Skip to content

Q-Syte 385157Recalls & reported problems

Becton, Dickinson And Company · Bi-Extension Set 15 cm (6 IN) · Class 2

All models of Q-Syte (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00382903851577), received between Sep 18, 2024 and Feb 2, 2026 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem5
  • Leak/Splash4
  • Backflow2
  • Material Integrity Problem2
  • Break1
  • Contamination1
  • Fluid/Blood Leak1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • Discomfort1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2026Leak/Splash; Connection Problem

    INCIDENT HAPPENED IN ER DEPARTMENT WHERE AFTER CANNULATION STAFF ATTACHED QSYTE BI EXTENSION CAT NO 385157 ON IV CATHETER 18 G AND WHEN THEY TRIED TO LOCK EXTENSION BY ROTATING SPIN NUT IT GOT FREED DURING TIGHTENING OF IT WHICH RESULTED IN SPILLAGE OF BLOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24235733 ↗
  • MalfunctionDec 16, 2025Manufacturing, Packaging or Shipping Problem; Manufacturing, Packaging or Shipping Problem

    NO ADDITIONAL INFORMATION.

    Patient problems: Discomfort

    MAUDE report 23822536 ↗
  • MalfunctionNov 3, 2025Backflow; Backflow

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23453130 ↗
  • MalfunctionOct 27, 2025Backflow; Backflow

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23390508 ↗
  • MalfunctionAug 1, 2025Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    IT WAS REPORTED THAT BD Q-SYTE BROKEN THREAD WITH DISCONNECTION AND LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE THREADS OF Q-SYTE BI EXTENSION SPIN NUT IS GETTING BROKEN DURING ATTACHING WITH CANULA. CONNECTION GETTING ACCIDENTALLY DISLODGED FROM CANULA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22686517 ↗
  • MalfunctionJun 25, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET LOOSE COMPONENT AND LEAK THE SPIN NUT KEEPS SPINNING AND DOESN'T FIX TO THE CANNULA, LEADING TO DISCONNECTION AND LEAKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22328937 ↗
  • MalfunctionJun 16, 2025Connection Problem; Connection Problem

    DESCRIPTION THE LUER LOCK FIXATOR DOESN'T GET FIXED TO THE CANNULA, THE THREAD KEEPS ROTATING AND GETS DISCONNECTED, WHICH LEADS TO LEAKAGE OF BLOOD AND MEDICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22210070 ↗
  • MalfunctionJun 9, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET HAD CONNECTION ISSUES INCIDENT HAPPENED IN ER DEPARTMENT WHERE AFTER CANNULATION STAFF ATTACHED QSYTE BI EXTENSION CAT NO 385157 ON IV CATHETER 22G AND WHEN THEY TRIED TO LOCK EXTENSION BY ROTATING SPIN NUT IT GOT FREED DURING TIGHTENING OF IT HENCE CAUSING RISK OF ACCIDENTAL DISLODGEMENT OF CATHETER, THEY HAVE PROVIDED SAMPLES OF 2 BATCHES AND REPLACED QSYTE BI EXTENSION IN PATIENT WITH NEW PIECE OF QSYTE BI EXTENSION WHICH WORKED WITHOUT ANY PROBLEM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22169250 ↗
  • MalfunctionJun 9, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET HAD CONNECTION ISSUES INCIDENT HAPPENED IN ER DEPARTMENT WHERE AFTER CANNULATION STAFF ATTACHED QSYTE BI EXTENSION CAT NO 385157 ON IV CATHETER 22G AND WHEN THEY TRIED TO LOCK EXTENSION BY ROTATING SPIN NUT IT GOT FREED DURING TIGHTENING OF IT HENCE CAUSING RISK OF ACCIDENTAL DISLODGEMENT OF CATHETER, THEY HAVE PROVIDED SAMPLES OF 2 BATCHES AND REPLACED QSYTE BI EXTENSION IN PATIENT WITH NEW PIECE OF QSYTE BI EXTENSION WHICH WORKED WITHOUT ANY PROBLEM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22168792 ↗
  • MalfunctionMay 27, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET COLLAR SPINS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LUER LOCK PLUG KEEPS ROTATING, THERE WAS AN EARLIER PIR RAISED, TODAY WE RECEIVED MORE UNITS FROM WARD SIDE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22091755 ↗
  • MalfunctionMay 26, 2025Leak/Splash; Connection Problem; Leak/Splash; Connection Problem

    IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET HAD A POOR CONNECTION WITH LEAK. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE LUER LOCK FIXATOR DOESN'T GET FIXED TO THE CANNULA, THE THREAD KEEPS ROTATING AND GETS DISCONNECTED, WHICH LEADS TO LEAKAGE OF BLOOD AND MEDICATIONS. 08-MAY-2025- RECEIVED AN EMAIL RESPONSE. QUANTITY INVOLVED-1. CAN YOU CONFIRM IS THERE ANY PATIENT IMPACTED? - NO. CAN YOU CONFIRM THAT THERE ARE ANY SERIOUS INJURIES THAT OCCUR TO THE PATIENT? NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22084658 ↗
  • MalfunctionMay 15, 2025Loose or Intermittent Connection; Loose or Intermittent Connection

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22028187 ↗
  • MalfunctionMay 7, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE BD Q-SYTE 15CM SMALL BORE BI-EXT SET HAD A LOOSE CONNECTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE LUER LOCK PLUG KEEPS ROTATING AND GETS DISCONNECTED SOON AFTER SECURING. ADDITIONAL INFORMATION PROVIDED: 1. CAN YOU CONFIRM IS THERE ANY PATIENT IMPACTED? - YES,THERE WAS A BLOOD SPILLAGE FOR ONE PATIENT, INCIDENT REPORT HAS BEEN RAISED BY THE STAFF NURSE. 2. CAN YOU CONFIRM THAT THERE ARE ANY SERIOUS INJURIES THAT OCCUR TO THE PATIENT? - NO SERIOUS INJURIES. 3. WHAT IS THE CURRENT NURSE STATUS? IS SHE/HE OK? - NURSE IS FINE. 4. QUANTITY INVOLVED - (B)(4) QUANTITIES IN TOTAL OF THE SAME BATCH. 5. DATE OF OCCURRENCE- (B)(6) 2025. THERE WAS A TYPE IN THE DATE OF OCCURRENCE, IT IS (B)(6) 2025.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21981528 ↗
  • MalfunctionJan 21, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET LEAKED DRUG LEAKAGE FROM Q-SYTE BI EXTENSION IN RADIOLOGY DEPARTMENT. THEY HAVE GIVEN CT CONTRAST AT 5.5ML PER SECOND WITH OVERALL VOLUME OF 80ML THROUGH PRESSURE INJECTOR. THEY HAVE NOTICED DRUG LEAKAGE FROM Q-SYTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21203262 ↗
  • MalfunctionSep 18, 2024Contamination; Contamination

    IT WAS REPORTED THAT BD Q-SYTE 15CM SMALL BORE BI-EXT SET HAD FOREIGN MATTER . DRIED BLOOD LIKE STAIN OBSERVED IN ONE PORT OF Q-SYTE BI EXTENSION (NEAR THE HUB) IN NEW SEALED PACK. OBSERVED BY NURSE IN CHARGE OF ICU OF (B)(6). THE INCIDENT IS ESCALATED TO ICN AND (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20253045 ↗

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page