEVES SHE IS MEASURING THE DOSE INCORRECTLY. ALSO, THE PET'S NUMBERS HAVE FLUCTUATED AND THE PET IS MORE TIRED LATELY. VERBATIM: PET OWNER STATED SHE HAS U-100 INSULIN AND WAS PRESCRIBED THE U-100 SYRINGES. STATED THE PHARMACY INFORMED HER OF THE U-40 SYRINGES AND ADVISED SHE SHOULD TRY THEM. PET OWNER STATED SHE IS USING THE INSULIN QUICKER THEN WHAT SHE SHOULD BE. STATED SHE GIVES 4 UNITS A DAY AND BELIEVES SHE IS MEASURING THE DOSES INCORRECTLY RESULTING IN USING THE INSULIN FASTER. STATED HER PET'S NUMBERS HAVE BEEN FLUCTUATING AND HER PET IS MORE TIRED LATELY. IT WAS RECOMMENDED THAT PET OWNER CONSULT WITH HER VET. "
Patient problems: No Information
MAUDE report 11060876 ↗MalfunctionDec 10, 2020Break; Break
IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF VET SYRINGE 0.3ML X 29G X 12.7MM 10 BAG BROKE AT THE CANNULA DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT THE NEEDLE HEAD COMES OFF WHEN THE CAP IS REMOVED. VERBATIM: EMAIL RECEIVED¿2020-11-17 14:30:21. I PURCHASE A BOX OF 100 PET SYRINGES AND SEVERAL HAVE BEEN DEFECTIVE... NEEDLE HEAD COMES OFF WHEN CAP IS REMOVED. LOT # 9266921.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10988600 ↗MalfunctionSep 8, 2020Failure to Deliver; Failure to Deliver
IT WAS REPORTED THAT VET SYRINGE 0.3ML X 29G X 12.7MM 10 BAG WAS UNABLE TO ASPIRATE. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO:323000 BATCH NO: 9266921. IT WAS REPORTED THAT THE CONSUMER, A PET OWNER, FEELS THE PLUNGER IS UNABLE TO DRAW UP VETSULIN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10505524 ↗MalfunctionAug 20, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED DURING USE THE VET SYRINGE 0.3ML X 29G X 12.7MM 10 BAG EXPERIENCED SHIELD/CAP DOES NOT ATTACH/DETACH AND NEEDLESTICK INJURY - SELF. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CUSTOMER STATED THAT WITH FOUR SYRINGES IT WAS HARD TO REMOVE SHIELDS AND THE NEEDLE HUB ASSEMBLY SEPARATED WHEN REMOVED. ALSO REPORTED, ON ONE OF THESE SYRINGES SHE PULLED HARD AND NEEDLE STUCK HER FINGER. NO MEDICAL ATTENTION SOUGHT.
Patient problems: Needle Stick/Puncture
MAUDE report 10434288 ↗InjuryAug 18, 2020Material Puncture/Hole; Material Protrusion/Extrusion; Material Puncture/Hole; Material Protrusion/Extrusion
IT WAS REPORTED THAT VET SYRINGE 0.3ML X 29G X 12.7MM 10 BAG NEEDLE WENT THROUGH THE SHIELD AND CAUSED A NEEDLE STICK. THIS WAS DISCOVERED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 323000 BATCH NO. UNKNOWN. IT WAS REPORTED THAT WHEN RE-CAPPING SYRINGE, NEEDLE WENT THROUGH SHIELD AND CAUSED NEEDLE STICK. NO MEDICAL ATTENTION RECEIVED. CONSUMER REPORTED IN HER PRIOR BOX USED FOR HER CAT HAD A SYRINGE WHEN RECAPPED THE SHIELD TO THE NEEDLE, AFTER INJECTING HER CAT, THE NEEDLE WENT THRU THE SHIELD AND STUCK HER FINGER. NO MEDICAL ATTENTION RECEIVED. CLAIMS FINGER HEALED FINE WITHOUT ISSUES. ADVISED THIS PET OWNER NOT TO RESHIELD THE SYRINGES, JUST PLACE THEM IN THE SHARPS CONTAINER.
Patient problems: Needle Stick/Puncture
MAUDE report 10418561 ↗
Other Syringe, piston devices
- BD Luer-Lok 309657Becton, Dickinson And Company · Syringe, piston1,266 reports
- BD Luer-Lok 309653Becton, Dickinson And Company · Syringe, piston948 reports
- BD Luer-Lok 309628Becton, Dickinson And Company · Syringe, piston862 reports
- N/A 302995Becton, Dickinson And Company · Syringe, piston1 recall537 reports
- BD Plastipak Syringe 300865Becton, Dickinson And Company · Syringe, piston515 reports
- N/A 302830Becton, Dickinson And Company · Syringe, piston462 reports
- N/A 302832Becton, Dickinson And Company · Syringe, piston424 reports
- BD Ultra-Fine 328468Becton, Dickinson and Company Inc · Syringe, piston323 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page